MODIFICATION
J -- REPAIR OF VG-FISONS PLATFORM II LC/MASS SPECTROMETER
- Notice Date
- 4/27/2006
- Notice Type
- Modification
- NAICS
- 811219
— Other Electronic and Precision Equipment Repair and Maintenance
- Contracting Office
- Environmental Protection Agency, Office of Research and Development, Ecosystems Research Division (LAB), 960 College Station Rd., Athens, GA, 30605
- ZIP Code
- 30605
- Solicitation Number
- EP06L000116
- Response Due
- 5/8/2006
- Archive Date
- 5/23/2006
- Point of Contact
- Rick Pittman, Contract Specialist, Phone 706-355-8010, Fax 706-355-8026, - Rick Pittman, Contract Specialist, Phone 706-355-8010, Fax 706-355-8026,
- E-Mail Address
-
pittman.rick@epa.gov, pittman.rick@epa.gov
- Description
- The Fisons Platform II LC/mass spectrometer has not been operational for several months. The problem with the instrument is that it appears to be pumping down (after final replacement of seals, etc.), but the ion gauge is not coming on (and the instrument is not operational until this ion gauge comes on and out of its vacuum tripped status). We replaced the penning ion gauge (in case oxidation was preventing it from operating properly), and this did not fix the problem. This instrument is over 10 years old is not one that is commonly serviced.
- Place of Performance
- Address: 960 College Station Road, Athens, GA
- Zip Code: 30605
- Country: US
- Zip Code: 30605
- Record
- SN01036807-W 20060429/060427221337 (fbodaily.com)
- Source
-
FedBizOpps Link to This Notice
(may not be valid after Archive Date)
| FSG Index | This Issue's Index | Today's FBO Daily Index Page |