Loren Data's SAM Daily™

fbodaily.com
Home Today's SAM Search Archives Numbered Notes CBD Archives Subscribe
FBO DAILY ISSUE OF APRIL 16, 2010 FBO #3065
SOLICITATION NOTICE

66 -- Low Background Shielding for Detecters in Gamma Spec

Notice Date
4/14/2010
 
Notice Type
Presolicitation
 
NAICS
334516 — Analytical Laboratory Instrument Manufacturing
 
Contracting Office
Idaho OperationsU.S. Department of EnergyIdaho Operations1955 Fremont AvenueIdaho FallsID83415
 
ZIP Code
83415
 
Solicitation Number
DE-SOL-0001718
 
Response Due
4/28/2010
 
Archive Date
5/28/2010
 
Point of Contact
Minor, Kelleen L 208-526-0144, minorkl@id.doe.gov
 
E-Mail Address
Minor, Kelleen L
(minorkl@id.doe.gov)
 
Small Business Set-Aside
N/A
 
Description
The Radiological and Environmental Sciences Laboratory (RESL) has a requirement for eight (8) High Purity Geranium Gamma-Ray Detectors Shields. The specifications for these shields follow: The interior shield dimensions must have at least a 28-cm diameter by 40-cm height. The shield must be of a graded shield design that has at least 4 inches of solid-cast lead with at least O.5-mm of tin sheet liner and 1.6-mm copper sheet liner on the interior. The lead, tin and copper shield should be contained within at least 9.5 mm thick low-carbon steel casing. The bottom of the shield should have at least a 3.5-inch opening with a lead plug that will accommodate ORTEC X-COOLER systems. The shield should be supported on a rigid steel table with the dimensions approximately 2-ft x 2-ft by at least 30-in high. The shield must have access from the top with either a split or single opening door and must be easy to open and close. The shield must have access from the front with either a split or single open door and must be easy to open and have a latch that will keep the doors closed during use. The shield should include an installation and maintenance manual, touch-up paint and lifting eyes. The lid must stay shut when in use and the bidder should describe how this is achieved. Bids that contain shields constructed out of different materials than described in these specifications must demonstrate how they are equivalent to the 4-inch lead graded liner shield design and will also be considered. Bids must also include freight charges to Idaho Falls, Idaho. The bid that is most advantageous to the government will be selected.
 
Web Link
FBO.gov Permalink
(https://www.fbo.gov/spg/DOE/PAM/HQ/DE-SOL-0001718/listing.html)
 
Record
SN02122196-W 20100416/100414235513-8e16e07a01ee788c57d0c0030e7f0626 (fbodaily.com)
 