MODIFICATION
43 -- high-vacuum turbomolecular pumps, cryogenic pumps and dry roughing pumps - Revised Table 1
- Notice Date
- 8/6/2013
- Notice Type
- Modification/Amendment
- NAICS
- 333911
— Pump and Pumping Equipment Manufacturing
- Contracting Office
- Department of Commerce, National Institute of Standards and Technology (NIST), Acquisition Management Division, 100 Bureau Drive, Building 301, Room B129, Mail Stop 1640, Gaithersburg, Maryland, 20899-1640, United States
- ZIP Code
- 20899-1640
- Solicitation Number
- SB1341-13-RP-0076
- Archive Date
- 9/4/2013
- Point of Contact
- Eric Bubar, Phone: 3019753635
- E-Mail Address
-
eric.bubar@nist.gov
(eric.bubar@nist.gov)
- Small Business Set-Aside
- N/A
- Description
- Revised Table 1 with list of 18 pumps Amendment 1 The purpose of this Amendment 1 is to provide Questions and Answers. Q1: In regards to Pump 18, the cryo pump, does NIST want an ISO-200 inlet flange? A1: They are CTI 8 cryo pumps. A conflat flange is required. See the attached revised Table 1. Q2: In regards to Pump 18, the cryo pump, what type of temperature monitoring is required? A2: Silicon diode sensor Q3: In regards to Pump 18, the cryo pump, does NIST want an onboard pump? A3: No