COMMERCE BUSINESS DAILY ISSUE OF JANUARY 6,1995 PSA#1257U.S. DEPARTMENT OF THE INTERIOR, U.S. BUREAU OF MINES-CONTRACT SECTION
B, BUILDING 20 - DENVER FEDERAL CENTER, DENVER, CO 80225-0086 66 -- ICP MASS SPECTROMETER SOL 1432029510001 63 DUE 022395 POC
Contract Specialist, JANICE L COFFELT, 303-236-0257 X223 The U.S.
Bureau of Mines has a requirement to purchase an Inductively Coupled
Plasma Mass Spectrometer (ICP-MS). The ICP-MS system shall be a
production instrument, not a prototype. The system shall be capable of
performing automated, unattended multi-element analysis with high
precision and accuracy. The following requirements are listed: (1)
inductively coupled plasma ion source, a quadrupole mass filter, and an
ion detection system, (2) a recirculating water chiller that will
provide an adequate circulating temperature to maintain stable system
operation, (3) an autosampler capable of handling at least 200 samples
and standards, (4) two sets of operation, methods, and repair manuals.
An industry standard Pentium computer with math coprocessor operating
at 50 MHz with SVGA video card, 16MB RAM, 452 MB hard disk drive, two
1.44 MB disk drives, Laser Jet 4 printer. SVGA monitor and all cables
necessary to connect peripherals to the computer. Software must consist
of a multi-tasking software package for complete operation of the
ICP-MS and data processing. The RF generator must be built into the
instrument and a shelf provided for the autosampler. Warranty, service
and technical support must be provided. Instrument performance tests
must be conducted by the service engineer after the ICP-MS has been
installed and before the instrument can be accepted.(0004) Loren Data Corp. http://www.ld.com (SYN# 0232 19950105\66-0003.SOL)