COMMERCE BUSINESS DAILY ISSUE OF APRIL 21,1995 PSA#1330--the Manufacturer Must Specify the Procedure for the, Mr. Yaakov
Kruthammer, Deputy Administrative Director, Chaim Sheba Medical Center,
Tel Hashomer 52621, Israel 99 -- ISRAEL: DUE 050795 POC Tel 972-3-530-3030 FAX 972-3-530-3568
9022 apparatus based on the use of radiation the Chaim Sheba medical
Center invites suppliers to submit a quotation for a remote after
loader for high dose rate brachytherapy according to the following
specifications tender 133/mtz. --The radiation source Must be of type
ir 192 with 370GBQ (loci); --the machine Must have at least 18 channels
for transmission of the radiation source; --each channel of the
radiation source Must have a number of dwell points to enable treatment
the number of dwell points shall correspond to the size of the
radiation source detailed data are required regarding the size of the
radiation source and the number of corresponding dwell points;
--maximum range of radiation source should be not less than 130CM; --
distance between radiation source dwell points Must be variable the
distance between the dwell points Must be specified; preference will be
given to equipment with maximum variability in distances; --accuracy in
targeting the radiation source Must be about lmm; --the machine Must be
equipped with a radiation protection system to prevent radiation dose
rate exceeding 2 5 micro sievert/hr at a distance of lm from the
machine (when the radiation source is inside the exchange of the
radiation source (once every three months) preference will be given to
a machine with an exchange Procedure that is largely automatic in
order to avoid exposure to the staff the Manufacturer Must give
assurances that the first four exchanges of the radiation source are
performed by his representative and on his account the total Cost
involved in the exchange of the radiation source over a period of five
years as of the second year Must be specified; -- five-year warranty
for all parts of the radiation source guiding tubes to ensure accurate
and proper functioning listing of connectors of all types and locking
parts of the tubes is to be attached; --the radiation source guiding
tubes Must have the ability to bend at a mimimum radius of 15MM without
affecting the transmission of the radiation source, --the machine will
have a computerized system for planning and treatment with the
following abilities -automatic correction for the decay of the
radiation source; -shielding radiation calculation; -direct transmittal
of data from the planning station to the operation and control station
of the machine; -registration of all the treatment data and the
results of the calculations; --s a F e T y -the machine Must have a
doubling timer system; -the machine Must have an additional control
system to enable the treatment according to the original plan; -the
machine Must have a self-testing system to include -battery-backup
-doubling timer system -computer control system -internal electric
interconnectors -range limit sensors -the machine Must comply with
international safety standards such as fda or a parallel objective
Institution the Manufacturer Must Specify the standard with which the
machine complies and provide appropriate documentation; -the
Manufacturer Must supply an external radiation testing system
independent of the treatment machine, to be positioned in the treatment
room; --preference will be given to a machine with the ability for
upgrading to a pulse treatment mode the Manufacturer Must Specify this
ability and its costs in detail; --a P P L I C a T 0 R s the
applicators shall be suitable for the following usages -intracavelary
treatment -interstitial treatment -intraluminal treatment - surface
treatment --preference will be given to a machine with a wide range of
applicators; --the machine Must have all the necessary accessories for
proper functioning, such as -storage containers to exchange a radiation
source with 2 channels - storage C upboard for the guiding tubes
-dosimetry system including proper phantom -accessories for cessation
and support of the guiding tubes and applicators -emergency container
for the storage of the radiation source --the Manufacturer Must provide
details regarding the disposable accesories supplied with the machine
and their Cost; --the machine Must be suitable for a one-phase 230V,
50HV Power supply in case the Power supply is three- phased this should
be clearly stated; --the Manufacturer Must have a known local agent
able to provide technical support and know-how, service and spareparts
according to the professional standards; --T R a I N I N G assurances
Must be given for training of hospital staff at the Manufacturer's
production facility, as follows -to two physicists - for planning of
treatment -to two operating technicians - for operation of machine -to
two technicians - for service and maintenance (at the same level as
the course given to Manufacturer's service technicians) --the system
Must be supplied with two operating manuals and two service manuals; --
W a R R a N T y 24 month-warranty period for the machine after
completion of a 6 month-running-in period from the date the equipment
is installed in the hospital if during the running-in period faults are
found in the machine that prevent its proper functioning, the
running-in period will be prolonged by the number of days the machine
was not functional the Manufacturer, through his local agent, Must give
written assurances that the the machine will not be disfunctional for
more than 96 hours the Manufacturer Must also assure immediate response
to calls for repair; --the Manufacturer Must submit a list of critical
accessories and parts and their costs --the offer Must include price,
weight, total fob price from port nearest to Israel the supplier is
requested to indicate the date of supply quotations should be submitted
not later than May 7, 1995 at 14 00 at tender box number 1, building
64B, Chaim Sheba medical Center, Tel Hashomer 52621 for additional
information contact Yaakov Kruthammer, Tel 972-3-530-3225, fax:
972-3-530- 3568 for assistance, U.S. companies May contact irit wolff,
FCS Tel Aviv, Tel 972-3-510-7213, fax 972-3- 510-7215 Please send a
copy of your response to: Commercial Section (FCS TOP), American
Embassy, Tel Aviv, Psc 98 Box 100, APO AE 09830, Refer to T0039..