COMMERCE BUSINESS DAILY ISSUE OF MAY 23,1995 PSA#1352NASA, George C. Marshall Space Flight Center, Procurement Office,
Marshall Space Flight Center, AL 35812 66 -- INSTRUMENTS AND LABORATORY EQUIPMENT SOL 8-W-5-02-79516 POC
Contact Patricia Fundum, GP26-V/Contract Specialist, 205-544-8436, or
Teresa Foley, GP26/Contracting Officer This notice constitutes a
Request for Quotation (RFQ) for an -- 8-inch Turbine Flowmeter which
will be used to accurately measure cryogenic liquid propellants into
development rocket engines; All internal flowmeter components (flow
straightners, rotor, shaft, and cone) are designed to prevent
cavatation at all flow conditions; They must be materially compatible
and be capable of measuring liquid hydrogen, liquid oxygen, and KP1
kerosene; All metal to metal contacting parts, where ''fretting'' can
occur, are to be dry film lubricated; Examples of this are the flow
straightener-to-housing contact points and the shaft-to-support points;
All clearances and loading of the flowmeter parts are to be designed to
prevent any ''Chattering'' which could lead to localized burning in
liquid oxygen; Flowmeter will have provisions to prevent rotation of
the internal assembly within the housing; All parts of the meter will
have positive retention, such as the shaft retaining nut; This nut will
be held in place by a lock (tab) washer that is bent around the nut;
Manufacturing specifications and tolerances are very critical for the
correct operation of the flowmeter thereby reducing the risk of any
damage or failure to the development engine--Performance Specifications
shall include the following--(1) Length-16 inches, flange face-to
flange face--(2) Flange-2-8 hole, 150 #ASA, raised face--(3) Flow
Range-60 to 7,000 GPM--(4) Linearity-Better than 0.5 percent over the
range of 60 to 6,000 GPM--(5) Accuracy-0.25 percent or better
throughout the linear range--(6) Repeatability- 0.25 percent throughout
the linear range--(7) Pressure Drop- No greater than 0.4 PSID in liquid
hydrogen at a flow rate of 6,000 GPM @ -423 degree F, or 6.4 PSID in
liquid oxygen @ -297 degree F--(8) Output-Minimum of 500 millivolts RMS
in the linear flow range--(9) Pickoff Coils-Will have two diametrically
opposed coils and coils must be able to operate at temperatures of -425
degrees F--(10) Cavatation Limit-Flow meters are to be designed such
that cavatation will not occur when flowmeter back pressure drops to 6
PSIG (21 PSIA) in liquid hydrogen @ -423 degrees F--Component
Specifications shall include the following--(1) Rotor- Designed to
operate at minimum net positive suction pressure (NPSP) requirements
set by NASA--The rotor blades are helical (twisted) airfoil design to
preclude cavatation at all flow conditions--The final rotor assembly
will be dynamically balanced and designed to yield a precision of 0.15
percent or better--Entire rotor is cast out of 304 CRES--Rotor will
have imbedded magnets to provide higher signal strength--(2) Rotor
Bearings-Must be self-lubricating (dry film)--Rotor bearings must be
sized to minimize individual ball loading and therefore reduce wear and
heating--Rotor is also designed with safety bearings of AMPCO 18 such
that in the event of primary bearing failure, the safety bearings would
prevent radial and axial movement of the rotor--Must provide fluid
coolant holes to rapidly cool bearings and shaft to maintain assembly
preload--In addition these holes help to maintain cooling of the
bearings to prevent wear and overheating--(3) Housing-Will consist of
one 8-inch pipe, SCH 805, TPL 304L seamless and two 8-inch 150LB ANSI
weld neck TP 304L flanges and will have two diametrically opposed
bosses to accommodate pickoff coils--(4) Magnets-Will have a magnetic
strength of 1000 to 1300 gauss and consists of ALINCO 8--(5)
Inspection-100 percent X-Ray and Hydrostst at 300 psig and housing to
be designed for 200 psig working pressure with a 4 to 1 safety
factor--(6) Vendor Requirements-Vendor must have extensive experience
in the manufacturing, calibration, and design to cryogenic turbine
flowmeters--Must be an ISO 9002 certified company--(7)Calibration-Ten
point water calibration is required on the turbine flowmeter with
equivalent Reynolds number calibrations for liquid hydrogen, liquid
oxygen, and RP1 kerosene--(8)Other-If vendor performs other design
modifications or changes, a rationale of each change shall accompany
the quote--The offeror shall provide a description (including
manufacturer, brand, and model no.) in sufficient detail to show that
the product offered meets the specifications included in this
notice--If the end product(s) offered is other than domestic end
product(s) as defined in the clause entitled ''Buy American Act--
Supplies,'' the offeror shall so state and shall list the country of
origin--The offeror shall state in their offer their size status for
this procurement--The SIC code and the size standard for this
procurement are 3824 and 500 employees respectively--Questions
regarding this acquisition must be submitted in writing--Transmission
by facsimile to 205-544-9081 or by Internet e-mail to
patricia.fundum@msfc.nasa.gov is acceptable--Quotations are due 30 days
after publication of this notice in the Commerce Business Daily (CBD)
to the address specified in Item 7 above to the attention of Patricia
Fundum/GP26-V-- Selection and award will be made to the lowest priced,
technically acceptable offeror--Technical acceptability will be
determined by information submitted by the offeror providing a
description in sufficient detail to show that the product offered meets
the Government's requirement--The period of time expected for the
Contracting Officer to solicit/evaluate vendor quotes and issue the
purchase order will be approximately 45 days from publication of this
notice in the CBD--This procurement is a total small business
set-aside.(0139) Loren Data Corp. http://www.ld.com (SYN# 0391 19950522\66-0001.SOL)