to conform with the following specifictions: Analytical Column -
spherical ion-exchange gel analytical column, 4mm inner diameter x
15cm, capable of performing 1000 analysis, Buffer System - buffers
necessary to perform HPLC A1C testing packaged in quantities capable of
performing 1000 analysis, Hemoglobins Separated - F, A1a, A1b, A1c, Ao,
S, and C must identify hemoglobin variants when present, Range of
Values - 4.3-6.1%, Precision - within-run and between run below 5%,
Specificity - measures HbA1C specifically, Interferences - must include
Schiff base removal. An estimated use of 4 packs per year. Packs are to
be compatible with Bio-Rad MDMS-A System. Contract to begin 10/1/95
through 9/30/96 with two one year options. Issue date on or about
8-7-95, bid opening on or about 9-7-95. All requests must be in writng
or by FAX referencing RFP 660-033-95, Attn: D. Simonich, FAX