COMMERCE BUSINESS DAILY ISSUE OF AUGUST 11,1995 PSA#1408U.S. GEOLOGICAL SURVEY, PROCUREMENT AND CONTRACTS, P.O. BOX 25046, MS
204, DFC, DENVER, CO 80225 66 -- HIGH DYNAMIC RANGE ACCCELEROGRAPH WITH INTERNAL TRIAXIAL
FORCE-BALANCE ACCELEROMETERS AND 25-FT. LONG GPS RECEIVER SOL
95-9950-7367 1434-CR-95-QQ-00461 DUE 092595 POC Purchasing Agent, Iris
D. Turner, (303)236-5900 x348 Concerns having the ability to furnish
the supplies decscribed herein are requested to give written
notification (including the telephone number for a point of contact) to
the procurement office above within(45 calendar days from the date of
this synopsis. The contractor shall provide 4 K2 High Dynamic Range
Accelerograph with internal triaxial force-balance accelerometers and
25-feet long GPS Receiver. The K2 High Dynamic Range Accelerograph
include the following specifications: Replacement accelerographs must
be compatible (without modification) with existing accelerographs now
in the field are manufacture by Kinemetrics, Inc. of Pasadena,
California. Three-component high dynamic range accelerograph with
internal triaxial force-balance accelerometers. These units will be
installed in free field locations for the recording of strong ground
motions produced by earthquakes and significantly smaller amplitude
accelerations produced by local mining induced events. Each unit shall
be self-contained and include all eletronics, sensors, removable
memory card and battery in rugged housing. Each unit shall be supplied
with a charger, all appropriate mating connectors, required mounting
hardware and operating instructions. Power: The unit shall operate from
a 115/220 volt AC supply and shall operate for a minimum of 36 hours
after the loss of the AC supply. In addition, the unit shall accept an
external 12 volt DC battery for longer autonomy. The charger supplied
with the unit shall be sufficient to charge both the internal main
battery of the unit and an optional external battery. Dynamic Range: In
order to effectively acquire the range of events envisioned for this
project, the unit shall have a minimum dynamic range of 110 dB. In
order to insure fidelity of the recorded data and maintain a full
effective resolution of 19-bits, gain ranging is not permissible.
Bandwith: The unit's response shall be nominally flat from DC to 50 Hz.
The effective sample rate shall be 200 sample per second. This sample
rate shall be accomplished by over-sampling the data and digitally
processing the data (using a DSP). Digital FIR filters shall be used
for anti-aliasing to insure sufficient attenuation (100 dB) at the
Nyquist frequency and continuous accurate control of the bandwidth.
There should be no channel-to-channel skew meaning that all three
channels are to be sampled simultaneously rather than sequentially.
Triggering Level: The unit's trigger threshold shall be selectable
anywhere within the range of 50 full-scale (i.e. 1 g to 200 micro g for
a +/-2 g full-scale unit). Recording Medium: Each unit shall have one
5 Mbyte PCMCIA compatible (fully compliant to PCMCIA Specification
2.01) removable memory card. An additional PCMCIA complaint slot shall
be provided for a second memory card (the second memory card not to be
supplied as part of this bid). Full-Scale Range: The units shall be
available in both +/-2 g or +/-1 g full-scale ranges. Firmware: The
unit's firmware shall be multi-tasking to allow the unit to be
interrogated (locally or via modem) and to allow data retrieval without
any interruption of data acquisition. The unit's firmware shall be
stored in flash memory rather than EPROM to allow firmware upgrades
(even remotely via telephone) without any hardware change. Intelligent
Alerting: The unit shall be configurable (as supplied with only the
addition of a modem) to provide automatic alerting when the unit is
triggered, the memory usage exceeds a preprogrammed level, the primary
battery voltage drops below a preprogrammed level or other
preprogrammed criteria are met. This alert shall be in the form of an
automatic telephone call via modem (initiated by the accelerograph) to
the user's host computer. The unit shall use a packet oriented
communications protocol which shall allow high priority packets
(alerting information) to be passed to the host computer without delay
even when event data is being transferred. GPS Timing: Each unit shall
include a GPS synchronized timing system (including antenna and 5 meter
cable) that will automatically maintain the unit's internal time
(including sample clock) accuracy to within 1 millisecond of absolute
time. Housing: Each unit shall be housed in a waterproof case with
three leveling feet and shall allow secure mounting by using only one
threaded mounting stud. All I/O connections (less GPS antenna) shall be
protected from the effects of ESD (electrostatic dishcarge) with
transient protection devices and from the effects of radiated and
conducted EMI (electromagnetic interference) and RFI (radio frequency
interference) vulnerability with ferrite filter elements. For
reliability reasons the unit shall contain no lithium batteries.
Support Software: Support software for unit set-up, parameter
modification, data retrieval, intelligent alerting, system diagnostics,
and data display shall be provided. All such support software shall be
Windows based and shall operate on IBM compatible personal computers.
It shall also be possible to operate the unit and retrieve data with
standard PC communications software. Based on information available at
this time, and in accordance with 41 u.s.c. 253(c)(1), implemented by
FAR 6.302-1, the Government considers the item specified as a sole
source from Kinemetrics Inc., Pasadena, California and proposes to
negotiate a contract with that firm on that basis. However, should
additional sources be identified, they will be considered. Firms who
feel that they can furnish the required item or its functional
equivalent (based upon the criteria specified herein) are invited to
sumit in writing an affirmative response to this announcement. An
affirmative response would include literature, brochures, and such
other materials which correspond to the required item stated herein.
This information will be evaluated and used to determine if competitive
opportunities exist. Since no request for quotation document exists,
requests for such documents without accompanying information will be
considered non-responsive to this request without further
consideration. This notice may represent the Government's only official
notice of this procurement. See Note 22. (0221) Loren Data Corp. http://www.ld.com (SYN# 0467 19950810\66-0022.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|