COMMERCE BUSINESS DAILY ISSUE OF MAY 24,1996 PSA#1602Department of Commerce, Mountain Administrative Support Center,
Procurement Division, MC3, 325 Broadway, Boulder, CO 80303-3328 66 -- PULSED NEODYMIUM-DOPED YAG LASER SYSTEM SOL 52RANB60C033 DUE
061796 POC Marilyn Scherer, 303-497-3657 The National Institute of
Standards and Technology intends to purchase a neodymium-doped YAG
laser system that will operate continuous wave, mode-locked, or
combined mode-locked and Q-switched. The specifications for this laser
system are given below. Note that ALL the laser performance
specifications assume that the laser is operating in the lowest order
TEM00 transverse spatial mode. 1. Continuous-wave (cw) and pulsed
operation at 1064 nm wavelength: a. In the cw operation state, the
minimum optical output power is 12 W. b. In mode-locked operation the
repetition rate must be between 80 MHz and 100 MHz. The output must be
at least 1.5 kW peak power with the pulses not exceeding 100 ps in
duration. The average output power must be at least 12 W. c. In
combined mode-locked and Q-switched operation the Q-switched envelope
of the mode-locked pulses must not exceed 300 ns. The Q- switched
repetition rate must be at least 400 Hz. The peak output power must be
at least 1 MW. The Q-switched pulse energy must be at least 2.0
millijoules. The pulse energy in the largest mode-locked pulse under
the Q-switched envelope must be at least 100 microjoules. The average
output power in combined mode-locked and Q-switched operation must be
at least 10 W. 2. Laser Stability, Beam Diameter and Point Stability:
The peak-to-peak amplitude fluctuation of the laser while operating in
mode-locked operation must not exceed one percent. Peak-to-peak
amplitude fluctuations of the laser while operating simultaneously
mode-locked and Q-switched must not exceed five percent. The beam
diameter of the of the laser output at 1064 nm must not exceed 0.6 mm.
The beam pointing variation of the laser, under any of its operating
conditions, must not exceed 80 microradians. 3. Physical Size of Laser
and Power Supply: The laser must require a space no more than 300 cm
in length and 30 cm in width. The power supply must be no higher than
60 cm, no wider than 100 cm, and no deeper than 80 cm. 4. Optional
Upgrades to Laser System for Operation at 1319 nm: The laser must have
the capability to be upgraded by the installation of optional
components that would permit it to operate cw, mode-locked, or
simultaneously mode-locked and Q-switched at a wavelength of 1319 nm.
For all of the specifications relating to the optional laser upgrades,
the laser is assumed to operate in the TEM00 lowest-order spatial
mode. a. In cw operation, the output power at 1319 nm must be at least
1 W. b. In mode-locked operation, the average output power must be at
least 1.5 W, the peak power must be at least 0.15 kW, and the pulse
width must be no greater than 120 ps. The mode-locked repetition rate
must be between 80 MHz and 100 MHz. c. In combined mode-locked and
Q-switched operation, the average power must be at least 1.0 W., the
peak power must be at least 50 kW, the duration of the Q-switched
envelope of the mode-locked pulses must not exceed 1300 ns. The width
of the central mode-locked pulse under the Q-switched envelope must not
exceed 150 ps. 5. Optional Upgrades to the Laser System for Operation
at 532 nm: The laser must have the capability to be upgraded by the
installation of optional components that would permit it to operate
mode-locked, or simultaneously mode-locked and Q-switched at a
wavelength of 532 nm via second-harmonic generation of the 1064 nm
fundamental. For all of the specifications relating to the optional
laser upgrades, the laser is assumed to operate in the TEM00
lowest-order spatial mode. a. In mode-locked operation, the average
power must be at least 1.0 W and the pulse width must not exceed 100
ps. Peak power output under mode-locked operation must be at least 0.2
kW. b. In combined mode-locked and Q-switched operation the average
power must be at least 1.0 W and the peak power must be at least 500
kW. The Q-switched envelope of the mode- locked pulses must not exceed
300 ns. The pulse width of the mode-locked pulses under the Q-
switched envelope must not exceed 100 ps. The energy of the largest
mode-locked pulse under the Q-switched envelope must be at least 50
microjoules. c. Peak-to-peak amplitude fluctuations of the laser in
mode-locked operation must not exceed two percent. Peak-to-peak
amplitude fluctuations of the laser in combined mode-locked and
Q-switched operation must not exceed eight percent. 6. Optional
Upgrades to Produce Ultraviolet Light: The laser must be upgradable at
some future time for the production of ultraviolet light through
third-harmonic generation and forth-harmonic generation of the
fundamental wavelength at 1064 nm. The Government intends to buy only
the laser system now, and will, at its discretion, include the upgrade
items as options to the contract for purchase within one year after
date of contract award. This synopsis is published for informational
purposes only; solicitation is not available; no telephone requests
will be accepted. Only qualification statements will be considered at
this time. In the absence of other qualified sources, it is the intent
of the Government to negotiate a sole source contract under the
authority of 41 USC 253 (c) (1) with Excel Quantronix, 45 Adams Avenue,
Hauppauge, NY 11788. See Note 22. (0143) Loren Data Corp. http://www.ld.com (SYN# 0409 19960523\66-0013.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|