|
COMMERCE BUSINESS DAILY ISSUE OF JUNE 17,1996 PSA#1617NASA Lewis Research Center, 21000 Brookpark Road, Cleveland, Ohio
44135 66 -- SCANNING ELECTRON MICROSCOPE SYSTEM - PART 2 SOL SFO3-051357 POC
Glen M. Williams, Contracting Officer, who may be contacted at (216)
433-2885, Mail Stop 500-306 I This is a combined synopsis/SFO for
commercial items prepared in accordance with FAR Subpart 12.6, as
supplemented with additional information included in this notice. This
announcement constitutes the only solicitation; offers are being
requested and a written solicitation will not be issued. This SFO and
incorporated provisions and clauses are those in effect through Federal
Acquisition Circular 90-37. The Standard Industrial Classification
(SIC) Code for this acquisition is 3826. The Small Business Size
Standard is 500 employees. Under the Defense Priorities and Allocations
System (DPAS)(CFR 15 Part 350), the rating assigned to this acquisition
is DO-C9. This procurement is being conducted under the NASA Midrange
Pilot Test Program approved by the Office of Federal Procurement Policy
on April 16, 1993. THE CONTINUATION OF THE SPECIFICATIONS FOR THE
SCANNING ELECTRON MICROSCOPE SYSTEM FROM THE SYNOPSIS PART 1 ARE AS
FOLLOWS: Item 1F. Imaging System 1. The imaging system shall provide a
magnification range from 25x to 1,100,000x at 200KV. Radial distortion
shall be less than one percent and spiral distortion shall be less than
1.5 percent over a 100 mm diameter on the plate. 2. Automatic lens
normalizing shall be provided to give a reproducibility of
magnification of less than 1.5 percent. 3. Automatic correction for
electron energy losses up to 1/1000th of the primary beam energy caused
by chromatic errors in thick samples shall be corrected up to 100 mm
diameter on the plate. 4. Microprocessor shall allow free lens control
of any or all lenses starting from the last used settings. Storage and
recall of a minimum of six configurations shall be possible. External
computer control of lens configurations shall also be provided. 5. Main
viewing screen shall be separate from the focusing screen. A binocular
microscope shall be provided for observing the image on the focusing
screen at a minimum of 12X magnification. Binoculars shall have high
point eye pieces for operators wearing glasses. 6. Microprocessor shall
provide a facility to accurately measure distances, heights, ratios,
and angles of all TEM images and diffraction patterns. 7. Software
shall be provided to allow an Apple Macintosh computer to allow storage
and recall of lens settings. Item 1G. Diffraction 1. Selected area
aperture holder in the diffraction lens shall accept a minimum of four
standard 3 mm apertures which can be selected by a click-stop
mechanism. 2. Dark field imaging shall be performed by tilting the beam
using either X-Y or conical controls. Tilt parameters shall be
displayed by the microprocessor . 3. Microprocessor control shall
provide a calibrated shift capability to determine lattice spacings and
crystal structures directly in the microscope. 4. STEM shall be able to
form convergent beam diffraction patterns of areas down to 2 nm
diameter with the specimen in the eucentric position. Convergence angle
shall be continuously variable to make the diffraction discs as large
as possible without overlapping. Item 1H. Film Camera 1. STEM shall
include a film camera capable of recording a minimum of fifty-five (55)
3 1/4'' x 4'' electron imaging films. 2. Camera control shall include
both automatic and manual exposure settings. Automatic exposure shall
be controlled by the electron intensity on the main screen or the
electron intensity on the focusing screen. 3. Microprocessor shall
provide fully automatic through-focus series of exposures by selection
of number of exposures and the focus increment. 4. Notation on the
negative shall include as a minimum; magnification, user code exposure
number, date, and at least 60 free text characters. 5. Film camera
shall in no way interfere with installation of the imaging filter (Item
2). 6. The contractor shall provide a minimum of three (3) sets of film
carriers, three (3) film supply boxes, and three (3) exposed film
boxes. Item 1I. Scanning System 1. Scanning optics shall produce a
stationary diffraction pattern in the objective lens back focal plane
under all conditions. 2. Shall be able to perform spot sizes less that
2 nm at the eucentric position. 3. Detectors shall be provided for
bright field, dark field, and back scattered electron imaging.
Detectors shall not alter in any way the specification of the basic
instrument. Back scattered electron detector shall not limit specimen
tilt range. 4. Continuous scan rotation shall be provided. 5. Dual
monitors for viewing images via different detectors simultaneously
shall be included. 6. A high resolution monitor shall be used for
photography using a Polaroid 4x5 camera. 7. Microprocessor control
shall provide accurate measurement of distances, heights, ratios, and
angles for all scanned images. 8. Selected area channeling patterns
shall be possible to be displayed or recorded while viewing the area of
the specimen from which it is obtained on the TEM screen. Beam rocking
up to twenty degrees shall be required for producing channeling
patterns. 9. Scanning system shall allow descanning of the scanning
beam so that each image point is transferred back onto the optical axis
and into the entrance slit of the imaging filter. Item 1J. X-ray Energy
Dispersive Spectroscopy Provisions 1. Design of the STEM shall provide
for an x-ray collection geometry with a minimum eighteen (18) degree
take off angle and 0.15 steradians solid angle on a thirty (30)
millimeter square detector crystal. 2. Microprocessor shall provide for
an x-ray detector protection function. 3. Scanning system shall allow
beam control by the x-ray energy dispersive system. Item 1K.
Miscellaneous 1. Contractor shall provide a mains matching transformer
capable of inputs from 190 to 440 volts and output of 220 volts.
Transformer shall include proper isolation and filtering to prevent
interference from accessories. 2. Contractor shall supply an air
compressor to meet the STEM's requirements. 3. Contractor shall provide
a radiation alarm monitoring system and sulfur hexafluoride detector.
4. Contractor shall provide a water cooled water chiller, film
dessicator with mechanical vacuum pump, and nitrogen gas regulator.
Item 2. The contractor shall provide an Electron Imaging Filter which
shall have the following features and capabilities; Item 2A. Design 1.
The Electron Imaging Filter shall be compatible with Item 1 and shall
be capable of operating up to and including 200 KV. 2. The Electron
Imaging Filter shall be of a post column design. 3. The Electron
Imaging Filter shall be able to form energy-filtered images when the
STEM is operated in image mode, energy-filtered diffraction patterns
when the STEM is operated in diffraction mode, and collection of
parallel electron energy-loss spectra. 4. The Electron Imaging Filter
shall correct for second-order aberrations in the energy loss spectrum.
5. The Electron Imaging Filter shall include an energy selecting slit
with an opening of at least 50 eV. The slit's opening and retraction
shall be pneumatically and computer controlled. 6. The Electron Imaging
Filter shall include a minimum of three (3) entrance apertures.
Selection of aperture shall be pneumatically actuated under computer
control. Apertures shall be pre-centered and require no manual
adjustment. 7. The Electron Imaging Filter shall consist of a minimum
of four quadrupole and five sextupole lenses to provide rotation free
operation when switching between energy dispersions. The range of
energy dispersions shall be equal to or better than 0.05 eV per channel
to 1.0 eV per channel. 8. The Electron Imaging Filter's detection
chamber shall include both a TV rate camera and a Grade A 1024 by 1024
pixel slow scan CCD. The TV rate camera shall include a digital
averaging device to improve the signal to noise during live viewing.
The slow scan CCD shall have a minimum dynamic range of 16000:1. Item
2B. Computer and Applications Software 1. The Electron Imaging Filter
System shall be include an Apple Power Macintosh computer with a
minimum of a 100 mHz processor, 48 Mb of memory, 1.4 Mb floppy drive,
1.2 Gb hard drive, removable 1.0 Gb drive, CD-ROM drive, network
capability, keyboard, track ball, and twenty-one (21) inch color
monitor. 2. Application software shall be included for image
acquisition and analysis. 3. Application software shall be included for
energy loss spectrum acquisition and quantification. Elemental mapping
of 512 x 512 pixels shall be included. Software for standard-less
quantification using built in cross-sections for all elements shall be
included. 4. Electron Imaging Filter System shall be able to read a
minimum of five elctron energy loss spectra per second. (0165) Loren Data Corp. http://www.ld.com (SYN# 0326 19960614\66-0001.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|