66 -- VIBRATION ISOLATION SYSTEM SOL 52SBNB6C9248 DUE 090396 POC Marsha Rodgers (301) 975-6398 The National Institute of Standards and Technology (NIST) has a requirement for a Vibration Isolation System including an optical quality vibrational isolation table and an active vibration isolation control system. The system shall consist of a custom optical quality non-magnetic table system and an active vibration isolation control system. The active control system shall provide isolation starting at 0.3 Hz or below, and provide for six degrees of motion control. The active control system shall meet or exceed vertical and horizontal isolation of 15db at 1 Hz, 20dB at 2 Hz, and 30dB at 5 Hz, and 40 DB at 100 Hz. -- F.O.B. Destination, Gaithersburg, Maryland.--Required Delivery Time 10 weeks- Evaluation criteria shall include the degree to which the proposal meets the specification, experience and capability, and delivery schedule and product quality. All responsible sources may submit a proposal which shall be considered by the agency. (0201)