COMMERCE BUSINESS DAILY ISSUE OF AUGUST 19,1996 PSA#1661NIH, National Heart, Lung, and Blood Institute, COB, Procurement
Section, 6701 Rockledge Drive MSC 7902, Bethesda, MD 20892-7902 66 -- C-ARM UPGRADE TO CARDIAC CATHETERIZATION SUITE SOL
NHLBI-HL-P-96-112 POC Kathleen J. Van Steelant, Contracting Officer,
(301) 435-0364. It is the intent of the National Institutes of Health
(NIH), National Heart, Lung, and Blood Institute (NHLBI) to purchase a
digital imaging system for assembly and installation in the Laboratory
of Animal Medicine and Surgery Cardiac Catheterization Suite. Following
are the minimum specifications: 1) VIDEO INPUT: a) 2 inputs: camera VTR
(or other device); b) Video line rate: 525,625 progressive. 1023, 1049
and 1249 are interlaced or progressive; c) Video input levels: 1 volt
peak to peak composite, adjustable from 0.7 to 1.3 volts; d) A/D
converter: 100 MHz 8 bit device; e) Sampling rate: Up to 50 MHz; and f)
Signal to noise: 50 dB; 2) VIDEO OUTPUT: a) 2 outputs: Live cine,
Fluoro, Loop Replay and Reference Images; b) Video line rate: 1023/1049
60Hz Interlaced, 1249 50Hz Interlaced; c) Video output levels: 1 volt
peak to peak composite, adjustable from 0.7 to 1.3 volts. Independent
look up table provided on each DCA output; d) Concurrent Low Lines Rate
Output: Monitor 1 output (live fluoro, cine and loop images) also
available as 525 line RSI 70 or 625 line CCIR output; 3) IMAGE STORAGE:
a) Disk Storage - Real time disk (Gbyte per drive, expandable to 2
drives). Specifically, the successful offeror shall: 1) Dismantle and
remove (dispose of) components not required for upgrade to the existing
system; 2) Modify the tube crane assembly to accept 2 lateral carriages
for support of monitor yoke radiation shield; 3) Provide and install
new Acoma Iso-Arc C-Arm; 4) Mount X-Ray tube on to new C-Arm; 5) Mount
new TV, Video, Cine components; 6) Install Digital Imaging System; 7)
Reload existing X-Ray tube into replacement housing; 8) Reposition AP
generator components in radiology suite; 9) Add Line Lock modification
to existing Cine camera; 10) Interface generator with digital system
and C-Arm; 11) Install new table; and 12) Calibrate system for on-line
use. Proposals will be evaluated on the following: 1) The offeror's
demonstrated ability to integrate partial or existing radiology
equipment with new system components; 2) The offeror's demonstrated
ability to integrate existing cine system to new digital video upgraded
system; 3) The upgraded system shall provide image digitization with
the ability for storage of images on video tape or laser disk; 4)
Real-time imaging capabilities suitable for cardiac catheterization
procedures; and 5) Completion of proposed upgrade within 120 calendar
days after contract award. A firm fixed-price type contract will be
negotiated. It is anticipated that Request for Proposals (RFP) Number
NHLBI-HL-P-96-112 will be issued on or about August 23, 1996. Copies of
the RFP may be obtained by writing to the National Institutes of
Health, National Heart, Lung, and Blood Institute, Contracts Operations
Branch, Procurement Section, Rockledge Building 2, 6701 Rockledge
Drive, Room 6143, MSC 7902 Bethesda, Maryland 20892-7902, Attention:
Kathleen J. Van Steelant. See Numbered Note(s): 1. (0228) Loren Data Corp. http://www.ld.com (SYN# 0270 19960816\66-0003.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|