COMMERCE BUSINESS DAILY ISSUE OF AUGUST 27,1996 PSA#1667Naval Research Laboratory, 4555 Overlook Ave., S.W., Washington, D.C.
20375-5326 66 -- INSTRUMENTS AND LABORATORY EQUIPMENT SOL N00014-96-R-LS08 POC
Lisa Bridgan, Contract Specialist, Code 3230.LS (202) 767-3739 Helen
Paul, Contracting Officer, Code 3230.HB (202) 767-3739. The
Specifications previously provided in the the combined
synopsis/solicitation for commercial items in the 15 August 1996 issue
of the Commerce Business Daily is hereby revised as follows: The
Contractor shall provide a high performance UV/VIS/NIR double beam
spectrometer with double monochrometer and reference beam attenuation,
1 each that meets or exceeds the following specifications: The optical
unit shall include a large interchangeable sample compartment and
detector module, and monochrometers with holographic gratings in
Littrow configuration with 1440 L/mm for the UV/VIS range and 360 L/mm
for the NIR range. The spectrometer must have a side on
photomultiplier with especially high sensitivity for the UV range and
a high performance thermostatted PbS cell for the NIR range with
automatic detector switching. The spectrometer must accurately measure
material properties in the following wavelength range: .185 - 3.3
microns (and be expandable to .175 microns with a nitrogen purge). The
instrument must have bandpass of .05 - to 5.0 nm in .1 nm increments
in the UV/VIS range and .2 - 20 nm in .04 nm increments in the NIR
range. The spectrometer must provide scan speeds of .1 to 2400 nm in
the UV/VIS range and up to 9600 nm/min. in the NIR range. Stray light
must not exceed .00008%T @ 220, 340, and 370 nm. measured according to
ASTM E-387. Noise levels must not exceed: .000050 A RMS @ 500 nm, 0A,
.000160 A RMS @ 500 nm, 1A, .000300 A RMS @ 500 nm,. 2A, 001000 A RMS
@ 500 nm, 4A, .001000 A RMS @ 190 nm, 0A. The spectrometer must
contain an optical system that includes a large interchangeable 150 mm
integrating sphere, a motorized sample and reference beam attenuator,
prealigned tungsten-halogen and deuterium lamps (with lifetime
counter) and automatic source change. The contractor shall provide
operative software for the spectrometer. The software must provide
recording, manipulation and storage of spectral data and complete
control of instruments and accessories. The synopsis/solicitation
closing date is hereby extended from 03 September 1996 to 05 September
1996. All of the other terms and conditions previously stated remain
unchanged. **** (0236) Loren Data Corp. http://www.ld.com (SYN# 0230 19960826\66-0003.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|