COMMERCE BUSINESS DAILY ISSUE OF OCTOBER 24,1996 PSA#1707VAMC (593) 1701 W Charleston Blvd.Rm 623, Las Vegas NV 89102 - FAX
702-386-3157 65 -- PROCUREMENT - VASCULAR DOPPLER 1 EA, - COMMERCIAL ITEM SOL
593-P01-97 DUE 111996 POC Kathleen Randall, Contracting Officer - FAX
ONLY FOR INFO This is a combined synopsis/solicitation for commercial
items prepared in accordance with Subpart 12.6, as supplemented with
additional information included in this notice. The solicitation number
593-P01-97 is issued as a Request For Proposal (RFP) subject to full
and open competition. Requirement for 1 each Noninvasive vascular
diagnostic system to include all probes, cuffs, clips, straps,
controls, connectors, tubes, cart, software, and start-up supplies with
minimum capabilities: Must offer A) Two (2) frequencies of
Bi-directional Doppler, 4MHz and 8MHz, user selectable for full
bi-directional & inverted bi-directional waveform recording, flow
toward probe recordings, or flow away from probe recordings. Doppler
recording to provide user selectable lo pass output filtering of 28 Hz,
14 Hz, 7 Hz, 3.5 Hz and mean flow. B) Bilateral (2channel) PPG and VPR,
AC & DC (Arterial & Venous) operation modes. Must provide calculation
of cuff volume in milliliters. C) Automatic Cuff Inflator with Blood
Pressure and VPR operation modes with BP having linear bleed down, user
select rate; and VPR having user selectable preset inflation levels -
to provide automatic BP cuff selection, provide hands free inflation
routing to cuff(s) appropriate for the site being tested. D) Must
provide for storage of a minimum of 20,000 full patient reports on
primary hard drive, allowing instant, on-line access to complete
patient reports- on non-dedicated PC or software capable for use on any
Pentium PC system with minimum requirements to run this program; allow
creation and storage of library of user created technical impressions,
retrievable into any patient report- to become a permanent part of the
patient record in the computer; must allow typed insertion of
comments, notes, clinical observations directly into the patient
report. Inserted notes must also be a permanent part of the computer
stored patient record. E) Must be equipped with a 28.8 modem to allow
remote software updates (updates to be included at no added cost) and
remote system diagnostics via telephone line. F) System to include
DATACHECK vascular accreditation software package to assist in the
compilation of accreditation/quality assurance statistics. G) system to
offer OPTIONAL VGA color display of diagnostic quality; Single channel,
40m thermal strip chart recorder; Pentium level microprocessor with
minimum 100 MHz, 845 MB HD, 1.44 floppy, 16MB RAM; and a color printer
for automatic report generation. These OPTIONAL items may be supplied
by the VA if deemed in the best interest of the VA to do so. The
system must also operate in either DOS or Windows 95 operating systems,
and allow use of commercially available IBM compatible type software.
On-site training initially & with each major software upgrade is to be
included. WARRANTY - a minimum 2 year full warranty is required. Your
offer shall include existing commercial product literature and
brochures on the item being offered that will meet the governments
needs. Mail to address listed above, no later than C.O.B. 11/19/96. The
provisions of FAR part 52.212- 1 Instruction to offers-commercial &
52.212-2 Evaluation commercial items, apply to this acquisition with
evaluation criteria to be Technical Capability #1 (includes specs,
upgrades, training & warranty information), Past Performance #2, and
Price #3. Offerors to include a completed copy of the FAR provision
52.212-3 with their offer. Also included in this solicitation are :
52.212-4 Contract Terms and Conditions- Commercial Items; 52.212-5
Contract Terms and Conditions Required to Implement Statutes or
Executive Orders - Commercial Items; 52.222-3 Convict Labor; 52.233-3
Protest After Award; 52.222-26 Equal Opportunity. Offers must be
received no later than C.O.B. Tuesday 11/19/96 4:00 PM PST. Fax
702-386-3157 - a copy of this CBD will be provided if needed, there
will be no other solicitation issued. SIC Code 5047. (0296) Loren Data Corp. http://www.ld.com (SYN# 0238 19961023\65-0001.SOL)
65 - Medical, Dental and Veterinary Equipment and Supplies Index Page
|
|