COMMERCE BUSINESS DAILY ISSUE OF JANUARY 23,1997 PSA#1767NASA, Procurement, John F. Kennedy Space Center, Kennedy Space Center,
FL 32899 66 -- ACQUISITION OF EDDY CURRENT TUBE TESTER POC Ginny Powers,
BOC-110, 407-867-7869 EG&G Florida, Inc., in support of NASA at the
Kennedy Space Center, is seeking sources to provide one (1) each Eddu
Current Tube Tester. The system supplied shall incorporate the
following features: 1) the eddy current unit must incorporate dual
frequency capabilities; 2) the system must be able to be used as a
conventional eddy current unit for locating surface and sub-sur-face
cracks and corrosion; 3) the system must have the capability to measure
conductivity and coating thickness; 4) the system must be able to
inspect copper tubing; 5) the system must be able to inspect carbon
steel tubing and pipes which incorporates a remote field eddy current
interface; 6) the system must be capable of adapting an automatic high
speed probe transport system for tub-ing; 7) necessary software and
hardware training shall be provided on site; 8) the system shall
consist of the following hardware and software: Tube sheet diagram
software compatible with Microsoft Windows 3.1xx; tubing data analysis
software compatible with Mocrosoft Windows 3.1xx; portable -- less
than 15 lbs; operat-ing frequency of 60Hx -- 6 Mhz; dual frequency
mode: 100 Hz -- 2 Mhz; Internal battery operated; Ac Inlet (100 -- 120V
or 200-240V at 50/60 Hx option); 6" minimum screen size; measure
coating thick-ness of 0-0.5"; measure conductivity range: 1% to 100%
IACS; analog outputs for both channels and mix; RS232 output; storage
capability of minimum 50 traces and 50 settings; Dual channel, Dual
probe, Dual frequency capabilities. Suppliers capable of providing a
system that meets these specifications are invited to send a data
package including sufficient technical data to establish a bonafide
capability of meeting the requirement, price sheets, GSA schedule (if
applicable) to EG&G Florida, Inc, attn: Ginny Powers, BOC-110, Kennedy
Space Center, FL 32899. A request for proposal will not be issued
until after any submitted data packages have been reviewed for
technical acceptance. Data packages should be received within 15 days
from date of publication. (0021) Loren Data Corp. http://www.ld.com (SYN# 0270 19970123\66-0013.SOL)