58 -- PORTABLE REAL TIME AEROSOL MONITORS SOL RFP 97-30 POC Bill Wise, Contracting Officer, 513/366-2025 The U.S. Environmental Protection Agency's Environmental Response Team Center in Edison, New Jersey has a requirement for portable real-time aerosol monitors, including ambient sampling inlet sets. This equipment is required for air monitoring and sampling of ambient particulates during emergency responses, Superfund site investigations, and for worker health and safety. The monitors must meet the following specifications: (1) must be handheld and transportable by commercial aircraft, (2) must have a concentration monitoring range from 0.1 ug/m3 to 400 mg/m3, (3) must have an adjustable flow rate, (4) must have data logging and alarm capabilities, (5) must have a built-in clock, (6) must have built-in calibration check capability, (7) must be able to measure on a real-time basis, (8) must have auto zero function which can be programmed by the operator, (9) must be battery operated with continuous operation capability greater than 12 hours, (10) must have a continuous digital or analog output, (11) must have capability to sample while monitoring, (12) must have the ability to measure total aerosols, PM 10 aerosols, and PM 2.5 aerosols, and (13) must be able to operate unattended. Requests for RFP 97-30 must be in writing to the above address or faxed to Bill Wise at 513/366-2004. Telephone requests will not be honored. (0057)