COMMERCE BUSINESS DAILY ISSUE OF MARCH 13,1997 PSA#1801National Institute of Standards & Technology, Acquisition & Assistance
Div., Bldg. 301, Rm B117, Gaithersburg, MD 20899 66 -- HIGH RESOLUTION GONIOMETER SYSTEM FOR X-RAY DIFFRACTION SOL
52SBNB7C1112 DUE 042597 POC Contract Specialist, Kathleen Lettofsky
(301)975-6342, Contracting Officer, Lisa K. Jandovitz (301)975-6342
WEB: NIST Contracts Homepage,
http://www.nist.gov/admin/od/contract/contract.htm. E-MAIL: NIST
Contracts Office, Contract@nist.gov. The National Institute of
Standards and Technology (NIST) has a requirement for the purchase of
a High Resolution Goniometer System (HRGS) for X-ray Diffraction. The
HRGS shall provide high resolution angular positioning of specimens for
x-ray diffraction studies. It shall include 5 rotation axes: coaxial
Theta and Two-Theta, Phi(Kappa) and Kappa which form a cradle of
"kappa" geometry, and a fifth axis, Phi(sample), perpendicular to
Phi(Kappa) and normal to the surface of a powder specimen, as seen in
figure 1 (available by fax). The Theta and Two-Theta rotation axes
shall be equipped with optical encoders and shall provide 0.2 arcsec
accuracy in angular positioning. The Kappa, Phi(kappa) and Phi(sample)
axes will allow the sample to be rotated about the Euler angles, Phi,
Chi and Omega with a sphere-of-confusion of less than 40 micrometers
in diameter. The controlling hardware and software shall be capable of
collecting data scans through arbitrary directions in reciprocal space
and slewing to within0.5 arcsec of target angle angles in Theta and
Two-Theta. Delivery required Freight on Board (FOB) destination,
Gaithersburg, MD. Delivery is required 120 days after contract award.
All responsible sources may submit a proposal which shall be
considered. Requests of copies of the solicitation may be telefaxed to
(301)963-7732. (0070) Loren Data Corp. http://www.ld.com (SYN# 0329 19970313\66-0009.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|