|
COMMERCE BUSINESS DAILY ISSUE OF APRIL 11,1997 PSA#1822National Institute of Standards & Technology, Acquisition & Assistance
Div., Bldg. 301, Rm B117, Gaithersburg, MD 20899 66 -- SCANNING PROBE MICROSCOPE CONTROL SYSTEM SOL 53SBNB760102 DUE
042497 POC Contracting Officer, Joan M. Smith -- (301) 975-6458,
Contract Specialist, Tamara Grinspon -- (301) 975-4390 WEB: NSIT
Contracts Homepage, http://www.nist.gov/admin/od/contract/contract.htm.
E-MAIL: NIST Contracts Office, Contract@nist.gov. The National
Institute of Standards and Technology (NIST) has a requirement for a
quantity of one (1) each RHK Technology SPM1000 and DTI-100 scanning
probe microscope control system and digital interface for pulse
counting, or equal. All parts, including computer boards, software,
electronics and hardware shall be completely interchangeable with
existing RHK SPM1000 instrumentation. In addition, the following
specifications shall be met. SPECIFICATIONS FOR SCANNING PROBE
MICROSCOPE CONTROL SYSTEM: Unit shall consist of scanning electronics,
hardware and software for driving a variety of scanning probe
microscopes, including but not limited to scanning near-field optical,
scanning tunnelling, and atomic force scanning microscopes. An analog
feedback loop for Z-position is required, as are at least 10
individual high voltage amplifier outputs for control of the scanning
head. These voltages shall be configurable by the user to drive a
number of commercially available or home made scanning heads, as
detailed below. A complete set of electronics schematics and block
diagrams, with component-level detail, shall be included. The system
shall be Intel Pentium based. HARDWARE REQUIREMENTS: Electronics for
driving a scanning probe microscope shall include the following: Piezo
driver outputs -- Separate drivers (hardware and software) for 5
scanning voltages (+X, -X, +Y, -Y and Z) and 5 offset voltages (+X, -X,
+Y, -Y, and Z), shall be provided for offsetting and scanning the probe
and sample. These voltages shall run between +130V and -130V. They
shall have a peak-to-peak noise of less than 2mV over a broad band
between DC and 100 kHz, and less than 0.5 mV peak-to-peak power line
harmonics. In order to accommodate dual or single tube designs, with 4
or 5 quadrant piezos, split tubes, tripods, or multiple tube designs,
it shall be possible to use X, Y, and Z separately; to sum Z into X
and Y; to sum X offset into X scan; to sum Y offset into Y scan; and to
sum Z offset into Z scan. Furthermore, it shall be possible to sum a
fraction of the X and Y scan signals into the Z signal in order to
compensate for sample slope. These various functions shall be easily
changeable through the use of switches, potentiometers, or jumpers.
Other outputs -- The following signals are provided to the computer and
are also accessible via a BNC connector: X scan (-10V to +10 V signal,
representing the -130 V to +130 piezo range); Y scan (-10V to +10V
signal, representing the -130V to +130 piezo range); Z position (2
outputs) (-10V to +10V signal, representing the -130 V to +130 piezo
range). Bias voltage (-10V to +10V). Tunnel current or feedback loop
input (2 outputs) (-10V to +10V signal representing the signal to the
feedback loop). In addition, outputs to the computer shall include all
signals required to perform the functions outlined under the software
specifications. Finally, two DAC outputs for computer control of
external equipment shall be provided, along with software drivers for
them. Front panel controls -- All systemparameters shall be accessible
from the front panel of the electronics. These include X and Y scan
range, position offset, slope compensation, feedback loop parameters,
bias voltage, tunnel current, scan speed and resolution. Available scan
speeds shall be from less than or equal to 1ms/line to 9 seconds/line,
with at least 9 settings available in each decade. Scan ranges shall
be symmetric around zero, amplitudes from 0 to 130 V; values to be set
by user with a resolution and readout of 0.1V. Position offset shall
be values to be set by user with a resolution and readout of 0.1V
between -130V and +130V. Feedback loop time constants shall be from
approximately 0.1 ms to 0.1s. Gain on feedback adjustable loop shall be
from zero to at least 2. Feedback amplifier shall be linear or
logarithmic, switchable. Lines per frame to be set by user in at least
the range from 128 to 1024. Square frames OK. Set point voltage
(tunnel current) shall be set by user over a range from 0 to 1 or 0 to
10V (0.1 nA to 10 nA or 0.1to 100). STM bias voltage shall be set by
user over a range from -10V to 10 V. Feedback loop -- An analog
feedback loop with proportional and integral control shall be provided
to control the Z position signal. The input to this loop shall be
accessible on a BNC connector and a voltage between 0 and 10V or 0 and
-10V, switchable. Gain adjustable shall be between 0 and 2 and time
constant adjustable from approximately 0.1 ms to 0.1s. This loop must
be capable of controlling Z position for a number of different
applications, and in particular, lateral force AFM signals, non-contact
AFM, and STM applications shall be supported. Data inputs -- At least
6 independent data channels with the ability to simultaneously acquire
data on at least three channels is required. One of these channels
shall be dedicated to reading the Z position. Another shall be
dedicated to reading the input of the feedback loop described above. 3
other analog inputs shall be available, and all analog inputs shall be
interfaced to A to D converters with at least 12-bit resolution. A gain
of 1 to 8 shall be adjustable through the software provided, and an
additional gain of 1 to at least 128 (which shall be located on the
front panel) shall be available for the Z-position input channel. The
final channel shall be a TTL pulse counting input with at least 15 bit
resolution, upgradeable to 32 bit. All shall be accessible with BNC
connectors. All external analog inputs shall be buffered through
differential inputs. External control and inputs -- The following BNC
inputs (approx. 0-10Volts,analog) shall be provided, each DC coupled,
separately and differentially buffered, and internally summed into the
appropriate output: Two bias voltage control inputs (summed into bias
voltage, unity gain); two Z modulation inputs (summed into Z scan,
unity gain); one Z position input (summed into Z offset, gain of 13);
and one X modulation input (summed into X scan, unity gain). In
addition, X scan and Y scan inputs shall be provided. These inputs
shall be analog inputs that connect directly to the inputs of the
high-voltage amplifiers that control the X and Y scans. It shall be
possible to disable internal scanning in order to use these inputs. An
"inertial approach" input shall be available to sum a sawtooth into
the Z offset piezo voltage in order to accommodate inertial approach
style scan heads. Software -- Shall provide an interactive environment
for all aspects of SPM control and data analysis including data
acquisition, image processing, display and output. The software shall
operate on an Intel Pentium running DOS 3.3 or later and be compatible
with Windows 95. The program shall be menu-driven and mouse
controlled. Scanning Probe Microscope (SPM) control -- The following
shall be controllable through software: Direction of the raster scan --
User must be able to stop a scan at any point; the dwell time at any
point during a scan shall be adjustable from 0s to at least 10 minutes
with 1 microsecond resolution and accuracy. Bias voltage shall be
adjustable by computer control during a scan. Two D to A outputs shall
be fully programmable through the software. In particular, the program
shall be capable of generating sawtooth waveforms useful in the control
of an inertial approach. Sample and hold circuitry shall be
software-controlled. SPM configuration -- All system parameters shall
be configurable through the software, meaning that the gain of inputs
and the calibration factors shall be accessible in software. In
addition, software for SPM calibration shall be provided. An
oscilloscope and spectrum analyzer shall be built into the software.
Data acquisition -- Real time images shall be acquired from both, or
just one, scan directions. Real time display of selected image(s) shall
be possible. It shall be possible to acquire data from 3 input channels
simultaneously, and to display any one of the resulting 6 images during
the scan. Data shall be stored at the same time as it is acquired. It
shall be possible to stop the scan at every point and acquire data at
that point for a given period of time, up to at least 10 minutes. STM
Spectroscopy options shall include number of points/curve, number of
sweeps/point, delay time before and after IV acquisition, and sample
and hold on/off select. Data analysis and image processing -- The
software shall provide at least the ability to correct for thermal
drifts, the ability to plot a cross section of the data, the ability to
do statistics on a plot, to difference two plots, draw histograms, do
2-dimensional FFTs, callculate correlations, X,Y, averages and X,Y,
slopes. Image processing routines shall include parabolic fits and
subtractions, X,Y offset or slope subtractions, background plane
subtractions on complete or partial images, smoothing, sharpen, expand,
filter (apply FFT or a user-programmed filter). This package shall be
capable of displaying images in 1280 by 1024 bit resolution and 256
color resolution. Bitmap, solid surface, or mesh display shall be
possible. Interactive color contrast, 3D tilt, rotate, and shade
functions are required. Output shall be possible in at least color
postscript and ascii text format, and it shall be possible to acquire
the full resolution image output of a screen capture, window capture,
or user-defined area in Bitmap. THIS IS A COMBINED
SYNOPSIS/SOLICITATION FOR COMMERCIAL ITEMS PREPARED IN ACCORDANCE WITH
THE FORMAT IN FAR SUBPART 12.6, AS SUPPLEMENTED WITH ADDITIONAL
INFORMATION INCLUDED IN THIS NOTICE. THIS ANNOUNCEMENT CONSTITUTES THE
ONLY SOLICITATION ; QUOTATIONS ARE BEING REQUESTED, AND A WRITTEN
SOLICITATION DOCUMENT WILL NOT BE ISSUED. This solicitation,
#53SBNB760102, is a Request for Quotation (RFQ). The solicitation
document and incorporated provisions and clauses are those in effect
through Federal Acquisition Circular 45. The associated Standards
Industrial Classification Code (SIC) for this procurement is 3826, and
the small business size is 500 Employees. The FOB Point shall be
Destination, Gaithersburg, Maryland. Delivery is required within 10-12
weeks from the Date of Award. Delivery location is the National
Institute of Standards and Technology (NIST), Building 301, Receiving
Room, Gaithersburg, Maryland 20899-0001. Acceptance location will be
NIST. The following FAR provisions apply to this solicitation:
52.212-1, Instructions to Offerors-Commercial; All vendors are to
include with their quotes, a completed copy of provision 52.212-3,
Offeror Representations and Certifications-Commercial Items. The
following FAR clause applies to this acquisition: 52.212-4, Contract
Terms and Conditions-Commercial Items. The following additional FAR
clauses, which are cited in clause 52.212-5(a) are applicable to this
acquisition and (b) as follows: 52.222-26, Equal Opportunity;
52.222-35, Affirmative Action for Special Disabled and Vietnam Era
Veterans; 52.222-36, Affirmative Action for Handicapped Workers;
52.222-37, Employment Reports on Special Disabled Veterans and Veterans
of the Vietnam Era; 52.225-3, Buy American Act-Supplies; 52.225-21, Buy
American Act, NorthAmerican Free Trade Agreement Implementation
Act-Balance of Payments Program. All quotes should be sent to the
National Institute of Standards and Technology, Acquisition and
Assistance Division, Building 301, Room B117, Gaithersburg, Maryland
20899-0001. Copies of above-referenced clauses are available upon
request, either by telephone or fax. All responsible small business
firms may submit a quote by 3:00 PM, Washington, DC time, on April 24,
1997. See Note 1. (0099) Loren Data Corp. http://www.ld.com (SYN# 0316 19970411\66-0010.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|