COMMERCE BUSINESS DAILY ISSUE OF JUNE 9,1997 PSA#1862DHHS/PHS/FDA/Office Of Facilities, Acquisition, And Central Services,
Division of Contracts and Program Management, HFA-512(COB), 5600
Fishers Lane, Rockville, MD 20857 66 -- LAB EQUIPMENT SOL E27814 DUE 062097 POC Trudy Dartouzost,
Contract Specialist, ph: 301/443-2597, fax: 301/443-3651 The Food and
Drug Administration (FDA), intends to award a purchase order for 17
X-ray monitors equivalent to the RADCAL Corporation Model 1015FBH with
ionization chambers 10X5-6L and 10X5-6M. The monitors must be able to
measure exposure, exposure rate, exposure integrated over a single
pulse of radiation of 1 ms or greater, and the duration of the pulse.
The monitors shall be equipped with one ionization chamber which shall
be compatible with the Center for Devices and Radiological Health test
stand and shall have an active volume not greater than 8 cc. This
chamber shall have a measurement range of 1 mR/min to 650 mR/min and
energy dependence less than 7% from M30 (HVL = .36 mm AL HC = 14) to
the H150 (HVL -- 17.0 HC = 100). A second chamber shall have the same
nominal volume but be designed for the measurement of mammographic
x-ray beams. This chamber shall have less than 2% energy dependence in
x-ray beams from molybdenum anodes for added filtrations of either 30
micrometers of molybdenum or 30 micrometers of molybdenum plus 2 mm
aluminum. The calibration accuracy for both chambers shall be 5% or
better. The monitors and chambers shall be contained within a case no
larger than 8" by 10" by 12" and not to exceed 15 pounds. The monitors
shall have a digital display and shall be powered by alkaline "D" cell
batteries. The batteries shall have a lifetime for 750 hours of
operation. Written quotations required by COB 6/20/97. (0156) Loren Data Corp. http://www.ld.com (SYN# 0303 19970609\66-0010.SOL)