|
COMMERCE BUSINESS DAILY ISSUE OF JUNE 30,1997 PSA#1877National Institutes of Health, National Heart, Lung, and Blood
Institute, Contracts Operations Branch, Procurement Section, 6701
Rockledge Drive, Room 6143, Bethesda, Maryland 20892-7902 66 -- CONSOLE UPGRADES TO NUCLEAR MAGNETIC RESONANCE SPECTROMETERS SOL
NHLBI-HL-P-97-111 DUE 081197 POC Kathleen J. Van Steelant, Contracting
Officer, (301) 435-0364 It is the intent of the National Heart, Lung,
and Blood Institute e (NHLBI), Division of Intramural Research,
Laboratory of Biophysical Chemistry, to acquire console upgrades to
Nuclear Magnetic Resonance Spectrometers. The upgrades must meet the
following minimum specifications: 600 MHZ: 1) The upgraded console
shall be capable of operating at or above 600 MHZ 1H (proton) frequency
and in the range of at least 30 -- 243 MHZ (with 15 -- 320 MHZ being
preferable) for X nuclei up to and including 31P. Thi s upgraded
console shall be interfaced with the existing Oxford magnet, upper and
lower shim stacks, and existing probes and related hardware or the
necessary modifications, replacements or substitutions made (at the
discretion and approval of the project officer) to this existing
equipment in use such that the working spectrometer (console, probes,
interfaces, magnet, computers and related hardware devices) is
identical in all respects (except for age of hardware not actually
replaced) to spectrometers cur rently being marketed by the
manufacturer. The Contractor shall take full responsibility for the
entire system, including probe heads, shim system and superconducting
magnet; 2) The upgraded console shall be equipped with and use a
minimum of 28 room temperature shim gradients optimally designed for
the highest magnetic field strength homogeneity commensurate with
minimal heating. The upgraded console shall have the capability to
perform computer controlled automated shimming of the magnetic field.
Adjus tment of the superconducting magnetic field strength or shims
required to obtain optimal room temperature specifications and shim
performance shall be the responsibility of the Contractor and shall be
carried out after consultation and in accord with the project officer;
3) The upgraded console shall have the capability to perform inverse
detection NMR experiments for X frequencies in the range specified in
1 above; 4) The upgrade shall include a power supply capable of
generating simultaneously and indepen dently power to drive three
orthogonal gradients in three axis gradient probes. The shape,
amplitude, duration and polarity of these gradients shall be under
pulse program control and the use of various waveforms including sine
shall be possible. Gradient amplifiers shall employ eddy current
compensation and be linear, low noise, and preferably constant current.
This system shall be capable of generating a field gradient along the
z axis (axis of the magnetic field) with a strength of at least 25 G/cm
on the 5-mm probe head(see next item). Gradients in the x and y
dimensions of the three-axis probe shall be at least 25 G/cm. The
vendor shall describe in detail in their proposal their strategy and
procedure, including automation when applicable, for performing
gradient shimming (magnetic field mapping). The vendor shall optimize
the homogeneity of the region sampled by the 1H coil in e (below)
including mapping of the field using gradient techniques on the
existing 600 MHZ Oxford magnet in the buyer's la boratory; 5) The
system shall include a triple resonance 5-mm three-axis pulsed field
gradient probe for observation of 1H while decoupling 13C and 15N,
include 2H lock, and the 90o pulse widths must be < 10 5s for 1H for
an aqueous non ionic sample, < 15 5s ( 12 5s is preferred) for 13C
for an aqueous non ionic sample, and < 44 5s ( 35 5s is preferred)
for 15N for an aqueous non ionic sample. The probe head shall be
capable of withstanding the required power without arcing for 220 5s.
In addition, the 1 H RF homogeneity of the triple resonance probe head
must be such that a continuous 810o pulse (2 X 360o + 90o) must retain
at least 50% of the integrated magnetization relative to a single 90o
pulse, using an aqueous solution containing 100 mM NaCl; 6) Disruption
of 2H lock operation shall be minimized during application of gradient
pulses. The peak-to-peak amplitude for the largest signal in the
difference spectrum calculated from two one-transient spectra, one
recorded 200 microseconds after a25G/cm 1-m s square gradient pulse and
one recorded without the gradient pulse, shall be less than 10% from
the corresponding peak in the spectrum. The phase difference between
the two one-transient spectra shall be minimal; 7) The system shall be
equipped with a variable temperature control unit capable of providing
less that 0.02o C sample temperature variation per 1o C room
temperature change. The sample temperature control, using the pulsed
field gradient triple resonance 5-mm probe head must operate over a
rang e of -10o C to 60o C (100o C is preferred) with setting and
control being independent of room temperature; 8) The 1H sensitivity
specifications for the triple resonance triple-axis pulsed field
gradient 5 mm probe using Wilmad-535 sample tubes shall be: (I) ASTM
0.1% ethylbenzene 700:1; (ii) 0.5 mg sucrose/ml D2O 80:1; (iii) 0.5 mg
sucrose/ml in 250 mM KCl in D2O 67:1; 9) The upgraded console shall
provide at least four frequency channels (with five being preferred)
with direct frequency synthesis on all channels together with the
capability of producing different asynchronous or synchronous composite
pulse decoupling and different shaped pulses on all of these channels
simultaneously and independently. The system shall be completely
broadband and ca pable of observing all nuclei up to 1H (protons). The
composite pulse decoupling schemes shall be freely programmable by the
user and include the option for pulse shaping of the individual
elements of the composite pulse. All four (or five) channels shall have
identical specifications at low rf power levels, except that only two
channels shall require operation over the frequency range from 555 to
605 MHZ (for 19F and 1H), and a minimum of three channels shall operate
over the range of frequencies from 30 -- 243 MHZ (with 15 -- 320 MHZ
being preferable) for all other (i.e., X) nuclei; 10) The console
electronics shall be capable of 0.25 degree phase resolution, 0.1 Hz
frequency resolution (with 0.01 Hz being preferable) and a 90 dB
attenuation range on all four (or five) channels. The rf phase shall
vary by no more than 10 degrees over an attenuation range of 40 dB from
full power. The settling time after a phase shift during a pulse
sequence shall be minimal; 11) The pulse programming system shall
permit i mplementation of user-developed pulse sequences and shall be
capable of conducting all experiments published in the volumes 1 -- 3
of the Journal of Biomolecular NMR. It shall be possible to specify at
least 30 independent pulse sequence interval lengths and 30 pulse
widths in a single pulse program. The minimum interval length shall be
no greater than 500 ns. It shall be possible to specify each interval
length in increments of 25 ns or less. It shall be possible to write
and execute pulse programs with at least 256 steps (without loops),
with each step representing a time period during which any combination
of pulses on the four channels is given. The amplitude and phases of
the pulses shall be arbitrarily programmable, subject only to the
resolution limits on the phase shifters and rf attenuators. It shall be
possible to include at least twenty loops within a single pulse
program, with independent loop counters. It shall be possible to nest
loops; 12) The long term pulse amplitude stability on all ch annels
shall be 1% and phase stability shall be better than 1 degree, as
measured over a 24 hour period in a room where the temperature changes
by less than 1.5o C. This phase stability is measured from the
relative difference between 1D spectra, recorded with a 30 degree flip
angle pulse, at various times during a 24 hour period; 13) The console
electronics shall permit application of user defined shaped pulses to
operate independently on each of the four (or five) available channels.
The shaped pulses shall allow a minimum of 16000 user-definable
elements within a single pulse shape. Waveform memory shall be at least
64K ram with 256K being preferred. The dynamic range of each channel
for shaped pulses shall be at least 60 dB total range with 0.1
dBresolution at half range; 14) The system shall be capable of
independent and simultaneous control over amplitude, frequency, phase
and duration of pulses for all four (or five) frequency channels and
over internal real-time clock pulses for triggering of e xternal
devices; 15) Before high-power amplification, pulse rise and fall times
shall be < 300 ns, measured between 10% and 90% amplitude levels.
Fast power switching ( 4 us) shall be provided on all four (or five)
channels; 16) The RF amplifiers shall be 100 Watt linear for 1H and 19F
transmit pulses (at 550 -- 600 MHZ) and 300 Watts linear for all other
transmit pulse amplifiers up to at least 243 MHZ for X nuclei. The 1H
amplifier shall be capable of operating at 10 Watt CW and the X
nucleus amplifier shall be capable of operating at 30 Watt CW.
Operation of all transmit pulse amplifiers shall be under computer
control. The RF amplifiers shall be capable of performing spin lock
experiments on the 5-mm triple resonance gradient probe for up to 200
ms for 1H(gB1 = 12kHz) and up to 50 ms for the other channels (gB1 =
12kHz at 175 MHZ; gB1 = 4kHz at 60 MHZ) with droop 2%; 17) The upgraded
console shall be equipped with a digitizer of at least 16 bits allowing
simultaneous sampling of the quadrature receiv er channels using two
digitizers and two sample and hold amplifiers for spectral widths of up
to at least 150 kHz (200 kHz is preferred). The system shall be capable
of oversampling (at least 2X) and should be capable of accepted
industry standard digital and analog filtering of the NMR audio signal;
18) The console shall be equipped with a minimum 1280X1024 17 inch (19
inch is preferable) high resolution color monitor; 19) A minimum of 16
additional TTL lines that can be addressed by the pulse program for
external device control shall be provided; 20) The dynamic range of the
upgrade console shall be at least 60,000:1 as measured for a 90o 1H
pulse by a S/N of at least 60:1 on a t-butanol peak in a 1/10,000 1H
molar ratio to water; 21) The console shallbe equipped with ethernet
TCP/IP capability. The host computer system shall operate under the
UNIX operating system and provide standard UNIX utilities and a C
compiler. The host computer shall contain a minimum of 128 M-byte of
memory. The system shall i nclude a laser printer with software
packages for plotting spectra. The system shall contain at least two
serial ports, a keyboard and an industry standard mouse. The system
shall include as a minimum a 6X CDROM drive. The vendor may propose in
addition a quarter inch tape drive with 150 M-byte per cartridge
(unformatted); 22) The system shall be equipped with two fast access
disk systems, one containing at least 1G-byte of storage (unformatted)
and one containing at least 4 G-byte of storage (unformatte d). Upon
consultation with the vendor, one or more of the mass storage devices
may be purchased separately by NHLBI and installed by the vendor who
will assume responsibility for the entire system including the mass
storage devices. The system shall includea 4 G-byte DAT tape drive; 23)
The system shall be equipped with a SCSI interface (two are preferred)
for control of up to at least 8 devices per SCSI controller; 24) The
system shall have the capability of processing and analyzing data from
a previous experiment while simultaneously acquiring data for a new
experiment. This statement shall hold true for both one- and
multi-dimensional experiments, up to four dimensions; 25) The
transverse excitation profile for a G3 inversion pulse (Emsley and
Bodenhausen, Chem. Phys. Lett., 165, 4 69 (1990)) of 3 ms duration
shall be within a factor of two from the simulated value over a
bandwidth of 1 4kHz; 26) The upgraded system including probes supplied
new shall be covered under full hardware and software warranty for a
period of a minimum of one year after all guaranteed specifications
resulting from this contract have been met in the buyer's laboratory;
27) The vendor shall take full responsibility for the entire system,
including console, computer and supplied peripherals, shim system and
sup erconducting magnet. The vendor shall also warrant that all
applicable specifications quoted in their published documentation will
be met in the buyer's laboratory; 28) The acquired data shall be
available as a file with equally spaced data points for ready off-line
processing with standard software packages; 29) The peak-to-peak
difference spectrum calculated from two spectra (four scans each)
recorded under identical conditions, 30 minutes apart without removing
the sample, shall be less than 2% of each individual spectrum; 30) The
peak-to-peak spin echo difference signal for the anomeric proton in
sucrose (two scans) shall be less than twice the intensity of the
satellites for 10 consecutive experiments; 31) There shall be no
sidebands or other spurious artifacts in the spectrum larger than 10%
of the anomeric proton's satellite intensities. The signal-to-noise
shall be larger than 20:1 for a provided sample containing 50 mg
sucrose in 0.5 ml D2O. The spectra are to be recordedat 30o C unless
otherwise indicated. 360 MHZ: 1) The upgraded console shall be capable
of operating at or above 360 MHZ 1H (proton) frequency and in the
range of at least 20 -- 200 MHZ (with 10 -- 243 MHZ being preferable)
for X nuclei up to and including 31P. This upgraded console shall be
interfaced with the existing 89 mm clear bore (wide bore) Oxford
magnet, upper and lower shim stacks, and existing probes and related
hardware or the necessary modifications, replacements or substitutions
made (at the discretion and approval of the project o fficer) to this
existing equipment in use such that the working spectrometer (console,
probes, interfaces, magnet, computers and related hardware devices) is
identical in all respects (except for age of hardware not actually
replaced) to spectrometers currently being marketed by the
manufacturer. The Contractor shall take full responsibility for the
entire system, including probe heads, shim system and superconducting
magnet. Note: Since 360 MHZ is not a widely available frequency,
offerors may quote spe cifications for 300 MHZ and provide data at 300
MHZ although the delivered console shall be capable of observing
proton resonances at 360 MHZ and using a deuterium lock; 2) The
upgraded console shall be equipped with and use a minimum of 17 room
temperature shim gradients for a wide bore magnet with a clear bore of
89mm designed for the highest magnetic field strength homogeneity. The
upgraded console shall have the capability to perform computer
controlled automated shimming of the magnetic field. Adjust ment of the
superconducting magnetic field strength or shims required to obtain
optimal room temperature specifications and shim performance shall be
the responsibility of the Contractor and shall be carried out after
consultation and in accord with the project officer; 3) The upgraded
console shall have the capability to perform inverse detection NMR
experiments for X frequencies in the range specified in 1 above; 4) The
upgraded console shall include a power supply capable of generating
simultaneously and independently power to drive three orthogonal
gradients in three axis gradient probes. The shape, amplitude, duration
and polarity of these gradients shall be under pulse program control
and the use of various waveforms including sine shall be possible.
Gradient amplifiers shall employ eddy current compensation and be
linear, low noise, and preferably constant current. This system shall
be capable of generating a field gradient along the z axis (axis of the
magnetic field) with a strength of at least 25 G/cm on the 5-mm probe
head (see next item). Gradients in the x and y dimensions of the
three-axis probe shall be at least 25 G/cm. The offeror shall describe
in detail in their proposal their strategy and procedure, including
automation when applicable, for performing gradient shimming (magnetic
field mapping). The offeror shall optimize the homogeneity of the
region sampled by the 1H coil in e (below) including mapping of the
field using gradient techniques on the existing 89 mm bore 360 MHZ
Oxford ma gnet in the buyer's laboratory; 5) The system shall include
a triple resonance 5-mm three-axis pulsed field gradient probe for
observation of 1H while decoupling 13C and 15N, include 2H lock, and
the 90o pulse widths must be < 10 5s for 1H for an aqueous non ionic
sample, < 15 5s ( 12 5s is preferred) for 13C for an aqueous non
ionic sample, and < 44 5s ( 35 5s is preferred) for 15N for an
aqueous non ionic sample. The probe head shall be capable of
withstanding the required power without arcing for 220 5 s. In
addition, the 1H RF homogeneity of the triple resonance probe head
shall be such that a continuous 810o pulse (2 X 360o + 90o) must retain
at least 50% of the integrated magnetization relative to a single 90o
pulse, using an aqueous solution containing 100 mM NaCl; 6) Disruption
of 2H lock operation shall be minimized during application of gradient
pulses. The peak-to-peak amplitude for the largest signal in the
difference spectrum calculated from two one-transient spectra, one
recorded 200 microseconds after a 25G/cm 1-ms square gradient pulse and
one recorded without the gradient pulse, shall be less than 10% from
the corresponding peak in the spectrum. The phase difference between
the two one-transient spectra shall be minimal; 7) The system shall be
equipped w ith a variable temperature control unit capable of
providing less that 0.02o C sample temperature variation per 1o C room
temperature change. The sample temperature control, using the pulsed
field gradient triple resonance 5-mm probe head shall operate over a
range of -10o C to 60o C (100o C is preferred) with setting and control
being independent of room temperature; 8) The 1H sensitivity
specifications for the triple resonance triple-axis pulsed field
gradient 5 mm probe using Wilmad-535 sample tubes should be: (I) ASTM
0.1% ethylbenzene 300:1; (ii) 0.5 mg sucrose/ml D2O 30:1; (iii) 0.5 mg
sucrose/ml in 250 mM KCl in D2O 20:1; 9) The upgraded console shall
provide at least three frequency channels with direct frequency
synthesis on all channels together with the capability of producing
different asynchronous or synchronous composite pulse decoupling and
different shaped pulses on all of these channels simultaneously and
independently. The system shall be completely broadband and capable of
observing all nuclei up to 1H (protons). The composite pulse
decoupling schemes shall be freely programmable by the user and include
the option for pulse shaping of the individual elements of the
composite pulse. All three channels shall have identical specifications
at low rf power levels, except that only two channels shall require
operation over the frequency range from 333 to 365 MHZ (for 19F and
1H), and a minimum of three channels shall operate over the range of
frequencies from 20 -- 200 MHZ (with 10 -- 243 MHZ being preferable)
for all other (i.e., X) n uclei; 10) The console electronics shall be
capable of 0.25 degree phase resolution, 0.1 Hz frequency resolution
(with 0.01 Hz being preferable) and a 90 dB attenuation range on all
four (or five) channels. The rf phase shall vary by no more than 10
degrees over an attenuation range of 40 dB from full power. The
settling time after a phase shift during a pulse sequence shall be
minimal; 11) The pulse programming system shall permit implementation
of user-developed pulse sequences and shall be capable of c onducting
all experiments published in the volumes 1 -- 3 of the Journal of
Biomolecular NMR. It shall be possible to specify at least 30
independent pulse sequence inte (0176) Loren Data Corp. http://www.ld.com (SYN# 0343 19970630\66-0001.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|