COMMERCE BUSINESS DAILY ISSUE OF JULY 17,1997 PSA#1889DHHS, PHS, CDC, NIOSH, ALOSH, 1095 Willowdale Rd., Morgantown, WV
26505-2888 66 -- AMENDMENT 001A TO CUSTOM TUNNEL, AUTOMATIC CAGE AND UTENSIL
WASHING SYSTEM SOL 97-019 DUE 072497 POC Brenda L. Goodwin, (304)
285-5882/Rex S. Wolfe, Jr., (304) 285-5883 THIS AMENDMENT IS A COMBINED
SYNOPSIS/SOLICITATION FOR COMMERCIAL ITEMS PREPARED IN ACCORDANCE WITH
THE FORMAT IN SUBPART 12.6, AS SUPPLEMENTED WITH ADDITIONAL
INFORMATION INCLUDED IN THIS NOTICE. THIS ANNOUNCEMENT CONSTITUTES THE
ONLY SOLICITATION; PROPOSALS ARE BEING REQUESTED AND A WRITTEN
SOLICITATION WILL NOT BE ISSUED. This notice represents AMENDMENT 001A
to Request for Proposal No. (RFP) 97-019, FURNISH AND INSTALL A CUSTOM
TUNNEL, CAGE AND UTENSIL WASHER published in the Commerce Business
Daily on July 1, 1997, PSA #1878. Amendment 001 contained an electronic
file conversion error each time degrees Fahrenheit was specified.
Amendment 001 published on July 16, 1997, is hereby deleted in its
entirety and replaced with the following Amendment 001A. As a result of
questions received the following additions and clarifications are
incorporated into the Specification: (1) Installation of the tunnel
washer shall include final utility connections; (2) Roughing drawings
and equipment descriptions shall be submitted by the offeror within one
week of acceptance of the bid; (3) The supply and exhaust requirements
for the tunnel washer shall be submitted by the successful offeror to
NIOSH within one week of notification of acceptance of the bid; (4)
The drain discharge cooling system will be provided by NIOSH, unless it
is a standard feature of the unit offered; (5) A high pressure steam
line (50 Psi), room air exhaust, and central acid neutralization will
be provided by NIOSH; (6) The wash and recirculated rinse temperature
should be able to be adjusted by the operator to 190 F. The system
should be able to monitor and guarantee specific temperatures up to 190
F so that the conveyor belt stops until a temperature set by the
operator has been achieved. Under normal operating conditions, stops to
achieved desired temperatures will not slow the operation of the tunnel
washer by more than 5% in an 8 hour work day; (7) The prewash section,
wash section, recirculated rinse section, and non-recirculated final
rinse section shall each result in a spray sufficient to thoroughly
cover each cage being washed in all positions across the belt so that
at the end of the process each cage is visibly thoroughly clean, free
of visible soap residue, and culturable bacterial numbers are reduced
by a minimum of 99% (determined by counts from swabs taken from cages
before and after the washing process).; (8) Jet spray headers shall be
provided and shall be secure but easy to remove without tools; (9) The
tunnel washing system shall operate so that light plastic cages are
hydraulically held onto the conveyor belt; (10) The smoothness of the
interior of the chamber and tanks will be assessed by sight, touch, and
performance (preventing the build up of debris); (11) The booster
heater with instantaneous steam to water heat exchange shall raise the
temperature of the NIOSH tap water supply by a minimum of 80 F; (12)
The statement in the original specifications "The supply filter
assembly will be complete with a periodic automatic reversingflush"
refers to a self cleaning filter to strain debris provided by the wash
pump to a size that will prevent the plugging of jet sprayers. This
filter assembly shall be easily accessible by the operator; (13) The
statement in the original specifications "the washer interior will be
automatically cleaned and descaled" means that the cleaning/descaling
system can be activated so that each tank is drained, cleaned and
descaled, flushed, and refilled automatically in an operator
programmable manner that greatly reduces or eliminates the need to
manually clean the tanks; (14) A water hammer eliminator will be
provided by NIOSH; (15) Acid neutralization of effluents will be done
by NIOSH; (16) The heat in the dryer shall be adjustable with a maximum
temperature which should be known, disclosed to NIOSH, specified by the
offer and between 205 F and 225 F; (17) The dryer provided shall be
able to dry polycarbonate cages so that they are free of visible beads
of water and can be filled with bedding or stored immediately; (18) An
exhaust system provided by NIOSH will be available in the room on the
unload side. The requirements at FAR 52.214.21 "Descriptive Literature"
apply. Proposals/Information provided must demonstrate the unit's
ability to meet the revised specification and that the supplier has at
least two years of experience in the production and installation of
tunnel washers. All specifications not affected by this amendment
remain as stated in the original specification (including the belt
width and tunnel height requirements). The due date for proposals is
extended to July 24, 1997. 2:00 p.m local prevailing time. All other
terms and conditions stated in the July 1, 1997, notice remain the
same. Fascimile submissions are not authorized. Proposals should be
submitted as follows: (1) one original and two copies of a completed SF
1449, signed by an individual authorized to bind the organization; (2)
a Schedule of Offered Supplies/Services containing pricing for the
system being offered, published price lists (if applicable) and a
delivery schedule; (3) completed Representations and Certifications --
Commercial Items, FAR 52.212-3; (4) descriptive literature and
demonstrated experience for evaluation purposes; (5) acknowledgment of
this Amendment 001A; and (6) exceptions, if any, to RFP. Submit to:
DHHS, PHS, CDC, NIOSH, ALOSH, Attn: Brenda Goodwin, MS/1019, 1095
Willowdale Road, Morgantown, WV 26505. Offers that fail to furnish the
required information, or reject the terms and conditions of the
solicitation may be excluded from consideration. Information regarding
this notice can be obtained from Brenda Goodwin at (304)285-5882/5876
or by fax at (304) 285-6083. All responsible sources may submit a
response, which shall be considered by the Agency. (0196) Loren Data Corp. http://www.ld.com (SYN# 0387 19970717\66-0007.SOL)