COMMERCE BUSINESS DAILY ISSUE OF AUGUST 20,1997 PSA#1913National Institutes of Health, OPM, OA/OD, 6120 Executive Blvd.,
Bethesda, MD 20892 66 -- X-RAY DIFFRACTION SYSTEM SOL 263-97-P(M7)-0047 DUE 090297 POC Al
Plyler, 301/402-3068/Contracting Officer, Edward B. Wilgus,
301/402-3070. This is a combined synopsis/solicitation for commerical
items description, prepared in accordance with the format in FAR 12.6,
as supplemented with additional information included in this notice.
This announcement constitutes the only solicitation, proposals are
being requested and a written solicitation will not be issued. The
solicitation, No. 263-97-P(M7)-0047 is issued as an Invitation for Bid
(IFB). The solicitation/contract will include all applicable
provisions and clauses in effect through FAR FAC-90-46. The Standard
Industrial Code is 3844. The NIH intends to contract with Molecular
Structure Corporation. 3200 Reseach Forest Drive, The Woodlands, TX
77381, under the authority of FAR 6-302-1 for this product. It is the
Government's belief that molecular Structure Corporation is the sole
source for this product. A. The X-Ray Diffraction System shall consist
of an image plate detector including controller, theta stage, x-ray
mirror, 6-piece set of erasing lamps, x-ray shutter, re-installation
andforty inch extension to existing radiation enclosure, data
processing software, and computer controlled crynogentic system. B. The
system shall allow users to select pixel sizes of 50 microns x 50
microns, 100 microns x 100 microns, or 200 microns x 200 microns. C.
The system detector shall consist of a dual photo multiplier
configuration having a dynamic range of 100,000. D. The sensitivity of
the system detector shall be I photon per ADC or better. E. The system
shall read plates in such a way that coordinate conversion to
orthogonal coordinates is not required. F. All system mirrors shall be
helium purged to limit air scatter and oxidizing effects. G. The
optical portion of the system shall have remote control of mirror
translation, curvature, and slit aperture. H. The diffraction system
shall include a subsystem having the capability to operate at -180
degree centigrade temperatures without the use of liquid nitrogen. I.
The system shall include a flexible transfer tube allowing for
positioning in any orientation from vertical to horizontal. J. The
Contractor shall furnish all items necessary to allow for mounting the
Diffraction system to an existing Rigaku RU200 rotating anode
generator. K. Shall include a HKL or equal, area detector processing
software package capable of reducing collected data. Offers must be in
writing and are due no later than 1:00 p.m. September 3, 1997, to the
attention of Al Plyler at the above address. (0230) Loren Data Corp. http://www.ld.com (SYN# 0294 19970820\66-0023.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|