Athens, GA 30605-2720 66 -- NMR SPECTROMETER SOL RFQ-029-4384-97 DUE 092297 POC Pat Lang,
Procurement Assistant, (706)546-3533; Alan Moore, Contracting Officer
(706)546-3533 WEB: Click here to download a copy of 52.212-3.,
http://www.nist.gov/admin/od/contract/repcert.htm. E-MAIL: Click here
to contact the contracting office via e-mail, saa-contr@ars.usda.gov.
Reference CBDNet Notice of August 26, 1997. This modification reference
correction to (v); (vi); (xii) under the description. (v) Contract Line
Item Number (CLIN) 1, is an NMR Spectrometer, 1 each. (vi) The NMR
Spectrometer console shall be compatible with a Spectrospin 89mm
vertical bore 7.05 Tesla superconducting magnet and with existing
Bruker probes. The spectrometer shall be capable of performing high
resolution CPMAS experiments on solid samples, high resolution,
multidimensional experiments in solution, and magnetic resonance
imaging on samples up to 25mm in diameter. Two frequency channels are
required, with the capability of the addition of more channels. Each
channel shall be capable of operating both asynchronously and
synchronously, with digital generation of offset frequencies and
phases. Each channel shall be capable of providing shaped pulses via
modulation and attenuation of the frequency synthesizer output. The
proton transmitter shall operate at 300 MHz and shall be capable of
simultaneously performing excitation and decoupling applications
requiring multiple power levels. The proton transmitter shall provide
up to 100 W in pulsed operation with an attenuation range of 110 dB.
Capability for utilization of an existing Bruker MSL 1kW high power
amplifier shall also be provided. The X frequency transmitter(s) shall
be capable of providing up to 300 W output power in pulsed operation
with an attenuation range of 110 dB. Capability for utilization of an
existing Bruker MSL 1 kW high power amplifier shall also be provided.
The receiver system shall be capable of digital oversampling and
digital signal filtering. An X,Y,Z gradient control system including
pre-emphasis and gradient current amplifiers or capability of utilizing
an existing Bruker MSL gradient control system shall be provided for
micro-imaging applications. Communication between the data acquisition
processor and the NMR workstation shall be through a dedicated
Ethernet/NFS link. The data acquisition processor shall be capable
ofcommunication with an existing 64 bit UNIX base operating system on
a Silicon Graphics Indigo 2 platform. (xii)FAR 52.212-5, Contract Terms
and Conditions Required to Implement Statutes or Executive Orders --
Commercial Items, applies to this acquisition. The following FAR
clauses cited in 52.212-5 are applicable to this acquisition: 52.203-6;
52.203-10; 52.219-8; 52.222-26; 52.222-35; 52.222-36; 52.222-37;
52.225-3; 52.225-9. (0239) Loren Data Corp. http://www.ld.com (SYN# 0295 19970902\66-0002.SOL)