|
COMMERCE BUSINESS DAILY ISSUE OF JANUARY 26,1998 PSA#2018U.S. Environmental Protection Agency, 26 W. Martin Luther King Dr.,
Cincinnati, OH 45268 66 -- MASS CHROMATOGRAPHIC ANALYZER SOL CI98-ORD22 POC Mike Hennessey,
513-487-2032 WEB: Office of Acquisition Management -- Contracts
Management Division -- Cincinnati, http://www.epa.gov/oam/cinn_cmd.
E-MAIL: hennessey, hennessey.mike@epamail.epa.gov. WRITTEN REQUESTS FOR
THE SOLICITATION WILL NOT BE ACCEPTED.The Environmental Protection
Agency has a requirement for a mass chromatographic analyzer that will
be used in conjunction with an existing HPLC pump (Waters 600E) and
autosampler (Waters 717plus). The EPA needs to detect, positively
identify, and quantify pesticides (and their metabolites) in tissues
from pesticide-treated animals. Currently, we have the capability to
detect and quantify the toxicants in question, but we do not have the
means for positive identification. The analyzer to be purchased will
provide that capability. In addition to the compatibility requirement,
analyzers must meet the following performance requirements: 1. Single,
gold coated ceramic quadrupole mass spectrometer with mass range of
30-3000 m/z, which will connect with existing HPLC pump (Waters 600E)
and autosampler (Waters 717 plus). 2. Ion transfer to mass spectrometer
shall occur without heat (which degrades labile samples), consisting of
a high pressure RF-only ion guide. 3. System shall have a
differentially pumped vacuum system with two 50 1/sec turbo molecular
pumps. Shall have fail-safe system protection with automatic shutdown
and restart after power failures. Turbo pumps shall not require a water
chiller even if room temperature exceeds 80 F. 4. Analyzer shall have
a continuous dynode electron multiplier (CEM) detector which operates
in the pulse counting mode and is capable of positive/negative ion
detection. The CEM polarity shall be switchable between alternate
scans. 5. Ion path optics, interface electronics, and gas supplies
shall be fully computer controlled. 6. System shall include
analyzer-computer interface that will communicate with embedded
processors to provide complete control of the vacuum system, API
sources, mass analyzer scan conditions, parameter settings and data
acquisition selections. 7. System shall include data acquisition and
processing software. 8. Analyzer shall be capable of CID/MS and
alternating CID/MS and MS experiments. 9. Resolution (M/ M) of
quadrupole shall be better than M (FWHH 0.8 u) throughout the mass
range of 30 to 3000 Da. 10. System shall include an atmospheric
pressure chemical ionization (APCI) source which has nebulizer
assisted, temperature controlled probe inlet with both probe and
discharge needle articulation. 11. APCI source shall have flow rate
compatibility of 0.2 to 2.0 ml/min, full gradient capability, and a
solvent compatibility of 100% aqueous to 100% organic. It shall also
have two viewing ports, with an off-axis sprayer configuration and a
discharge current of +/- 15 micro amperes. 12. System shall include an
atmospheric pressure ionization (API) source that accepts flow rate
from 50 L to 1000 L/min. 13. System shall include a dual channel
analog-to-digital converter to collect data from other detectors. All
responsible sources may submit a quote which shall be considered.
TELEPHONE REQUESTS WILL NOT BE ACCEPTED. The fax number is
513-487-2107. The E-mail address is hennessey.mike@epamail.epa.gov
(0022) Loren Data Corp. http://www.ld.com (SYN# 0313 19980126\66-0003.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|