COMMERCE BUSINESS DAILY ISSUE OF FEBRUARY 17,1998 PSA#2033DHHS/FDA/NCTR, Contracts Branch, HFT 322, 3900 NCTR Road, Jefferson, AR
72079-9502 66 -- SPECTROGRAPH, FULLY AUTOMATED, PC CONTROLLED. SOL R-5005-00-8
DUE 040898 POC Contract Administrator Karen Moore 870-543-7479,
Contracting Officer Marcia Park 870-543-7405 Spectrograph, fully
automated, PC controlled (Government provided PC), complete with all
necessary software. Device drivers must be compatible with MicroSoft
Windows 95. Double monochromator configuration, each having 1/4 meter
focal length, software controlled multiple grating turrets and equipped
with PMT detector for range of 200-900 nm. Second monochromator must
have two detection ports, allowing a choice between two detectors for
the system with the contorl software. Ports to accommodate CCD, PMT and
photodiode detectors. The operator must be able to choose between the
detector ports using provided software, and the mirror which diverts
the light between the two detection ports must be reproducible to at
least 0.05 nm. Filter assembly to eliminate 2nd-order radiation from
the gratings. Software to allow collection, printing and storage of
each spectrum and comparison of multiple spectra. Must have choice of
fixed slitwidths at the light entrance port, detector port and between
the monochromators. RFQ will be issued on or about March 5, 1998 with
a 4 week closing. Telephone requests for RFQ package will not be
honored. Mail or fax requests. If faxed, send to 870-543-7990 attention
Karen Moore. (0043) Loren Data Corp. http://www.ld.com (SYN# 0326 19980217\66-0012.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|