COMMERCE BUSINESS DAILY ISSUE OF MARCH 4,1998 PSA#2044U. S. ARMY ENGR. CEMVK-CT, CONSOLIDATED CONTRACTING DIVISION, 4155 CLAY
STREET, VICKSBURG, MS 39180-3435 66 -- HIGH RESOLUTION GAS CHROMATOGRAPH/HIGH RESOLUTION MASS
SPECTROMETER SOL DACW39-98-R-0011 DUE 042198 POC Contract Specialist
EMMA ROBERTS (601) 631-7257 (Site Code DACW39) Furnish and deliver a
high resolution gas chromatograph/high resolution mass spectrometer
(capable of performing EPA Methods 8290 and 1613). The instrument shall
be a new and unused high resolution, high sensitivity dual focussing
mass spectrometer system. This system includes: a fully laminated
magnet which can analyze 2000 amu at a minimum of 5 kV accelerating
potential with resolution of greater than 60,000 (10% valley), a plug
in source module with positive and negative ion capabilities, and a CI
reagent gas inlet. The instrument shall be computer controlled by a
computer system with a modern operating and graphics system. Data
system software shall permit all necessary acquisition routines
appropriate for mass spectrometry and perform complete automated
instrument control including autotune. The instrument shall be capable
of performance of USEPA Methods 1613 and 8290 for Dioxin and be
capable of satisfying the following performance factors: Signal to
Noise Ratio on TCDD: In the selected ion recording (SIR) mode at 10,000
resolution while selecting two masses from natural 2,3,7,8-TCDD (m/z
319.8965 and 321.8936) and two masses from 13C 2-2,3,7,8-TCDD (m/z
331.9368 and 333.9339) plus a lock mass and lock mass check (m/z
316.9824) from PFK. The signal to noise ratio at m/z 321.8936 from 100
femtograms of 2,3,7,8-TCDD standard will be: S/N 100:1 where noise is
peak to peak measurement encompassing 95% of noise data points
(Equivalent to four standard deviations, i.e., plus or minus 2 sigma
about the mean) and S/N 200:1 where noise is mean positive going noise
to mean negative going noise about mean (EPA protocol 8290)
(Equivalent to two standard deviations, i.e., plus or minus sigma about
mean). The instrument shall be capable of operating in a completely
automated mode. The instrument shall be demonstrated to perform methods
1613 and 8290 without any operator intervention beyond loading the
autosampler. Complete autotune, calibration, calibration check, and
re-autotune on failure of calibration check, operation from a remote
PC, notification at remote location by phone and computer of system
failure, failed autotune, calibration, or calibration check must be
demonstrated. The company shall provide an installation kit and manuals
kit and furnish technical representatives from the company to set up
the instrument in our laboratory and adjust it to meet established
standards of performance. Successful installation will be gauged by the
operating system achieving the operating specifications. Upon
completion of the applications training the system must be capable of
performing both EPA method 1613 and 8290 in a fully automated manner
(including recalibration and autotune). FOB USAEWES Reservation, 3909
Halls Ferry Road, Vicksburg, MS 39180, no later than 90 days after date
of award. Solicitation No. DACW39-98-R-0011, opening on or about 21
April 1998, 1400 hours CST. All responsible sources may submit a
proposal which shall be considered. This procurement will be available
only through the Internet on or about 17 March 1998. Potential
offerors may view and/or download this solicitation and all amendments
from the following Internet address: http://ebs.wes.army.mil. (0061) Loren Data Corp. http://www.ld.com (SYN# 0257 19980304\66-0007.SOL)