COMMERCE BUSINESS DAILY ISSUE OF APRIL 17,1998 PSA#2076USDA-ARS-MWA, Contracting Office, 1815 N. University Street, Peoria, IL
61604 66 -- INSECT ELECTROPHYSIOLOGY SYSTEM SOL 5114-8-0006 DUE 043098 POC
Lori J. Houghton, Contract Specialist, (309) 681-6616 The USDA,
Agricultural Research Service, Midwest Area, National Center for
Agricultural Utilization Research (NCAUR), Peoria, IL, is in need of an
Insect Electrophysiology System. This is a combined
synopsis/solicitation for commercial items prepared in accordance with
the format in FAR Subpart 12.6, as supplemented with additional
information included in this notice. This announcement constitutes the
only solicitation; quotations are being requested and a written
solicitation will not be issued. Solicitation No. 5114-8-0006 is being
issued as a Request for Quotation. This solicitation document and
incorporated provisions and clauses are those in effect through Federal
Acquisition Circular 97-04. Standard Industrial Classification Code
3821 with a corresponding size standard of 500 employees applies to
qualify as a small business. CLIN 0001 Data Acquisition Interface
board, Type IDAC-02, or equal, 1 each; CLIN 0002 Signal Interface Box
for IDAC-02, type ID-02, or equal, 1 each; CLIN 0003 EAG and GC/EAD
Signal Recording and Analysis Program, 1 each; CLIN 0004 Electrode
holder for glass pipette electrodes, 1 each; CLIN 0006 Electrode holder
for tungsten wire electrode, 1 each; CLIN 0007 Differential Probe, Type
PRD-1, or equal, 1 each; CLIN 0008 EAG Probe, Type MTP-4, or equal, 1
each; CLIN 0009 Stimulus Controller Unit, Type SFC-2, or equal, 1 each;
CLIN 0010 Effluent Conditioner Assembly, Type EC-03, or equal, 1 each;
CLIN 0011 Micro Knife Beveling Device, Type EG-03, or equal, 1 each;
CLIN 0012 Universal AC/DC Amplifier, Type UN-06, or equal, 1 each.
Salient Characteristics: Data Acquisition Interface board, Type
IDAC-02: Especially developed for electrophysiological signal
processing; 8 channel analog input; 16 bit resolution; max sampling
speed 100,000 sampl/s; 8 digital inputs; 8 digital outputs; 2 analog
outputs; trigger, pre-trigger, stimulus period, and status inputs;
software controlled on-board input amplifier; DSP (Digital Signal
Processor) for filter and conversion functions; Half size 16 bit
standard IBM type ISA extension slot compatible; Dedicated to process
AC and DC electrophysiological signals; Full signal capture to 100s;
Spike capture up to 7 hours continuous recording. Signal Interface Box
for IDAC-02, type ID-02 (required for signal interfacing IDAC-02 with
signal sources): Standard BNC input receptacles for external signals;
EAG input with automatic and manual base line control; Second DC input
channel; AC input with fine attenuation control; Anti Aliasing and
noise filters on analog signal inputs; Stimulus indication signal
input; Trigger command input; Direct input for Syntech probes (no
amplifier needed). EAG and GC/EAD Signal Recording and Analysis
Program: For use with Syntech IDAC data acquisition board; Windows user
interface; Auto zero on trigger signal of base line and time base;
Real-time recording and display of Electro Antennogram signals (EAG);
Calculation and display of EAG maxima and Normalized values;
Calculation and display of means and standard deviations; Dual trace
real-time recording of combined Gas Chromatography-Electro
Antennographic Detection (GC/EAD); max 5 hours continuous recording;
Peak area calculation and retention time marking on GC and EAD trace;
Amplitude and time-based zooming; Comments stored with data; Graphic
output in Windows metafile format; Printing of all graphic and numeric
data; ASCII data output. Electrode holder for glass pipette
electrodes: Stainless steel body; Easy replacement of silver wire; no
soldering; body on guard potential (no leaking current); Fits on all
Syntech probes. Electrode holder for tungsten wire electrode: To
replace holder for glass pipette electrodes; Fits on all Syntech
Probes. Differential Probe, Type PRD-1: Miniature input probe for
general AC and DC application; Adjustable input impedance 10 to the 6th
power to 10 to the 13th power Ohm in 6 steps; BNC input receptacles for
input cables; Low noise guarded input cables with electrode holders;
Stainless steel electrode holders; Easy replaceable silver wire (no
soldering); Can be connected directly to IDAC interface box. EAG Probe,
Type MTP-4: For use with electrically conductive gel; Interchangeable
holders for different sizes of antennae; Suitable for filiform
antennae; Stainless steel contact surfaces; High input impedance 10x
input stage; Can be connected directly to IDAC interface box. Stimulus
Controller Unit, Type SFC-2: Dual channel electronic mass flow
control; Flow: 0-50ml/s, digital read-out; Accuracy 1% Full Scale;
Pulse Time: 0.02 -- 10 s, 12 steps; Pause Time: 0.1 -- 10 s, 12 steps;
Interval/Period pause selector; Simultaneous/Alternate mode selector;
External (computer) control interface connector; Air Supply:
Instrument quality air, max pressure 9 Bar.; Power requirements: 110V
(220V) (10%), 50-60 Hz, 20 Watts; 19", 3 HE metal enclosure. Effluent
Conditioner Assembly, Type EC-03: Heated transfer line for GC-EAD;
Digital temperature controller max. 350 degrees; Interchangeable flow
tube; Adapted to fit available Gas Chromatograph; Prepared for Syntech
"Flash Desorption" System. Micro Knife Beveling Device, Type EG-03:
For glass and tungsten micro knife and electrode beveling; Adjustable
speed 30 cm diameter, interchangeable diamond rotor; Adjustable
beveling angle; adjustable rotor speed; Tool holder chuck on x-y-z
positioner; 90 degrees click mechanism; Built-in power supply for 110
or 220 V. 40-60 Hz; Compact size; can be placed under stereo
microscope. Universal AC/DC Amplifier, Type UN-06: Independent AC and
DC amplifier sections; DC Section: Automatic and manual base line
control; low pass filter for EAG signals; Gain: 10, 20, 50, 100, 200,
500, and 1000 x; Fine gain adjustment range: 1:2; Calibrator: 0.1; 1
mV; Signal indicator meter; Input /status LED indicator; Audio
amplifier with speaker and phones output; Normal/inverse switch; 50/60
Hz notch filter. Standard commercial warranty shall apply. FAR
52.212-1, Instructions to Offerors -- Commercial, FAR 52.212-3 Offeror
Representations and Certifications -- Commercial Items, FAR 52.212-4
Contract Terms and Conditions -- Commercial Items, and FAR 52.212-5
Contract Terms and Conditions Required to Implement Statutes or
Executive Orders -- Commercial Items apply to this acquisition. Offeror
must submit descriptive literature to provide proof that the proposed
equipment meets or exceeds the salient characteristics listed above.
Offer must include a 30 calendar day acceptance period. Offers are to
be received Not Later than the Close of Business, April 30, 1998 at
USDA-ARS-MWA, Contracting Office, 1815 N. University Street, Peoria, IL
61604. The POC for this acquisition is Lori Houghton, Contract
Specialist (309) 681-6616. (0105) Loren Data Corp. http://www.ld.com (SYN# 0280 19980417\66-0002.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|