COMMERCE BUSINESS DAILY ISSUE OF SEPTEMBER 9,1998 PSA#217682d Contracting Squadron, 136 K Avenue Ste 1, Sheppard AFB TX
76311-2739 J -- MEDICAL EQUIPMENT SERVICE AGREEMENT SOL F41612-98-Q0766 DUE
091498 POC LGCS/2Lt William P. Surrey, Contract Specialist,
940/676-2804; Contracting Officer -- MSgt Kevin E. Slone, 940/676-5640.
E-MAIL: Send all e-mail inquiries to: 2Lt William P. Surrey.,
William.Surrey@sheppard.af.mil. Provide all tools, equipment, labor,
parts, materials, facilities, and transportation necessary to perform
maintenance on the following equipment manufactured by Pyxis
Corporation: (Index # /Model # /Serial #)10310/Medstation/A05420277;
10311/M1M6/09524072A; 10312/M1M6/09524093A; 10313/M1M6/09524074A;
10314/M1M5/09524076A; 10315/M1M6/09524075A; 10316/M1M6/09530417A;
10317/M1M6/09524116A; 10318/M1M5/09524136A; 10319/M1M2/0954121M;
10320/M1A6/09524137A; 10321/M2M7/095514638I; 10322/M2M/09530557I;
10323/M2M/09530505I. Statement of work is below. Place of work will be
Sheppard AFB TX 76311 (Wichita Falls). Base year of service will be 1
Oct 98-30 Sep 99. Government is considering four (4) option years to
run consecutively commencing 1 Oct 99. This is a combined
synopsis/solicitation for commercial services prepared with the format
in Subpart 12.6, as supplemented, with additional information included
in this notice. This announcement constitutes the only solicitation.
Quotes are being requested, and a written solicitation will not be
issued. Only faxed (940/676-3784) and mailed quotes will be acceptable.
Deadline for quotes is 14 Sep 98. This solicitation documents and
incorporates provisions and clauses in effect through Federal
Acquisition Circular 97-06. This solicitation is issued under full and
open competition. FAR Provisions 52.212-1 and 52.212-2, and Clauses
52.212-5 and 52.212-5, apply to this acquisition with no addenda.
Offerors are required to complete and mail FAR 52.212-3, Offerors
Representations and Certification -- Commercial Items, with their
quote. Offerors are required to register with the Central Contractor
Registry prior to award (888/227-3784 or http://ccr.edi.disa.mil).
Defense priorities and allocations system rating S10 applies to this
acquisition. The Government anticipates award of a firm-fixed-price
purchase order. The following additional clauses are in effect by
reference: 52.217-8, 52.212-9 and 52.232-18. The following is the
Statement of Work: All maintenance service shall be performed by
equipment-manufacturer-trained, English-speaking, technically-qualified
servicemen who shall record all work actions in accordance with current
applicable regulations and/or industry standard practice. Contractor
will perform two (2) preventative maintenance inspections, in Jan 99
and Jul 99; the Government reserves the right to make changes to this
schedule by notifying the contractor two weeks in advance. The
contractor will give two weeks notice to the contract monitor prior to
performing scheduled service. All repair must follow manufacturer
procedure. The preventative maintenance inspection shall include, but
is not limited to, performing operational check on all systems and
correcting any deficiencies noted, verifying the systems' internal and
external calibrations are within the manufacturer's specifications,
cleaning and lubricating mechanical parts in accordance with applicable
manufacturer's service manual, performing an electrical check of the
equipment and all accessories included in this contract, and making
electrical adjustments as required to ensure compliance with
manufacturer's specifications. Replacing worn or defective parts (worn
parts are defined as parts that may fail prior to the next
preventative maintenance visit if they are not replaced), visually
inspecting exterior of all systems for damage and cleanliness,
reporting any discrepancies to the contract manager, making mechanical
adjustments to ensure proper operation, and consulting with the
equipment operator regarding equipment performance. Emergency service
calls: On-site response within 24 hours after notification for any
outage occurring Sunday through Thrusday; response within 48 hours
after notification for any outage occurring on Friday or Saturday.
Systems must be serviceable, to include replacement of defective parts,
within 24 hours of arrival on site. Reporting requirements for service
calls: Prior to accomplishing any work specified in the contract, the
contractor shall report to the Medical Equipment Repair Center (MERC),
Room B506, in person, to check in with the contract manager and make a
contract log entry to establish arrival time. An escort will be
provided. The MERC will develop and maintain a legal log of system
performance, noting dates and times of notification of maintenance
requests, response, and completion. The log will provide data for the
Government to determine if service requirements are being met, and will
be available for review by the contractor. Upon departure, the
contractor must report to MERC control desk and present work
documentation. Contractor must ensure a work completion entry is made
in the call log. Replacement parts: all non-consumable parts are
covered. All defective replace parts will be returned to the contract
manager. All replacement parts and/or repair parts shall be either new
or remanufactured by the manufacturer's repair center, and then
quality assurance tested and approved by the manufacturer's quality
assurance department. Nonavailability of required parts will not be an
excusable delay for not completing repairs within the previously
specified time frames. Upon completion of each service visit, the
contractor shall supply written calibration documentation to the
contract monitor; this documentation must specify accuracy
specifications and tolerances to which the euqipment will be
calibrated. Actual and measured values normally required to certify
calibration or proper system operation must be listed. Contractor must
complete and affix the Medical Equipment Verification Certification
Label, DD Form 2163, to each system upon completion of a scheduled
service inspection or unscheduled service which required a major
component change that could affect system calibration. The form is
available from the contract manager. Upon the completion of each
service call, the contractor shall provide the contract manager a
legible service report. The contract manager will countersign the
report to certify that services were rendered. The Government will not
recognize any maintenance actions that have not been reported and
singed by the Government contract monitor. Service reports shall
include, but not be limited to: purpose of service, index # (equipment
ID tag), serial #, description of defect and action taken, signature
of service person(s) actually performing repairs or servicing, complete
list of parts used (to include part # and cost), travel time and
maintenance/repair time which was involved with the service, and
signature of the contract monitor. The contractor will be held liable
for errors and omissions made during the performance of this contract.
Payments will be made semiannually upon receipt of a certified
invoice. Failure in any facet of this contract may delay payment.
Invoices should be sent to: 82 CONS/LGCS, 136 K Avenue Ste 1, Sheppard
AFB TX 76311-2746. Posted 09/04/98 (W-SN246558). (0247) Loren Data Corp. http://www.ld.com (SYN# 0037 19980909\J-0001.SOL)
J - Maintenance, Repair and Rebuilding of Equipment Index Page
|
|