|
COMMERCE BUSINESS DAILY ISSUE OF OCTOBER 2,1998 PSA#2193U.S. Army Aviation and Missile Command, Acquisition Center, Bldg 4488,
Redstone Arsenal, AL 35898-5280 66 -- DOSIMETRY GRADE X-RAY CALIBRATION SYSTEM SOL DAAH01-99-R-0008
DUE 101698 POC Betty Carden, 256-842-7223 (fax: 256-955-6449) Contract
Specialist, or Shirley Hill, 256-842-7436 Contracting Officer THIS IS
A COMBINED SYNOPSIS/SOLICITATION for commercial items prepared in
accordance with the format in Federal Acquisition Regulation (FAR)
Subpart 12.6, as supplemented with additional information included in
this notice. This announcement constitutes the only solicitation.
Proposals are being requested and a written solicitation will not be
issued. This solicitation is a Request For Proposal (RFP) Number
DAAH01-99-R-0008. The RFP and incorporated provisions and clauses are
those in effect through Federal Acquisition Circular (FAC) 97-06. This
solicitation is competitive, restricted to small businesses. The
Standard Industry Code (SIC) is 3841, number of employees is 500. The
requirement is as follows: 1. SCOPE: This purchase description covers
a dosimetry grade x-ray calibration system. 2. COMPONENTS: The x-ray
system shall include the following: a. X-ray tube; b. High voltage
generators; c. Cooling system; d. Shielded enclosure; e. Beam shutters;
f. Filter assembly; g. Safety system; h. Rail systemwith movable
exposure table; i. Control console. When assembled, the components
shall form a complete 320 kilvolt (kV) constant potential (cp), high
stability x-ray calibration unit. 3. PURPOSE: The X-ray calibration
system will be used in the Primary Radiation Standards Center to (1)
calibrate x-ray detection instrumentation for U.S. Army hospitals and
medical activities worldwide as well as survey instruments and
dosimeters used in support of recently fielded DA non-destructive test
equipment (NDTE) x-ray systems. 4. SALIENT CHARACTERISTICS: The x-ray
system shall incorporate the following salient characteristics, which
are essential to the needs of the Government: a. X-ray Tube: (1) The
X-ray tube shall be a double focus tube and shall be a double pole tube
with stationary anode and oil insulation. It shall have a minimum 2 mm
thick beryllium (Be) window and no more than 4 mm Be or Be equivalent
total inherent filtration. The full solid angle of the emergent x-ray
beam shall not exceed 20 degrees and shall not be less than 10
degrees. (2) The tube head shall have minimum ratings of 320 kVcp and
13 milliamperes (mA) on a 4.5 mm focal spot, and 320 kVcp and 3 mA on
a 1.2 mm focal spot. (3) The tube shall be designed for protection
against leakage radiation. b. High Voltage Generators: (1) The unit
shall include two high voltage generators, one furnishing a minimum of
-160 kVcp to ground, the other furnishing +160 kVcp to ground. The
generators shall be voltage-doubling with smoothing capacitors and
solid state rectifiers. The high voltage generating components shall be
oil immersed in rugged steel tanks. (2) A solenoid grounding system
shall be provided to automatically ground any electrical charge
remaining in the capacitors and cables when the entire unit is turned
off. (3) The generators shall be supplied with interconnecting cables
to the control panels/console. (4) High voltage ripple shall not exceed
70 volts peak-to-peak at maximum system output. (5) External means of
monitoring kV and mA shall be provided. (6) A 1500 Hz regulated
filament power supply shall be provided to minimize modulation of the
x-ray output due to filament intensity variations. (7) High voltage
cables shall be provided which are shockproof and which shall be
provided with special rubber cone plugs and snap-type locks. The cables
shall be interchangeable (capable of being used for either anode or
cathode connection). A means of locking the high voltage cable ends to
their plugs shall be provided in order to prevent unauthorized
personnel from removing cables. Two protective terminal caps shall be
provided with each cable. c. Cooling system: An oil-to-water heat
exchanger to cool the x-ray tube head shall be provided. Water lines
used to connect the heat exchanger to the facility water supply shall
be provided. d. Shielded enclosure: A rectangular shielded enclosure
constructed of lead lined steel, which will completely enclose the
x-ray tube head and head mounts, shall be provided. The enclosure shall
limit leakage radiation from the tube head to less than 0.5 mR/hr at a
distance no greater than 5 cm from any part of the enclosure's
exterior surface when the tube is operating at maximum rated kVcp and
mA. The enclosure shall be configured such that the x-ray tube head is
mounted within with its long axis horizontal to the floor and the
x-ray beam port is 150 cm above the floor. e. Beam shutters: Two beam
shutters, a safety shutter and a high-speed shutter, shall be provided.
During normal operation, the safety shutter shall open first, followed
by the high-speed shutter. At the end of the exposure, the high-speed
shutter shall close, followed by the safety shutter. Both shutters
shall immediately close upon power-down of the system. (1) Safety
shutter. The safety shutter shall be mounted inside the shielded
enclosure. When closed, it shall limit the intensity of the useful beam
to less than 0.5 mR/hr at a distance 5 cm from the front surface of the
shutter for the maximum rated kVcp and mA. It shall move from fully
closed to fully open (and vice versa) in 1.0 second or less. (2)
High-speed shutter. A digitally timed high-speed shutter, used to
control the exposure period, shall be mounted between the safety
shutter and the tube head. The shutter shall move from fully closed to
fully open (and vice versa) in 0.25 +/- 0.01 second. f. Filter
assembly: An assembly which includes at least 10 slots, each of which
will hold a set of filters shall be provided and mounted to the front
surface of the shielded enclosure. The assembly shall rotate such that
the selected filter combination will be positioned directly in front
of the x-ray beam port. The slots shall be designed such that they can
hold filters of 8.3 cm diameter ranging in thickness up to 25.0 mm. g.
Safety system: The x-ray system shall be provided with the following
safety features: (1) A SCRAM control located within the room containing
the tube head which will enable power down of the system when
activated. (2) A door interlock which will immediately close the safety
shutter when the door to the x-ray room is opened. The shutter shall
not be able to be re-opened until the door is closed and the interlock
system reset. Interlock system reset shall be accomplished from the
control console. (3) An EMERGENCY OFF control to be located on the
control console which will shut the system down when activated. (4)
Warning lights on the control console and at the entry door to the
x-ray room which will indicate when high voltage is on the x-ray tube
and when the safety shutter is open. (5) Audible pre-warning signals
when (1) high voltage is applied to the tube and (2) when the shutters
are opened. h. Rail system with movable exposure table: A remotely
positionable exposure table which will hold instrumentation and
dosimeters to be irradiated shall be provided. The table shall ride on
a track which will in turn be mounted on a platform built into the
floor of the x-ray room (already in place). Track length shall be at
least 5 meters but shall not exceed 6 meters. The platform consists of
two heavy-duty aluminum rails with centers 75 cm apart. Table distance
from the face of the x-ray tube window shall range from 50 cm to at
least 550 cm. The table shall be height adjustable from the centerline
of the radiation beam to at least 30 cm below the centerline and
laterally adjustable to at least 25 cm either side of the beam
centerline. Table positioning in all three axes (distance, height
adjustment, lateral adjustment) shall be within 1 mm of selected
position. Dimensions of the table surface shall be at least 30 cm x 40
cm. Table positioning in all three axes shall be accomplished from the
control console. i. Control console: A control console, enabling remote
operation of the x-ray system and monitoring of x-ray system status,
shall be provided. The console shall include a keyed switch for turn-on
and normal turn-off of the system. It shall include an EMERGENCY OFF
control for emergency shut down of the system. Other specific
requirements for the control console are as follows: (1) The console
shall include controls enabling the operator to set and apply tube kV
and mA. Voltage shall be selectable from 10 kV to 320 kV in 0.1 kV
increments. Kilovoltage accuracy shall be better than +/- 1 percent of
set value. kV drift as a function of temperature shall not exceed
0.005 percent. Current shall be selectable from 0.1 mA to 13.0 mA in
0.05 mA increments. Current accuracy shall be better than 0.2 percent
of set value. The console shall include a display which indicates set
and actual kV and mA. kV display resolution shall be 0.1 kV and mA
display resolution shall be 0.1 mA. (2) The console shall include a
control whereby the operator may select the filter to be used. The
console shall include a display indicating filter in use. (3) The
console shall include controls whereby the operator may position the
exposure table. The console shall include a display indicating table
position in all three axes. (4) The console shall include controls for
opening and closing the safety and high-speed shutters. The console
shall include an automatic reset, six digit exposure timer to count
from 0 to 999.999 minutes in .001 min increments for control of the
high-speed shutter. The timer shall be capable of preset or manual
control of exposure time. The timer shall have a time-out feature that
will enable an irradiation interrupted by closure of the safety
shutter during either preset or manual operation to be resumed and
completed upon reopening the safety shutter. The timer shall begin
counting down from the indicated pre-set exposure time when the
high-speed shutter is in the fully open position. When the timer
reaches zero, x-ray exposure is terminated. (5) The console shall
include a control whereby the operator may select the tube focal spot
to be used. Focal spot in use shall be displayed on the console. (6)
Console controls shall include protective interlocks to prevent
equipment damage from loss of coolant flow, excessive voltage or
current, or prolonged exposure periods. Controls shall not allow
selection of operation outside the system's rated capacity. (7)
Automatic warm-up sequences selectable from the control console shall
be provided to correctly energize the x-ray tube when the system is
cold. At least three programs shall be provided to accommodate various
system down times (day, week, month). The appropriate mode shall be
automatically selected based upon battery-backed real time clock.
Manual override shall be provided for this feature. (8) The console
shall include an RS232 interface to enable potential computer control
of portions of the system. j. General requirements: (1) The system
shall be operable on 208 volt three phase, 60 ampere AC power and shall
require no additional line conditioning against noise or transients for
operational compliance with specifications. Appropriate
electrical/mechanical protection devices shall be provided to preclude
equipment damage due to overloads and component failures. (2) The
contractor shall provide all cables and accessories required to enable
proper operation of the x-ray system in accordance with these
specifications. The control console will be located approximately 50
feet from the tube head. The high voltage generators and heat exchanger
will be located no more than 20 feet from the tube head. k.
Installation and training: The system shall be installed and made fully
operational in the Primary Radiation Standards Center, Redstone
Arsenal, AL by the contractor. The contractor shall provide training in
the operation and maintenance of the system. The contractor shall
provide operation manuals for system elements. One manual is required
with each instrument. Proposal is firm for 90 days unless otherwise
stated. Required delivery is 90 days after contract award. Earlier
delivery is acceptable and desirable at no additional cost to the
Government. Inspection, acceptance, and FOB are destination: Redstone
Arsenal, AL. An evaluation shall be performed to determine if each
proposal is technically acceptable. A technically acceptable proposal
must clearly show that the offeror can meet the minimum requirements of
the Government. Each proposal will be reviewed to assure that each
function included in the requirement is addressed in the proposal.
Therefore, each offeror must respond to each functional requirement of
the specification. The Government intends to award a contract
resulting from this solicitation to the responsible, responsive,
technically acceptable and lowest priced offeror whose offer conforms
to the solicitation and will be most advantageous to the Government.
All FAR clauses and forms may be obtained on the internet at
www.arnet.gov/far/. The FAR provision at 52.212-1, Instructions to
Offerors -- Commercial Items, applies to this acquisition. A completed
copy of the FAR provision at 52.212-3, Offeror Representations and
Certifications -- Commercial Items, must be completed and submitted
with the proposal. The FAR clause at 52.212-4, Contract Terms and
Conditions -- Commercial items, applies to this acquisition, with the
following addenda: FAR 52.247-34, F.O.B. Destination,FAR 52.212-5,
Contract Terms and Conditions Required to Implement Statutes or
Executive Orders -- Commercial Items, applies to this acquisition; the
following clauses under paragraph (b) are applicable: (6) 52.222-26,
Equal Opportunity (E.O. 11246); (7) 52.222-35, Affirmative Action for
Special Disabled and Vietnam Era Veterans (38 U.S.C. 4212); (8)
52.222-36 Affirmative Action for Handicapped Workers (29 U.S.C. 793);
(9) 52.222-37, Employment Reports on Special Disabled Veterans and
Veterans of the Vietnam Era (38 U.S.C. 4212); Defense Federal
Acquisition Regulation Supplement (DFARS) provision 252.225-7000, Buy
American Act -- Balance of Payments Program Certificate applies to this
acquisition. DFARS clause 252.212-7001, Contract Terms and Conditions
Required to Implement Statutes or Executive Orders Applicable to
Defense Acquisitions of Commercial Items, applies to this acquisition;
the following clause under Paragraph (b) is applicable: 252.225-7001
Buy American Act and Balance of Payment Program. The proposal must be
signed by an official authorized to bind your organization. You must
submit two (2) copies to: U.S. Army Aviation and Missile Command, ATTN:
AMSAM-AC-SM-T/B. Carden, Redstone Arsenal, AL 35898-5280. Offers may be
submitted on letterhead stationary and, at a minimum, must show: (1)
the solicitation number, (2) the time specified in the solicitation for
the receipt of offers, (3) the name, address, and telephone number of
the offeror, (4) a technical description of the items being offered in
sufficient detail to evaluate compliance with the requirements in this
solicitation. This may include product literature, or other documents,
if necessary, (5) terms of any express warranty, (6) price and
discount terms, (7) a completed copy of the representations and
certifications. Offers that fail to furnish required representations or
information, or reject the terms and conditions of this solicitation
may be excluded from consideration. See Note 1 ***** Posted 09/30/98
(W-SN256403). (0273) Loren Data Corp. http://www.ld.com (SYN# 0213 19981002\66-0013.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|