COMMERCE BUSINESS DAILY ISSUE OF NOVEMBER 13,1998 PSA#2221U. S. Environmental Protection Agency, Contracts Management Division
(MD-33), Attn: ORDSC, Research Triangle Park, NC 27711 66 -- XAD DENUDER SEMI-VOLATILE ORGANIC AEROSOL SAMPLER SOL
RFQ-RT-99-E2D315 DUE 112698 POC Starry M. Jones, Contract
Administrator, (919)541-2897 or Julio Lopez, Contracting Officer,
(919)541-4474 The synopsis issued on November 9, 1998, for the Organic
Aerosol Sampler is cancelled and replaced by this submission: The
Environmental Protection Agency has a requirement for a semi-volatile
organic aerosol sampler system to consist of a cyclone, a fitting to
connect two denuders in parallel, 2-100 lpm denuders, housing to
maintain denuders at ambient 5 C, filter holder, 2-XAD holders,
2-filter holders, fitting to connect XAD holders to flow controller,
one flow controller and one pump, valves to allow computer operation of
either denuder system, data system to record output of flow controller,
ambient and denuder temperatures and relay to start the pump. Inlet:
cyclone with 50% cutpoint 2.5 m at 100 liters/minute. Denuder: 8
channel etched glass denuder with 1mm channel, 52mm OD, total length
650mm with etched of 600mm centered in denuder. Flow rate through each
denuder is 100 liters/minute. Each denuder to be coated with XAD
microresin. Denuder housing: sufficient in size to contain two
denuders, filter and XAD holders and maintain temperature within 5 C of
ambient. Flow controllers and pump: system to consist of flow
controller and pump. Flow controller and pump must maintain flow within
10% of 100 liters/minute under with ambient temperature between -20 to
40 C. If possible, a flow controller based on the volumetric flow at
ambient temperature and pressure is preferred. System to be sized to
collected ambient air sample of PM2.5 for 24 hours at up to 50 g/m3.
Sample flow is 100 liters/minute through cyclone and denuder, 3 m
Zefluor, quartz, or Teflon coated quartz filter (diameter 9cm), and bed
of XAD2. XAD2 bed must be sufficient to trap naphthalene under above
conditions. Flow controller and pump to be inclosed in weather
protected housing. Data acquisition system: sensors for monitoring and
storing ambient and denuder temperature, outputs from mass flow meter
controller, vacuum before flow controller, time of day, and relay to
turn pump system on and off. Data to be recorder with data acquisition
system. The Environmental Protection Agency anticipates an award of a
fixed price purchase order using Simplified Acquistion Procedures.
This procurement is a total small business set aside. All interested
parties must submit a written request for the RFQ within fifteen (15)
days of the date of this announcement to the Attention of Starry M.
Jones, U. S. Environmental Protection Agency, Contracts Management
Division, MD-33, 79 T. W. Alexander Drive, Research Triangle Park, NC
27711. No telephone request will be honored. All responsible sources
may submit a quote which shall be considered by the Agency. Posted
11/10/98 (W-SN270026). (0314) Loren Data Corp. http://www.ld.com (SYN# 0398 19981113\66-0001.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|