COMMERCE BUSINESS DAILY ISSUE OF DECEMBER 8,1998 PSA#2237DHHS/FDA/NCTR, Contracts Branch, HFT 322, 3900 NCTR Road, Jefferson, AR
72079-9502 66 -- HAMILTON THORNE RESEARCH NO. 790500 MULTI-SPECIES SPERM ANALYSIS
SYSTEM SOL R-5007-00-9-00 DUE 012199 POC Contact Karen Moore, Contract
Administrator, 870/543-7479/Priscella Thomas Sullivan, Contracting
Officer, 870/543-7271 66-It is the Government's intention to negotiate
on a Sole Source basis with Hamilton Thorne Research for a 790500
Multi-Species Sperm Analysis System which includes; 1 EA. Multi Species
Sperm Analysis System, 1 EA. Oval Head Sperm Motility Software, 1 EA.
Metrix Morphology Software, 1 EA. Rat Metrix Morphology Software, and
1 EA. 40x Fluorescent Objective. The system must have the following
minimum specifications: Specimen Holder; Temperature Control must range
from Room Temperature to 45 deg. C, Temperature stepsize of 0.1 deg.C,
Temperature stability of 0.5 deg. C, Programmable Stage Position.
Specimen Chambers; Cannula with depth of 100 micron and >30 fields,
slide with 20 and 50 microns and 30 fields, programmable user
definition. Internal Optical System; Imaging device, high resolution
CCD array, objective minimum of 4x, 10x, 10 UV, image type, dark field,
bright field, phase contrast, negative and positive, image output,
NTSC, RS-170 60Hz. Illumination System; light source, stroboscopic,
photometer with an on screen scale, exposure to include source on only
during acquisition and focus, pulse 1.3 millisecond, intensity,
computer controlled, monitor, digitized image with 640 x 480 pixels,
spectra-indent fluorescent light source, illumination, epifluorescent,
source strobe Xenon, pulse, 5-8 millisecond, intensity, computer
controlled, filter, double bandpass, 450 mm and 660 mm. Data
Acquisition, frame rate, 60, 30, 15, 7.5 Hz, frames from 5 minimum and
100 maximum, system must have 1-20 fileds, designation, automatic or
selected. Analyzing System; Input signal, NTSC, RS-170, control from a
mouse and keyboard, analysis time, 30 seconds for 200 cells, software
to measure count (total, motile, progressive, % motile, %
progressive), concentrations (total, motile, progressive (m/mL),
distributions (LIN, STR, VCL, VSL, VAP, ALH, BCF, head
size-elongation), and total sperm count from homogenized samples.
System has a Quality Check Playback and Quality Control Graphs,
Analysis Sets to include two preset and five user defined. Video
Capability; Able to analyze both 50 Hz (PAL) and 60 Hz (NTSC). Standard
Toxicology Software; Includ software that does "counts",
"concentrations", "mean values", "distributions", and "Graphics". Data
Output; Internal hard drive, floppy disk drive and an external
monitor. Internal Computer System; This system must have 200 MHz
Pentium CPU, 32 MB RAM, 2 serial and 1 parallel ports, 1.44 MB 3.5"
high density floppy drive, 2.5 gigabyte hard drive, 17" Super VGA color
monitor and must be year 2000 compliant. This system must also come
with the following; Spectra Ident automated sperm concentration
analysis of homogenized cauda samples using DNA specific fluorescent
stain and integrated fluorescent illumination. Sort function which
determines fraction of cells within user specified limits on VCL, VSL,
LIN, STR, ALH, BCF, Head size and Head Shape. Track editing, which
allows you to view and store detailed data for individual sperm tracks
(used for validation procedures). HDATA ASCII exportwhich allows the
transfer of summary data and/or individual track to ASCII compatible
spreadsheets for database programs. Clinical filing which provides the
ability to design three one page reports and store analysis reports to
file. INFORMATION RECEIVED IN RESPONSE TO THIS NOTICE WILL BE USED FOR
DETERMINING WHETHER TO CONDUCT A COMPETITIVE PROCUREMENT. WRITTEN
SUBMISSIONS OF INTEREST SHOULD BE SPECIFIC IN IDENTIFYING THE
PROSPECTIVE OFFEROR'S CAPABILITY TO PERFORM. See numbered note 22*****
Posted 12/04/98 (W-SN277356). (0338) Loren Data Corp. http://www.ld.com (SYN# 0293 19981208\66-0017.SOL)