|
COMMERCE BUSINESS DAILY ISSUE OF FEBRUARY 22,1999 PSA#2287U.S. Environmental Protection Agency, 26 W. Martin Luther King Dr.,
Cincinnati, OH 45268 66 -- INDUCTIVELY COUPLED PLASMA OPTICAL EMISSION SPECTROMETER SOL
99-72 DUE 030599 POC Mike Hennessey, 513-487-2032 E-MAIL: EPA,
Contracts Division, Cincinnati, hennessey.mike@epamail.epa.gov. This is
a combined synopsis/solicitation for commercial items prepared in
accordance with the format in FAR Subpart 12.6, as supplemented with
additional information included in this notice. This announcement
constitutes the only solicitation; proposals are being requested and a
written solicitation will not be issued. Solicitation number 99-72 is
issued as a request for quotation. The solicitation document and
incorporated provisions and clauses are those in effect through Federal
Acquisition Circular 97-09. This is an unrestricted acquisition General
-- Simultaneous ICP-OES with dual viewed plasma, free-running 40MHz
Radio Frequency (RF) generator, purged echelle polychromator and Charge
Coupled Device (CCD) detector. RF Generator, Plasma ignition and
Viewing:1)The RF generator shall be free-running at 40 MHz and have a
optimal power output range of at least 800-1500 watts, as well as
computer controllable 5 watts increments. The out- put power stability
shall be better than 0.1%; 2)The instrumentshall maintain the plasma
power set point even when changing sample matrices. The power control
shall monitor power amplifier grid voltage, plate current and plate
voltage; 3) Plasma ignition and extinguishment shall be computer
controlled and totally automated; 4)Plasma viewing shall be switchable
between axial and radial, and be controllable by the computer. The
system shall be able to view sequentially the torch both axially and
radially in the same analytical method; 5)The instrument shall have a
means to displace the cool end of the axial plasma to minimize
self-absorption and physical interferences. Samples Introduction and
Plasma Torch -- 1)The instrument shall include both Scott and cyclonic
spray chambers, and both cross-flow and Meinhard nebulizers that can
handle at least 50% HCl, HNO3, H2SO4, H3PO4, 20% HF and 30% NaOH. The
nebulizer shall also be able to handle an hour spray of 20% NaCl
without clogging; 2)The system shall include a four channel, variable
speed, computer controlled peristaltic pump. The peristaltic pump shall
automatically pump the spray chamber drain outlet effluent to waste; 3)
The system shall include an easily accessible torch to allow a quick
change of torch. The plasma torch shall be of a demountable center tube
design. The alignment of the torch shall be computer controlled; 4)The
instrument shall include a shutter and mercury recalibration system
that monitors system conditions and ensures optical stability. The
shutter system shall be computer operated to shield the detector
between samples from long exposure to UV radiation. A mercury lamp
shall be built into the shutter system for an automatically updated
wavelength calibration. Optics System and Detector -- 1)The system
shall have a thermostatted Echelle based polychromator that has no
moving parts. The optical system shall be purged and shock-mounted to
the frame to prevent any adverse effect on the instrument performance
by any floor vibrations. 2) The instrument shall have a CCD detector
that is able to perform determinations across the entire spectrum, both
UV and visible (167 -- 766 nm), for all the desired elements in one
analytical method event. PC Control and Software -- 1)The instrument
controlling software shall run under Windows NT or Windows 95; 2)The
instrument shall be computer controlled including all gas flows, plasma
viewing height or X-Y torch position, ignition, RF power, safety
interlocks and utilities monitoring; 3) The software shall have the
ability to perform the automatic torch alignment and remotely control
the plasma igniting or extinguishing and start/stop gas flow and
sampling pump; 4)The software shall be able to read background and
emission signal simultaneously and allow for manual and automatic
background correction; 5) The software shall be able to display
calibration curves and all the peaks or numerical readings of all the
elements analyzed in each analytical run simultaneously; 6)The software
shall be able to correct for spectral interferences. Interfering
Element Correction (IEC) shall be calculated automatically; 7) The
instrument shall be able to run at least two quality control (QC)
samples with user defined limits for each element analyzed. If QC
fails, corrective action such as recalibration and rerunning of samples
shall be available; 8)The software shall have a library system
containing several analytical wavelengths for each element;
9)Calibration curves shall be stored and be available to recall for
later use. Raw data shall be saved for post run processing such as
changing of background corrections, IEC factors and selecting the
desired wavelength or elements to report; 10) The report format shall
be customizable to include various statistics such as daily calculated
detection limit of each element, standard deviations, besides the
analytical data that have been corrected for background and IEC; 11) A
PC with Pentium II 350 MHz, 64 MB RAM, 6.0 GB hard drive and 32X
CD-ROM is a minimum; 12) The computer shall have an Ethernet card with
10 MB on board and a serial port toallow for connection of an external
JAZ drive; 13) The system shall also contain an external JAZ drive of
2.0 GB, a 20" or larger color monitor and an inkjet printer. Analytical
Performance- 1)The instrument shall meet all EPA contract laboratory
required detection limits; 2) The instrument shall have the following
benchmark resolutions: As 193.7 nm 0.006nm; Ni 231.6 nm 0.008 nm; 3)
Detection limits (based on 3s, in mg/L) shall be at least or better
than the following: Al 2.0; As 2.0; Be 0.08; Cd 0.2; Cr 0.2; Hg 0.5; Mo
0.2; Ni 0.7; Pb 0.8;Sb 2.0; Se 2.0; V 0.4; U 15.0; Zn 0.2; ( 4. The
upper linear level (in mg/L) shall be at least or better than the
following: Na 6000; Ca 1000; K 1000; Fe 500; Al 400;Mg 350; Mn 400.
Miscellaneous -- The system shall include: 1)A water recirculator for
the instrument cooling system; 2) An autosampler with a rinsing
station, capable of housing at least 95 samples of 15 mL each; 3) A
spare kit includes commonly required replacement items and consumables
shall be included in the system. The warranty shall include:1) At
least one year on parts, labor and travel, three years for detectors
and computer; 2) A minimum of two days on-site training, plus the
tuition for one analyst for a 3 days training course on the instrument
maintenance in the manufacture training center; 3)A guarantee of
returning a user's service call within 24 hours and of scheduling a
service visit within 2 business days. Items are to be delivered to the
U.S. EPA/Robert S. Kerr ERC, 919 Kerr Research Dr., Ada, OK 4821-1198
4/1/99. F.O.B. point shall be destination. The provisions at 52.212-1,
Instructions to Offerors-Commercial, applies to this to this
acquisition. The provision at 52.212-2, Evaluation-Commercial Items is
applicable and the specific evaluation criteria are as follows:
(1)Technical capability of the item offered to meet Government
requirement and (ii) price. Offerors shall include a completed copy of
the provision at FAR 52.212-3, Offeror Representations and
Certifications-Commercial Items, withits offer. Clause 52.212-4,
Contract terms and Conditions-Commercial Items applies to this
acquisition. The Clause at 52.212-5, Contract Terms and Conditions
required to Implement Statutes or Executive Orders-Commercial Items,
applies to this acquisition, however, for paragraph (b), the following
clauses apply to this acquisition: 52.222-26 Equal Opportunity;
52.222-35, Affirmative Action for handicapped Workers; 52.222-37
Employment Reports on Special Disabled Veterans and Veterans of the
Vietnam Era; and 52.225-3 Buy America Act -- Supplies. Offers shall
also include sufficient descriptive literature to demonstrate that the
items offered meet the minimum requirements specified above. Offers
shall be submitted no later than March 5, 1998 at 4:00 p.m. E.S.T. to
Mike Hennessey at: regular mail: U.S. EPA, Contracts Management
Division, 26 West Martin Luther King Drive, Cincinnati, OH 45268, or by
FEDEX: U.S. EPA, 4411 Montgomery Road, Suite 300, Norwood, OH 45212.
Questions or comments may be directed to Mike Hennessey at
(513-487-2032), by e-mail (hennessey.mike@epamail.epa.gov). Posted
02/17/99 (W-SN299031). (0048) Loren Data Corp. http://www.ld.com (SYN# 0294 19990222\66-0001.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|