|
COMMERCE BUSINESS DAILY ISSUE OF MARCH 17,1999 PSA#2304U.S. Environmental Protection Agency, 26 W. Martin Luther King Dr.,
Cincinnati, OH 45268 66 -- GAS CHROMATOGRAPH -- MASS SPECTROMETRY (GC/MS) SYSTEM SOL
PR-CI-26510NL DUE 033199 POC J. Deanna Easley, Contract Specialist,
(513) 487-2027 WEB: Office of Acquisition Management -- Cincinnati
Contracts, http://www.epa.gov/oam/cinn_cmd. Minimum Needs The U.S. EPA
has a need for a Gas Chromatograph coupled with a Mass Spectrometry
Analytical Detection System which will primarily be used for the trace
analysis of human and/or environmental samples for pesticides,
endocrine disruptors and metabolites. The system must have the
following features: (1) Gas chromatograph must have the ability to add
two detectors in addition to the Mass Spectrometry Analytical
Detection System. The system must have the capability of capturing data
from all three detectors simultaneously. This will allow for multiple
detector capability to aid in the identification of components in
extremely complex environmental mixtures. (2) Retention time locking or
an equivalent method of reproducing analyte retention times and
ensuring instrument stability is required due to the complex nature of
the anticipated samples. In other words the GC shall have electronic
pressure control on all inlets and detectors and be capable of
reproducing analyte retention times independent of ambient temperature,
barometric pressure, column length or detector as long as columns of
same stationary phase, nominal phase ratio and diameter are used. This
will prevent recalibration with an entire suite of standards after a
column change, method transfer or routine maintainance. The neccesity
of retention time locking is to allow the use of an original
calibration curve and preclude a labourious re-injection of hundreds of
individual standards. The retention time locking library will narrow
the search to a few peaks from several hundred to identify unknown
peaks. This will prevent the neccesity of keeping, hundreds of
pesticide standards on site, but allow for tentative identification
that could be followed with injection of an actual pesticide standard.
(3) System must come with a library on CD-ROM, that contains relative
retention time and mass spectral data for over 500 pesticides,
endocrine disruptors and their metabolites. This library must be built
by actual runs on an instrument of the type described herein. The
system must be capable of identifying multiple pesticides in a single
analysis because multiple pesticides are used in a single growing
season, in limited geographic areas in most agricultural applications.
This will also save time by allowing multiple analyses in a single run
and conserve sample which may be of limited quantity. (4) Autoinjector
parameters must be controlled through the data system; the injector
shall not require gases, regulators, tubing or pneumatics for operation
and must be temperature controlled with the ability to be heated or
cooled to a constant temperature. Due to the volatility of some of the
anticipated analytes or of the extraction solvents, a fast injection
(100 milliseconds) is required to prevent sample decomposition or loss.
(5) The data system must be capable of simultaneous collection and data
handling from Flame Ionization Detector, Nitrogen Phosphorus Detector,
and Mass Spectrometry Analytical System Detector. This will allow for
multiple detector capability to aid in the identification of
components or materials in extremely complex environmental mixtures.
The data system must also allow automated control of all operating
parameters, including the electronic pressure control of the Gas
Chromatograph/Mass Spectrometry Detection system. (6) The system must
include test columns, gas purifying cartridges, and tubing neccesary
for installation and performance testing. Also included will be two
analytical columns of the same phase used to originate the CD-ROM
libraries described herein. All parts and labor for installation of
GC/MS system must be included. The Request for Quotation (RFQ) will be
released on 3/29/99 with a due date for quotations of 4/12/99. This
acquisition will be performed according to Simplified Acquisition
Procedures (FAR Part 13). Requests for a copy of the RFP can be sent
via facsimile to (513) 487-2109. All requests should include a complete
mailing address. Posted 03/15/99 (W-SN308597). (0074) Loren Data Corp. http://www.ld.com (SYN# 0254 19990317\66-0003.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|