|
COMMERCE BUSINESS DAILY ISSUE OF MARCH 17,1999 PSA#2304U.S. Environmental Protection Agency, 26 W. Martin Luther King Dr.,
Cincinnati, OH 45268 99 -- X-RAY FLUORESCENCE SPECTROMETER SOL RFQ-OH-99-00096 POC Sharon
Clement (513) 487-2030 WEB: Office of Acquisition Management --
Cincinnati Contracts, http://www.epa.gov/oam/cinn_cmd. E-MAIL: Sharon
Clement -- Cincinnati Contracts Management,
clement.sharon@epamail.epa.gov. The Environmental Protection Agency has
a requirement for a research grade, wavelength dispersive, sequential
X-ray Fluorescence Spectrometer to be used for the analysis of
environmental air pollutants. Specifically the spectrometer will
perform qualitative, semi-quantitative and quantitative measurement of
elements of atomic number range 6 through 92 with sensitivities at the
parts per million (ppm) level and with appropriate relative accuracies
pf 10 to 50%. The spectrometer software must perform state-of-the-art
"standardless" analysis of unknown specimens such as particulate
captured on filters from the sampling of polluted air for which no
standards exist. The Government's minimum requirements for this system
are as follows: (1) 4 kilowatt x-ray tube with inverted geometry so
that the sample is irradiated from below with an Rh anode and an end
window thickness of not more than 75 microns; (2) stable, medium
frequency, high-voltage generator sufficient to operate the x-ray tube
in a range from 20 to 60 kilovolts and from 5 to at least 120
milliamperes; (3) programmable beam filter system of not less than 3
positions with a filter configuration optimized for rhodium radiation;
(4) programmable collimator changer with at least four positions
fitted with at least three collimators of the general description:
fine, coarse, and extra coarse, the exact size of which will depend
upon the geometry and optical system of the spectrometer selected; (5)
at least an eight-position crystal changer fitted with at least the
following 6 crystals and multilayer materials upon installation: a)LiF
200 for the analysis of potassium to uranium, b)LiF 220 for the high
resolution analysis of vanadium to uranium; c)pentaerythrite (PET)
crystal for the analysis of aluminum to chlorine, d)germanium 111
crystal for the analysis of phosphorus to chlorine, and e)two
multilayer materials of sufficient d-spacing to cover the range of
analysis of the light elements carbon through fluorine with the
required sensitivity. The crystals must be temperature-controlled in
the spectrometer chamber to 0.1 C; (6) a goniometer system for required
positioning of the crystals and the detectors for analysis with a
2-theta angular range of not less than 4-148 degrees, a fast speed of
angular positioning, a positional accuracy of not less than 0.015
degrees, and a long-term reliable positional reproducibility of not
less than 0.0002. contained in a temperature-controlled vacuum chamber
with supplied vacuum systems and the spectrometer cabinet temperature
controlled to not less than 0.05 C; (7) a spectrometer detection
system consisting of a flow proportional counter and a scintillation
counter configured so that the required elements may be detected with
specified accuracies, the flow counter accompanied by at least 5
replacement windows; (8) a sample handling system of approximately 50
positions, specimen holder opening sizes of about 23mm, 29 mm, or 32 mm
including at least one specimen holder that holds small specimens in
the correct configuration forcounting. The number of specimen holders
required is not less than the number of sampling positions. The
spectrometer shall have specimen spinning capability during analysis;
(9) multichannel analyzer counting electronics with deadtime
correction. The counting electronics shall ensure linearity of response
of up to 1.5 million counts per second for the scintillation counter
and up to 2 million counts per second for the flow proportional
counter; (10) a helium system for liquid analysis; (11) a heat
exchanger/water chiller cooling system for the x-ray tube and the
spectrometer; (12) protected by integrated safety circuitry to monitor
the status of the instrument and shut it down if the limits of any of
its components are exceeded; (13) hardware and software with
multitasking capability to operate and monitor the spectrometer, to
diagnose problems including remote diagnosis of problems with the
spectrometer via modem, adequate to handle the requirements for
multiple and simultaneous functions, and to provide state-of-the-art
capability for spectrometer set-up for data collection and for
qualitative and quantitative analysis operating in a Windows 95 (or 98)
or a Windows NT environment; (14) state-of-the-art analytical software
to perform evaluations of data for qualitative, semiquantitative, and
quantitative analysis using multivariable regression analysis
accompanied by suitable mathematical correction models to minimize the
influence of interfering elements in multicomponent matrices. The
software must also be able to matrix-correct the intensities using
fundamental parameters; (15) analytical software to perform
quantitative analysis for specimens for which no standards exist
capable of selecting optimal conditions for the elements to be
analyzed, collecting the data, and providing a quantitative analysis
with an accuracy and precision that is state-of-the-art. The
standardless analysis software must be Uniquant, or, at minimum,
provide the performance of Uniquant for analysis of unknown specimens;
(16) equipped with calibration standards and calibrated to analyze the
following elements in the following ranges in pressed pellets in a high
calcium matrix (typical of cement kiln dust): (a) P, Mn, Cu: <0.1%,
(b) Na, Mg, Ti, Sr <1.0%, (c) Al, Si, S, Cl, K, Fe <10.0%, (d)
Ca and O <50.0%. Additionally, the purchase shall include
installation including UL listing and labeling satisfactory to the
North Carolina Building Code Council, on-site satisfactory calibration,
on-site comprehensive training, 2 copies of instrument documentation
and training materials, an 18 month warranty on the instrument, a
2-year non-prorated warranty on the x-ray tube, and timely on-site
service. The hardware and software supplied must be year 2000 compliant
and supported with service, parts, and upgrades for at least 10 years
after purchase. THIS IS NOT A REQUEST FOR PROPOSALS. This acquisition
will be accomplished under FAR Par 12, Acquisition of Commercial Items.
Only sources capable of fully meeting the unique technical performance
specifications/capabilities outlined above, without exception, will be
considered. Technical documentation must be submitted on or before
March 31, 1999 to the office identified above. Posted 03/15/99
(W-SN308430). (0074) Loren Data Corp. http://www.ld.com (SYN# 0295 19990317\99-0005.SOL)
99 - Miscellaneous Index Page
|
|