COMMERCE BUSINESS DAILY ISSUE OF APRIL 15,1999 PSA#2325USDA-ARS-MWA, Contracting Office, 1815 N. University Street, Peoria, IL
61604 66 -- BRUKER AVANCE CONSOLE -- NMR UPGRADE SOL 5114-9-3113 DUE 042999
POC Lori J. Houghton, Contracting Officer, 309-681-6616 The U.S.
Department of Agriculture, Agricultural Research Service, Midwest Area
has a requirement to purchase a High Performance Digital NMR
Spectrometer Console; an AVANCE 400 Console in exchange for an MSL 300
console (less probes and high power rack) and an ARX 400 console (less
probes and BVT-2000, VT dewar/transfer lines). The above is an upgrade
to the existing Bruker NMR spectrometers at the National Center for
Agricultural Research in Peoria, IL. This is a combined
synopsis/solicitation for commercial items prepared in accordance with
the format in FAR Subpart 12.6, as supplemented with additional
information included in this notice. This announcement constitutes the
only solicitation; proposals are being requested and a written
solicitation will not be issued. Solicitation No. 5114-9-3113 is being
issued as a Request for Proposal. This solicitation document and
incorporated provisions and clauses are those in effect through Federal
Acquisition Circular 97-11. The solicitation is a sole source
procurement to Bruker Instruments, Inc. Standard Industrial
Classification Code 3826 with a corresponding size standard of 500
employees applies to qualify as a small business. Offerors are to
propose prices for the following: CLIN 0001 Digital NMR Spectrometer
Console, 1 each. The salient characteristics that the AVANCE 400
Console possesses are as follows: Magnet Control System: Bruker
Orthogonal Shim System (BOSS), with 20 shim gradients, low current and
low heat dissipation design for use with 400 Mhz standard bore magnet.
Room temperature shim system with 17 shim gradients for use with 300
Mhz wide bore magnet, Bruker Sample Transport (BST) upper shim stack
enables high flow rates for vt gas and includes integral sensors for
Bruker Smart Magnet System in exchange for existing shim stack on both
400 Mhz standard bore and 300 Mhz wide bore magnets. Bruker Smart
Magnet System (BSMS)and Digital Lock control unit including, digital
lock frequency generation, digital quadrature lock receiver, fast-field
adjustments with sample-and-hold, flexible computer-controlled lock
regulation system, Shim Control Boards (SCBs) with ultra-stable high
precision, low noise shim current sources, lock transmitter board for
the 300 Mhz wide bore magnet and lock receiver board for the 300 Mhz
wide bore magnet. BSMS keyboard control unit with twenty meter cables,
for full manual and computer control of lock, sample spinning and room
temperature shims. Provides extensive flexibility for customization of
shim ranges, magnetic field offset, sample lift rate and many other
parameters. Spectrometer Cabinet: Double-width 19" rack mounted cabinet
with rf shielding and full front access to all sub-assemblies and
cabling. RF Section: Multilink preamplifier for use with 400 Mhz high
resolution and CPMAS probes, including: H1 preamplifier [BH0240], D2
preamplifier for lock and D2 observe [BH0243], BB preamplifier
[BH0249], Control unit and Built-in tune/match display. Multilink
preamplifier for use with 300 Mhz wide bore high resolution and CPMAS
probes including: H1 preamplifier, D2 preamplifier for lock and D2
observe, BB amplifier, Control unit and Built-in tune/match display.
ATR Quadrature detection receiving system with high dynamic range 22
Mhz intermediate frequency and inverse capability. Router for automated
high speed configuration of RF signals. Frequency synthesizers. Digital
Acquisition Control System: Communication Control Unit (CCU) with 100
Mhz RISC microprocessor, 16 Mbyte buffer memory, dedicated Ethernet
link (10 Mbit/sec) and RS-232 connection for communications between the
spectrometer and the host computer. Timing Control Unit (TCU),
ultra-fast VME pulse programmer, provides exact timing and full
flexibility in data acquisition with 12.5 nanosec resolution. Based on
Intel I960 RISC co-processor, the TCU allows control of up to 8 FCUs
using a master clock frequency of 80 Mhz. Frequency Control Unit (FCU)
for the F1 channel, interfaced to the VME Bus, for generation of
phase, amplitude, and frequency using Direct Digital Synthesis (DDS).
Provides 0.05 degree phase settability and 0.001 Hz frequency
resolution. Includes 64 Kword (256 Kbyte) wave form memory (256 kW
optional) for programming of experimental sequences (including pulse
shaping and composite pulse decoupling) that are executable
synchronously and asynchronously. Includes Amplitude Shaping Unit
(ASU), for gating and rf amplitude shaping control of the rf output
[BH2070]. Frequency Control Unit (FCU) for the F2 channel, interfaced
to the VME Bus, for the generation of phase, amplitude, and frequency
using Direct Digital Synthesis (DDS). Provides 0.05 degree phase
settability and 0.001 Hz frequency resolution. Includes 64 Kword (256
Kbyte) wave form memory (256 kW optional) for programming of
experimental sequences (including pulse shaping and composite pulse
decoupling) that are executable synchronously and asynchronously.
Includes Amplitude Shaping Unit (ASU), for gating and rf amplitude
shaping control of the rf output [BH2070]. Receiver Control Unit (RCU),
interfaced to the VME Bus for acquisition of NMR data with over
sampling and digital filtering. Provides >16 bit effective dynamic
range for typical proton measurement conditions. Includes Intel I960
co-processor and buffer memory for flexible real-time data management
and high throughput acquisition. 16 bit digitizer allowing simultaneous
acquisition with up to 2 Mhz sampling rate. Industry Standard
Windows/NT Host Workstation. Bruker NMR Suite software package for
acquisition and processing of 1D, 2D and 3D NMR data, including
XWINNMR, XWINPLOT, ICON-NMR, NMR-Check, and NMR-SIM. Includes media,
on-line hypertext manual and single node-locked license. Also includes
printed manuals for XWINNMR general operation, XWINNMR pulse
programming, XWINPLOT and ICON-NMR. Probes and other features: 400 Mhz
standard bore inverse broadband (BBO) probe for use in 400 Mhz magnet
(5 mm sample diameter), H1 observe with broadband decoupling coil
tunable from Ag109 to P31 [PH0010]. GRASP II version, including
actively shielded Z-gradient coil for use with GRASP II, VT range -150
to +180 C. Bruker engineer will install the customer's existing low
temperature accessory (dewar and transfer line) to ensure compatibility
with the AVANCE console. The Bruker engineer will install the customers
existing Bruker manufactured 5mm C13/H1 dual and 5mm H1/C13 inverse
probe and demonstrate lineshape, H1/C13 90's and decoupling powers to
a best effort on the 400 Mhz magnet only. Level I Solids Accessory for
the AVANCE series, suitable for CP/MAS or PE with addition of probe.
Includes Amplifier upgrade [BH2090], in lieu of standard BLAXH 100/50,
including BLARH 100 and BLAX-300 as described in the following items.
These amplifiers are appropriate for both high resolution and CPMAS
applications. BLARH 100, H1/F19 linear Array Amplifier for H1/F19
observe and homo/heteronuclear H1/F19 decoupling; 100 Watt linear pulse
power, 10 Watt linear CW, 0.8 Watt ultra linear stage, all operations
under computer control with total dynamic excitation range >120dB.
BLAX 300 RS linear broadband amplifier for X-nucleus pulse and decouple
with ultra-fast blanking and rise/fall times; 300 Watt pulse power, 30
Watt linear CW, all operations under computer control. Microprocessor
based MAS control unit with spinning rate counter, precision
regulating valves, pressure stabilization and sample eject. May be
controlled from the front panel or directly by the system computer.
NMR-Suite solids analysis software upgrade. Support and Installation:
two tuitions to Bruker training courses. Set of system documentation
and manuals provided in BASH CD ROM format. Complete console
installation provided by qualified personnel. Standard samples will be
provided on a temporary basis for meeting specifications. Accessories:
GRASP II Gradient Spectroscopy accessory for the AVANCE series.
Includes single channel GAB ultra-stable power supply, integrated into
the BSMS, with single channel gradient amplifier (10 Amp) for
generation of gradients with up to 50 Gauss/cm. Includes GCU but
requires Bruker GRASP II probe with built-in gradient coil. Deuterium
Lock Switch for RF deuterium decoupling with lock switching. Also used
for H2 Z gradient shimming [BII003]. FAR 52.212-1, Instructions to
Offerors -- Commercial, FAR 52.212-2, Evaluation -- Commercial Items;
FAR 52.212-3 Offeror Representations and Certifications -- Commercial
Items, FAR 52.212-4 Contract Terms and Conditions -- Commercial Items,
and FAR 52.212-5 Contract Terms and Conditions Required to Implement
Statutes or Executive Orders -- Commercial Items apply to this
acquisition. Proposals will be evaluated on price, past performance,
and proof of meeting the salient characteristics listed above. Offers
should include a list of at least four references to include company
name, point of contact, phone number and fax number. References should
reflect similar type of work or equipment purchases. Offers to include
a commercial warranty for equipment. Payment will be made 30 calendar
days after acceptance of equipment. Equipment willbe shipped FOB
Destination to: National Center for Agriculture Utilization Research,
1815 North University Street, Peoria, IL 61604. Delivery of equipment
desired to be within 3-6 months after receipt of award. Offer must
include a 30 calendar day acceptance period. Offers are to be received
Not Later than the Close of Business (4:00 p.m. CST), April 29, 1999
at USDA-ARS-MWA, Contracting Office, 1815 N. University Street, Peoria,
IL 61604. The POC for this acquisition is Lori J. Houghton, (309)
681-6616, or E-Mail: houghtol@ars.usda.gov, or FAX: (309) 681-6680.
Faxed Proposals will not be accepted. Posted 04/13/99 (W-SN318757).
(0103) Loren Data Corp. http://www.ld.com (SYN# 0608 19990415\66-0015.SOL)