|
COMMERCE BUSINESS DAILY ISSUE OF JULY 19,1999 PSA#2390Bechtel Nevada, Mail Stop NLV 018, Post Office Box 98521, Las Vegas, NV
89193-8521 66 -- CONTINUOUS X-RAY SOURCE, INCLUDING POWER SUPPLY AND COOLING
SYSTEM SOL BW002-DJ-99 DUE 072899 POC Dennis J. Jeffrey, (702) 295-2192
E-MAIL: Click here to contact the contracting officer via,
jeffredj@nv.doe.gov. CONTINUOUS X-RAY SOURCE Continuous X-ray source,
including power supply and cooling system (if required) which supplies
the following: REQUIREMENTS: Signal level: Unit must generate
sufficient x-ray flux to ensure that 109 x-rays per second pass through
an aperture 1/4 x 1/8 in located a minimum of six feet from the x-ray
head. It is preferred that the x-ray beam be focused or at least
collimated rather than omni-directional. X-ray energy spectrum:
Molybdenum has proven successful in a similar application. We must be
able to observe significant attenuation by 1" thick plastic (Lexan) as
it eclipses the beam transmitted through the aperture. Ripple:
Variation in the x-ray flux must be < 1.0% of average signal level
up to frequencies in excess of 10 MHz. This is essential to ensure
unambiguous definition of the time when the signal is eclipsed.
Continuous Operation: Unit must be capable of being operated
continuously for at least 20 min. Physical Constraints: X-ray head must
be separable from power supply by at least 6 ft. Prefer separation >
10 ft if possible. Power supply will be installed at floor level; x-ray
head will be approx eight to 10 ft above floor. Power Requirements:
Both 110 and 220-V power are available. DESIRED: X-ray Energy Spectrum:
Would appreciate the option to change tubes to optimize absorption by
the target. In the future, it may be determined that a Mo tube is not
optimum for the application. Physical Constraints: X-ray head will be
mounted on a vacuum- and pressure-sealed column which is four- to
six-inches in diameter. Would prefer flange on x-ray head to facilitate
mounting such that the x-ray beam is directed axially through the
mounting column. Beam power monitor: Would appreciate the ability to
monitor the power emitted by the x-ray head independent of the detector
being used to measure transient absorption by the target. Posted
07/14/99 (W-SN353778). (0195) Loren Data Corp. http://www.ld.com (SYN# 0812 19990719\66-0013.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|