COMMERCE BUSINESS DAILY ISSUE OF SEPTEMBER 2,1999 PSA#2424U.S. Environmental Protection Agency, 26 W. Martin Luther King Dr.,
Cincinnati, OH 45268 66 -- HEATED FLAME IONIZATION DETECTOR SOL 99-13 POC Pam Legare
E-MAIL: U.S. EPA -- Cincinnati Contracts Management Division,
legare.pamela@epamail.epa.gov. The U.S. Environmental Protection Agency
(EPA) seeks qualified sources to provide a Heated Flame Ionization
Detector (HFID) for EPA's National Vehicle and Fuel Emissions
Laboratory(NVFEL) Testing Services Division. The electrical output of
the HFID used for measuring hydrocarbons (or hydrocarbons plus
methanol) shall be recorded onto a data acquisition system. The HFID
must be functionally capable to accept AK protocol commands over an
RS-232 communications bus. It must be equipped on the back with a 9 pin
female DB connector for the RS-232 communication, a 3 pin male Molex
connector for power, a 3 pin female Molex connector for heated FID line
power, a 3 pin female Mini DIN connector for the heated FID
thermocouple, a 9 pin female DB connector and cable to connect to the
individual sample valve solenoids with connectors that match the
individual sample solenoids, and the HFID shall be delivered with the
ability to respond to commands from a host computer without any user
modifications. The HFID shall be mounted on a 19" rack. There shall be
a port (1/4" Swagelok or 6mm bulkhead connectors or equivalent) on the
HFID for span gases (4), FID fuel (1), zero air (1), and sample (1).
The HFID shall be automated such that it can be zeroed and spanned
during testing through the commands given on the RS-232 bus. The HFID
system shall have a minimum warranty of 12 months after commissioning
and shall guarantee the components and specified functions and include
labor costs for any work required under the warranty. The analyzer
shall be supplied with complete operational manuals, maintenance
manuals, and system electrical and pneumatic schematics. Delivery shall
be within 180 days after award. Interested parties should submit a
written request to Pam Legare at the above address, by fax
(513)487-2107 or by email (legare.pamela@epamail.epa.gov). TELEPHONE
REQUESTS WILL NOT BE HONORED. The Request for Quotation is scheduled to
be issued fifteen days after the date this synopsis is published.
Posted 08/31/99 (W-SN374576). (0243) Loren Data Corp. http://www.ld.com (SYN# 0316 19990902\66-0016.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|