COMMERCE BUSINESS DAILY ISSUE OF SEPTEMBER 10,1999 PSA#2429Commander, U.S. Army Aviation and Missile Command, Acquisition Center,
Research Development and Engineering Center, Redstone Arsenal, AL
35898-5000 A -- BAIRD ONESPARK CHARGE INJECTION DEVICE (CID) SOLID STATE
SPECTROPRAPH SYSTEM SOL PAN RG132-9 DUE 091599 POC Deidra Barnes,
Contract Specialist, Phone (256) 876-3139 or Harold Smith, Contracting
Officer, Phone (256) 876-0908 E-MAIL: click here to contact Deidra
Barnes, Contract Specialist, barnes-dd@redstone.army.mil. One (1) BAIRD
OneSpark Charge Injection Device (CID) Solid State Spectrometer System
as detailed below: 1. Spectrograph: The emission spectrograph shall
have the following minimum capabilities: a. Nitrogen or argon purged
optical path with an echelle grating and prism cross-disperser to
produce a two-dimensional emission spectrum. b. A wavelength range of
174-800nm. c. A minimum resolution of 0.004nm at a wavelength of 200nm
and 0.008nm at a wavelength of 400nm. d. Software controlled HR-400
Source Unit capable of operating at 200 or 400 Hz. e. Charge Injection
Device (CID) solid state detector system capable of operating in both
Random Access Integration (RAI) and Time Domain Spectroscopy (TDS)
modes. f. Detector cooling unit to maintain low detector dark current.
g. Ability to conduct simultaneous background corrections. h. Ability
to simultaneous analyze all element in a sample. i. Ability to
simultaneously conduct background data immediately adjacent to the
analytical wavelengths of interest. j. Ability to create a complete
spectral image over the wavelength range of the spectrograph, allowing
the visual identification of the specific location of elements in a
sample. 2. Software: System control and storage of generated data shall
be accomplished through a PC interface. Control software shall be
Microsoft Windows 98 compatible and shall as a minimum be capable of
the following: a. Performs quantitative analyses at any wavelength. b.
Automatic or manual selection of background points. c. Controls gas
flow and spark-source parameters. d. Utilizes unlimited number of
points per calibration curve. e. Simultaneously corrects the entire
emission spectrum of a sample. f. Provides full wavelength and
identification tools. g. Allows identification of spectral interference
lines and interelement corrections. h. Produces data output compatible
with Microsoft Word, Excel, and Access. 3. Computer Data System: The
emission spectrograph system shall include a computer data system with
the following minimum requirements: Motherboard, Processor, Cache,
Memory, Hard Drive, CD-ROM, Video Controller, Keyboard, Mouse, Sound,
Speakers, Ports, O/S, Printer, Monitor, 66MHz, 300MHz (Intel Pentium
Celeron), 128KB minimum, 32MB SDRAM, 4.3GB, 32X, 4MB SGRAM, Standard
Windows 95/98 104 Keys, 2 Buttons PS/2, SoundBlaster 16 Compatible, 2
Speakers, 2 Serial, 1 Parallel, Windows 98 CD, 1 Printer, 17" color. 4.
Voltage Regulator: The emission spectrograph system shall be provided
with a voltage regulator capable of maintaining the constant voltage
necessary for accurate, repeatable analysis. 5. Calibration:
Calibration shall be provided for three alloy sets. The alloy sets
shall consist of three different matrices: Aluminum, iron, and nickel.
6. Installation, Training, and Warranty: The purchase shall include
installation, minimum one-year warranty, and on-site training for two
operators at BAIRD's training facility, Franklin, MA. 7. Delivery:
Delivery shall be made to the following address not to exceed 120 days
from receipt of order: U.S. Army Aviation and Missile Command,
AMSAM-RD-PS-CM, Bldg. 7103, Redstone Arsenal, AL 35898. See Numbered
Note 22. Posted 09/08/99 (W-SN377541). (0251) Loren Data Corp. http://www.ld.com (SYN# 0005 19990910\A-0005.SOL)
A - Research and Development Index Page
|
|