VA Medical Center, Attn: Contracting (P4 CONT), 4th Plain Blvd. & St. Johns Road, Vancouver, WA 98661

65 -- REVERSE OSMOSIS PROCESSING UNIT SOL 648-67-99 DUE 092499 POC Julie M. Files 360-905 -- 1781 The subject combined synopsis soliciation is modified to reflect there is no installaiton required. Requirement is for hardware only, water production is to allow for 18,000 gallons of water to be produced per unit per day. Operating voltage is to be 480V. The R/O panel alarm indicators must minimally include: low inlet pressure; high/low Ph; low product flow; low % rejection; high membrane differential pressure; high permeate pressure; high temp; high/low primary pressure; high permeate conductivity; and high pre-filter differential pressure. R/O unit manufacturer must have authorized local service personnel available 365 days per year to provide 8 hour response time during warranty period. No other changes are being made at htis time. Posted 09/15/99 (W-SN380803). (0258)