COMMERCE BUSINESS DAILY ISSUE OF OCTOBER 5,1999 PSA#2448USDI, Bureau of Reclamation, 125 South State Street, Room 6107, Salt
Lake City, Utah 84138-1102. 66 -- XRF SPECTRUM ANALYZER NITON XL-722Q DUAL-SOURCE BULK SAMPLE
ANALYZER OR EQUAL SOL 00-SQ-49-0001 DUE 101299 POC TECHNICAL CONTACT:
Linda Calton (801) 379-1273; PURCHASING AGENT, Nancy Biddinger, phone
(801) 379-1119, fax (801) 379-1159, e-mail nbiddinger@uc.usbr.gov WEB:
Acquisition Office, Upper Colorado Region, Bureau of,
http://ideasec.usgs.gov. E-MAIL: Nancy Biddinger,
nbiddinger@uc.usbr.gov. Niton XL-722Q Dual-Source Bulk Sample Analyzer
or equal 888-006R 10 mCiCd-109, 14 mCi Am-241. DISPLAY: System must
display complete x-ray spectrum from 3 KeV -- 88 KeV for each reading
with sufficient resolution displaying spectrum conclusively confirms
the presence of absence of significant quantities of RCRA 8 metals
including lead, arsenic, selenium, mercury, cadmium, silver and barium.
AUTOMATED DATA ENTRY: Automated, barcode-based, data-entry system for
test site and sample information. Data-entry system must accommodate
user-defined inputs and be fully integrated with the system downloading
and reporting software. RADIOACTIVE SOURCE: Use a cadmium-109 (109Cd)
and americium-241 (241Am) for bulk sample analysis. RADIATION:
Radiation Exposure: Radiation on user's hand with safety shutter open
during reading not to exceed 0.1 mR/hour. Travel and Shipping
Restrictions: Must not require radiation label on carrying case.
Conforms to 49CFR 173.421 for "excepted package", instruments and
articles, N.O.S. UN2910. WEIGHT AND SIZE: Single, self-contained unit,
weighing no more than 3 pounds, including batteries. Exterior
dimensions must not exceed 9"x3"x2". Comes equipped with belt holster
for safe transport into soil excavation sites. ADDITIONAL FEATURES: 1.
Auto-calibration: System must always self-calibrate before testing is
enabled. 2. Automatic Detector Cooling: System must electrically
regulate the temperature of its x-ray detector to prevent inconsistent
and/or highly variable results caused by shifts in ambient
temperature. 3. Reading Speed ad a Function of Source-Age: System must
self-adjust reading times for age of radioisotope source. Reading
times (in real, or "live", seconds) with one-year-old (12-month-old)
source must not exceed twice the system reading times with a new
source. 4. Utilize a thermo-electrically cooled Silicon PIN dioide
detector with resolution less than 350eV at the 5.95keV Mn K-line. 5.
System must utilize NiMH battery packs which provide at least 8 hours
of continual use on afull charge. BULK SAMPLE ANALYSIS: System must
analyze prepared soil samples in 60 seconds with a lead-in-soil
detection limit of 20 ppm. Capable of demonstrating ELPAT proficiency,
and must meet the instrument specification for EPA method 6200: Field
Portable X-Ray Fluorescence Spectrometry For the Determination of
Elemental Concentration in Soil and Sediment. System must utilize
Compton Normalization technique for metal quantification in bulk
samples. System must not require matrix-specific calibration or
standardization. System accessory kit must include certified NIST soil
standards for quality control and system accuracy verification,
including NIST 2709, NIST 2710, and NIST 2711 standards. THIN SAMPLE
ANALYSIS: System must be able to be upgraded to include Thin Sample
Analysis capabilities, including ambient air filter, personal exposure
monitoring air filter, and dust wipe analysis modes. This is a
combined synopsis/solicitation for commercial items. This announcement
constitutes the only solicitation; proposals are being requested and
a written solicitation will not be issued. Please visit our electronic
commerce website: http://ideasec.usgs.gov. If you register with us,
you can view posted solicitations and awards and submit quotes on line.
All vendors with access to the internet and an e-mail address can
receive electronic notification of awards. All vendors with access to
the internet and an e-mail address can receive electronic notification
of awards. Vendors can also be sent an e-mail notification of any
solicitation directed to them. Numbered Notes (Footnotes about set
asides, sole source etc.) Posted 10/01/99 (W-SN387761). (0274) Loren Data Corp. http://www.ld.com (SYN# 0160 19991005\66-0008.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|