COMMERCE BUSINESS DAILY ISSUE OF FEBRUARY 9,2000 PSA#2533NASA/John F. Kennedy Space Center, Procurement, Kennedy Space Center,
FL 32899 66 -- HUMIDITY GENERATOR SYSTEM SOL S-1 DUE 022200 POC Sandra L.
Wheeler, Purchasing Agent, Phone (321)867-7869, Fax (321)867-7202,
Email sandra.wheeler-1@kmail.ksc.nasa.gov WEB: Click here for the
latest information about this notice,
http://nais.msfc.nasa.gov/cgi-bin/EPS/bizops.cgi?gr=D&pin=76#S-1.
E-MAIL: Sandra L. Wheeler, sandra.wheeler-1@kmail.ksc.nasa.gov. SPACE
GATEWAY SUPPORT, POC SANDRA WHEELER, 321-867-7869, KENNEDY SPACE
CENTER, FL, INTENDS TO PURCHASE A THUNDER SCIENTIFIC CORPORATION OR
EQUAL AUTOMATED TWO PRESSURE HUMIDITY GENERATING SYSTEM MODEL 9000 OR
EQUAL. THE SYSTEM MUST CONTAIN THE FOLLOWING SALIENT CHARACTERISTICS:
1-RELATIVE HUMIDITY RANGE FROM 5 TO 99%RH; 2-RELATIVE HUMIDITY
RESOLUTION, MIN. 0.01%RH; 3-RELATIVE HUMIDITY ACCURACY OF +OR-0.3%RH OR
BETTER; 4-BATH TEMPERATURE RANGE OF 0 TO 70 DEGREES C; 5-BATH TEMP.
MEASUREMENT RESOLUTION OF 0.005 DEGREES C; 6-BATH TEMP. CONTROL
STABILITY OF +OR-0.02 DEGREES C; 7-BATH TEMPERATURE UNFIROMITY OF 0.04
DEGREES C; 8-BATH TEMP. MEASUREMENT ACCURACY OF +OR-0.038 DEGREES C;
9-OPERATION MUST BE EITHER AIR OR NITROGEN; 10-TEST CHAMBER DIMENSIONS
BE A MIN. OF 12 INX12 INX12 IN WITH AN ACCESS OPENING THROUGH A FULL
REMOVABLE COVER WITH TWO CABLE PORTS OF AT LEAST 3.5 INCHES IN
DIAMETER. IF YOU HAVE A PRODUCT THAT MEETS THESE SPECIFICATIONS, PLEASE
SEND TECHNICAL DATA ONLY TO: ATTN: SANDRA WHEELER, DEPT.SGS-300,
KENNEDY SPACE CENTER, FL 32899. ALL TECHNICAL DATA MUST BE RECEIVED BY
2-22-00 TO BE CONSIDERED. Posted 02/07/00 (D-SN422703). (0038) Loren Data Corp. http://www.ld.com (SYN# 0262 20000209\66-0003.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|