|
COMMERCE BUSINESS DAILY ISSUE OF FEBRUARY 28,2000 PSA#2545Commander, U. S. Army Aviation & Missile Command, Acquisition Center,
Buiding 5303, Redstone Arsenal, AL 35898-5280 66 -- LIQUID FLOWMETER CALIBRATION SYSTEM (PART 1) SOL
DAAH01-00-R-0262 DUE 032000 POC Betty Carden, 256-842-7223 or Debra K.
Long, 256-876-4153 THIS IS THE FIRST PART OF THE DESCRIPITION FOR THIS
ACQUISITION. THE DESCRIPTION IS CONTINUED IN A NOTICE, SUBJECT: 66 --
LIQUID FLOWMETER CALIBRATION SYSTEM (PART 2). -- THIS IS A COMBINED
SYNOPSIS/SOLICITATION. SOLICITATION NUMBER: DAAH01-00-R-0262, NOUN:
LIQUID FLOWMETER (R078-00), PRON: B003HK21D2, NSN: 0000 00 000 0000,
CLOSING DATE: 20MAR00. This Request for Proposal (RFP) is for
commercial items prepared in accordance with the format in Federal
Acquisition Regulation (FAR) Subpart 12.6, as supplemented with
additional information included in this notice. This announcement
constitutes the only solicitation. This RFP and incorporated provisions
and clauses are those in effect through Federal Acquisition Circular
(FAC) 97-15. This RFP is competitive, restricted to small businesses.
The Standard Industry Code (SIC) is 3829, number of employees is 500.
The requirement is as follows: General: A positive displacement liquid
flow calibrator(LFC), a volumetric type measurement device, producing
and measuring the exact volume of fluid passing through a flowmeter as
a function of time while compensating for fluid viscosity, density, and
temperature. The LFC is to be used as a Primary Standard. The LFC is to
include a precision honed and polished piston and cylinder assembly
mechanically connected with a shaft to a linear encoder. The encoder
side of the piston shall be pressurized with air to provide the force
to move the piston. Calibration fluid will be loaded into the opposite
side of the piston, the meter under test, and a collection reservoir
tank hydraulically located after a manifold of flow control valves,
plumbing, etc. Air pressure should be supplied to the piston through
control valves and pressure regulators. Once the pressure is applied,
pressurizing the fluid, flow is created by opening one or more flow
control valves on the downstream side to adjust the desired flow rate.
Once the rate is established, data is acquired by the data acquisition
and control system and stored on a disk mediumfor later evaluation.
When the piston reaches the end of the cylinder, the flow stops. The
air pressure on the encoder side of the piston is vented and air
pressure is supplied to the collecting reservoir to provide motive
force to return the piston to the starting point. As the fluid returns
to the cylinder, from the reservoir, it is directed through a filter
to remove any minute particles that may be present in the line. A fluid
diverting check valve prevents this fluid from passing backward through
the meter under test. When the piston reaches the starting position,
the reservoir is vented and the piston is repressurized to restart the
process. SPECIFIC REQUIREMENTS FOR LIQUID FLOW CALIBRATOR(LFC): The
LFC will operate over a range of 0.5 to 300 gpm using hydrocarbon
fluids such as MIL-H-5606 and MIL-H-7024D. The uncertainty of the
generated flow rate will be 0.05% of reading or better at K=2(95%
Confidence Level). The repeatability will be 0.02% of reading or
better. The required viscosity range is 0.5 to 100 centiStokes within
the temperature range of 60 to 120 F. The LFC is to operate with an
existing air pressure supply of 100 PSIG at 20 CFM. If the delivery
volume is insufficient, the manufacturer will supply an air system,
which will provide the required air capacity for operating the LFC. The
LFC will be capable of operating at any ambient temperature from 60 to
100 F with no degradation in performance. The LFC temperature
conditioning system(TCS) will be capable of conditioning and
maintaining the temperature of the calibration media within 2.5 F over
a range of 60 to 120 F at flow rates over the range of 0.5 to 300 gpm.
The TCS will include all of the components required (chiller, heaters,
storage vessels, valves, plumbing, heat exchangers, control cables,
control, etc.) to achieve the required operation over the operating
ranges. Transient response to temperature changes will be stabilized
within 2.5 F within 60 minutes when changing from room temperature to
either operating extreme; and within 120 minutes when changing from one
extreme to the other in either direction (heating or cooling). The
chiller and other components, to be supplied and installed by
manufacturer, will be mounted outside of the calibration laboratory and
will not require environmental cover. The cooling media must not freeze
under out-door ambient temperatures of -10 F. They will be located
within 150 feet of the LFC with appropriate power and other services in
close proximity. The storage vessel for the conditioning fluid will be
insulated to limit the media temperature change to 5.0 per 8 hours
when the ambient changes from 35 to 100 F. The LFC will be housed in a
single enclosure. The data acquisition and control system will be
located on a separate console. The enclosure shall be moveable by
conventional means to the final location. The LFC does not have to be
rigidly mounted to the floor, but may be rigidly mounted if desired.
All power wiring will be in conduit with explosion proof junction boxes
and terminations. Instrumentation cables of low power and voltage will
be routed in other appropriate conduits and cableways. All components
of the LFC data acquisition and control will be housed in a single
mobile cart with 10 to 20 feet of shielded cable between the LFC and
the cart. The cart will have at least two drawers with minimum internal
dimensions of 16 inches wide, 5 inches high, and 18 inches deep. All
interconnecting computer cables will be shielded, and will enable the
LFC to be operated and controlled at a position contiguous to the
artifact being calibrated. A benchtop viscometer and densitometer will
be supplied by the manufacturer. These will be utilized to determine
the liquid media characteristics by sampling the fluid from the LFC.
These devices shall provide a means of control the temperature of the
sample being measured. The uncertainty in the measurements they provide
during there use shall not degrade the over system uncertainty of the
LFC. The LFC will use an optical linear encoder to determine the
displaced volume during the piston cycling. No alignment by the user
should ever be required. The LFC will include a minimum of three
pressure and six resistance temperature sensors to monitor flow media
conditions during calibration processes. Each of the above sensors will
be supplied with appropriate signal conditioners and connectors.
Interconnecting cables for each sensor will be supplied with a mating
connector. Each connector set will be appropriately marked to aid in
connecting each sensor to the correct electronics channel. Each sensor
will be supplied with an eleven point NIST traceable calibration
covering the applicable range of the sensor. The sensors in the LFC
shall be removable to allow calibration off line. The signal
conditioner, along with its sensor, shall be calibrated as a system.
The resolution of the temperature sensors shall be 0.010 C and the
resolution of the pressure sensors shall be 0.010 psig. The data
acquisition system shall be capable of recording all sensors in use by
the LFC. The placement of the pressure and temperature sensors will be
at approximate locations in the LFC. If these locations are incorrect,
it is up to the manufacturer to place the sensors where the
performance of the LFC meets the required flow rate uncertainty. One
pressure sensor shall be located near the exit manifold of the piston
cylinder tube. The other pressure sensors shall be located at the
upstream end of the test section as well as between flow meters under
test. One temperature sensor shall be located at each end of the piston
cylinder tube. The remaining temperature sensors shall be located at
along the test section, including but not limited to the upstream and
downstream location of each meter under test. The LFC will be capable
of calibrating pulse generating type flow meters manufactured by Cox,
FMSI, FT, EG&G, Potter, Foxboro, etc. The LFC electronic interfaces
will allow the connection of RF and magnetic pickoffs. Analog input
channels will be provided for 1 ea. over the following ranges: 0 -- 5
VDC, 0 -- 15 VDC, 4 -- 20 ma. 10 -- 40 ma. Each of these channels may
be used for any desirable purpose including but not limited to the
calibration of flow meters with analog outputs. These meters are
subject to the same response limitations as for the frequency inputs.
The LFC will have the capability to simultaneously calibrate at least
two separate turbine flow meters that may have magnetic or RF pickoffs.
Adapters, flow straighteners, fittings, etc., will be provided to allow
connection of flow meters having diameter sizes of 2", 1.5", 1", ", ",
3/8", etc. The manufacturing process by which the cylinder is
fabricated will be controlled to maintain a true cylinder to within
0.001" or better, the actual mean diameter being measured and recorded.
Filtering of the calibrating media will be provided in the LFC. A
filter with a nominal rating of 200 microns or better shall be
provided. All fittings and valves will be provided to gravity drain 99%
of the fluid from the LFC and to vent the air from the high points in
the system. Air or Nitrogen pressure may be used to expedite the
draining process. A stainless steel drip pan will be located to catch
fluid that drains from the test meter during removal from the LFC. The
total fluid requirements of the system will not exceed 75 gallons. The
manufacturer will supply storage tanks sufficient in capacity to house
all fluids required while using the LFC. Installation of the power
services for the LFC and its associated components, i.e., chiller, air
pressure system, data acquisition system, shall be performed by the
manufacturer of the LFC. The APSL will provide drawings to the
manufacturer that will give the needed information to make the
installation. Installation of cement pads for the chiller and air
compressor shall be the responsibility of the manufacturer. The LFC
dimensions will not exceed 25-ft. in length, 6-ft. depth, and 8-ft. in
height, not including the chiller, air compressor, or data acquisition
system. The LFC will fit through a door 12-ft wide and 10-ft high. The
controller cart will not exceed 5-ft in length, 3-ft in depth, and
6-ft in height and will be mounted on caster wheels. SPECIFIC
REQUIREMENTS FOR THE DATA ACQUISITION AND CONROL SYSTEM(DACS): The DACS
will include all of the computer hardware and software to manually or
automatically control the LFC, acquire, reduce and display the test
data in tabular and/or plotted form, and generate hard copy of the
tabulated and/or plotted calibration results on paper or magnetic
media. All original equipment manuals, supplied with each component or
software package, will be supplied with system. All data taken will be
stored to disk media and can be imported into other programs such as
Excel, Access, etc. The DACS will make use of instrumentation interface
cards which are placed in either ISA or PCI ports (or both) of the
computer itself. In the event of a computer failure or a card failure,
replacement and removal should be easy and quick. An Intel Pentium 300
MHz or faster computer with all of the supporting equipment, video
card, monitor, instrument interfaces, color printer, plotter, and
software will be supplied with the LFC. An additional hard drive and
zip drive will be supplied in computer system to allow data backup and
storage for historical and statistical use. A 17" monitor or larger
will be supplied. A HP DeskJet 895C or like printer will be provided
with the system. The software used with the DACS will be a flexible,
comprehensive, user friendly, menu driven and will run under Microsoft
Windows 98 or the latest version of Windows. The software will allow
the user to access and control all facets of the LFC data collect and
presentation. The software will allow access to all sensor data and
selection to view sensor data. Manual and automatic mode operation of
the LFC will be part of the software as well. In addition, a
calibration and system checkout section of the software will allow the
user to calibrate all instrumentation used in the LFC, i.e. encoder,
temperature sensor(s), pressure sensor(s), and all other inputs. An
operation manual will be included to afford manipulation visualization
of the software. The software includes calibration means for each
sensor including the clock, pressure, temperature, encoder, and analog
channels. The data may be curve fit using, a polynomial least squares
fit of any order up to 10th. Any number of data points may be taken
during the calibration process. All calibration data gathered while
using the LFC will be compensated for density and kinematic viscosity
changes as a function of temperature. Actual test conditions, the raw
data, and computed parameters are automatically recorded during flow
meter calibration at the conclusion of each data point. Calculations
are made by the system to correct the data in any desired manner,
including those to standard conditions such at 70 F and 14.7 psia. The
software must include a method to enter header information that is
configurable by the user. Information such as date, model, serial,
liquid medium, turbine meter manufacturer, etc. can be put in header.
Automatic flow calibration tables are stored in the program for future
use and may be retrieved at any time. All instructions are
communicated to the computer through the keyboard. Data output of the
text and graphics may be processed to the screen, hard or soft magnetic
media, printer, plotter, and/or serial or parallel ports. Software will
allow presentation of calibration data in a report format suitable to
satisfy most customer needs. Information in the form of graphs, tables,
standards used, the description of the measurement process, etc, can
all be put in this report. MAINTAINABILITY/RELIABILITY: Maintenance of
the LFC is to be kept to a minimum. The piston and shaft seals, as
well as valve seals, are normally the only items that will eventually
wear out with normal use. The layout of the LFC is such that the seals
are readily accessible. Removal of a minimum amount of plumbing and
the end flanges will facilitate the piston removal. All items that are
subject to failure such as transducers, electronic components, valves,
switches, etc. are positioned so that they can be removed and installed
with a minimum of effort and without requiring removal of other
components. Draining and filling the LFC must be readily accomplished
from one location. 99% of the fluid in the system can be drained
without removal of a single component. The system can be filled from a
single location by pouring the fluid into a reservoir or more easily
by pumping the fluid into the flow loop with a portable pump. The means
of draining and filling the LFC will be supplied by the manufacturer.
The electronics shall be designed in modular boards and discrete
components. Each board or discrete component can be removed without
unnecessary removal of other components. The manufacturer will provide
a mean time between failure of those components that are most likely
to fail in the LFC under normal usage. SAFETY ISSUE: The LFC will be
built to meet or exceed all requirements to NEC Class I Division II
GroupD for devices capable of producing arcing or sparking. MAINTENANCE
AND CALIBRATION SUPPORT PLAN: Manufacturer shall provide a complete
Maintenance and Calibration Support Plan. A minimum warranty period of
2-year will be supplied. REPAIR PARTS AVAILABILITY: A list of the
recommended spare parts, including manufacturing source, will be
incorporated into the manuals provided with the LFC. These will include
seals, gaskets, fittings, plumbing, electronic boards, sensors,
connectors, etc. This list will contain all of the items that are
subject to wear and those items that are most likely damaged by power
surges, lighting, or misuse of the machine. DOCUMENTATION: A set of
manuals on all components of the LFC and DACS will be supplied. Manuals
are to be comprehensive. Instrumentation, test data, standards,
equations, correction factors, etc. are to be listed and discussed for
all components within the LFC and DACS that affect the uncertainty of
the liquid flow measurement. (THIS DESCRIPTION IS CONTINUED IN CBD
NOTICE: SUBJECT: LIQUID FLOWMETER CALIBRATION SYSTEM (PART 2). Posted
02/24/00 (W-SN428235). (0055) Loren Data Corp. http://www.ld.com (SYN# 0249 20000228\66-0011.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|