|
COMMERCE BUSINESS DAILY ISSUE OF APRIL 14,2000 PSA#2579DOC; Mountain Administrative Support Center; Acquisition Management
Division; 325 Broadway MC3; Boulder, CO 66 -- TIME SCALE MEASUREMENT SYSTEM SOL 52RANB00C020 DUE 052600 POC
Procurement Technicians, 303/497-3221, FAX 303/497-3163 E-MAIL: NOAA;
MASC Acquisition Management, masc.solicitation@noaa.gov. The National
Institute of Standards and Technology (NIST) requires the upgrade of
the system that measures the time differences of the atomic clocks in
the NIST times scale. The measurement system is at the heart of
UTC(NIST) and requires sub picosecond stability and a very high degree
of reliability. Two independent and identical systems will be
procured. The two new systems shall have channels that accept
sinusoidal input signals from hydrogen masers or high stability cesium
beam atomic frequency standards at 5 and 100 MHz (10 MHz optional). To
provide improved accuracy and stability of the measurement system, a
low noise frequency synthesizer, which is referenced to one of the
atomic frequency standards, shall generate a reference output signal
that is coherently offset by a nominal 10 Hz. The nominal 10 Hz beat
frequencies between the output of the reference synthesizers and the 5
or 100 MHz outputs of each of a number of other atomic frequency
standards are the signals that are actually measured. The input
impedance of the RF channels shall be 50 ohms with a maximum return
loss of 30 dB at 5MHz and 20 dB at 100MHz. All RF inputs shall have DC
isolated grounds and the connectors shall be SMA. A minimum of 24
input channels in each system shall be required. There shall be a
minimum of 8 channels for 100 MHz and 16 channels for 5 MHz. At the
option of the Government, additional channels for 10 MHz may be added
or replace some of the 5 MHz channels. Ultimately 192 channels may be
required. The design shall be modular in nature so that more
measurement channels can be added at a later time. Designs based on
measurements at 1 pulse per second cannot satisfy the requirements, and
all such designs are unacceptable. The information outputted from each
measurement channel shall be: (1) the number of cycles of the nominal
10 Hz beat frequency. (2) The phase difference between the nominal 10
Hz beat from the reference oscillator and the test oscillators. (3)
Error and status information as described in the full set of
specifications. At least 24 channels shall fit in a standard 19-inch
rack with a maximum height of 34 cm (14 inches) and a maximum depth of
53 cm (21 inches). The unit shall run off of 120 VAC, single phase in
a normal room temperature environment. The computer interface shall be
standard computer hardware and shall use an industry-standard interface
design such as the PCI bus or the VME bus. Designs based on the (E)ISA
standards are not acceptable, since these interfaces are becoming
obsolete now and are unlikely to be widely supported in a few years.
The controlling computer and its operating system will be supplied by
NIST, and will be chosen based on the detailed interface design
proposed by the contractor. NIST will consider a design based on any
generally-available operating system that is currently supported by its
supplier. The maximum acceptable instability for a time difference
measurement on the 100 MHz channels is 0.06 ps TDEV at 1 s for a
measurement rate of once per second. The corresponding maximum Allan
deviation at 1s is 1e-13. For longer time intervals, tau, the Allan
deviation shall decrease as 1/tau to 1e-16 at 1000 s. At 100,000
seconds it shall not exceed 1e-17. Beyond tau equals 1e6 s the Allan
deviation shall not exceed 1e-18. Better performance is highly
desirable and a goal for the 100 MHz signal is an Allan deviation at 1
s of 1e-14. The stability goal should be such that the Allan deviation
decreases as 1/tau, or faster, to a level of 1e-19 and never exceeds
that value. The Allan deviation of the measurement systems shall be
determined by using two independent systems to measure the same clock,
and it shall be assumed that the two systems are equivalent. Similar
tests shall also be performed with two channels on the same system.
Delivery of the two systems shall be 240 calendar days after
Contractor's receipt of a notice of award. Delivery shall be F.O.B.
Destination to NIST, 325 Broadway, Boulder, Colorado 80303-3328.
Standard Industrial Classification code is 3825. Small business size
standard is 500 employees. See Numbered Note 1. Posted 04/12/00
(D-SN444122). (0103) Loren Data Corp. http://www.ld.com (SYN# 0248 20000414\66-0001.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|