COMMERCE BUSINESS DAILY ISSUE OF JUNE 1,2000 PSA#2612DOC; Mountain Administrative Support Center; Acquisition Management
Division; 325 Broadway, MC3; Boulder, CO 66 -- POSITIONING STAGE SOL NB853-004052DT DUE 061900 POC Doris
Turner, Purchasing Agent, 303-497-3872 or FAX: 303-497-3163 E-MAIL:
NOAA; MASC Acquisition Management, Doris.P.Turner@noaa.gov. The
National Institute of Standards and Technology (NIST) requires a
positioning stage to be used to position specimens for x-ray
diffraction studies. It will be mounted on an existing Huber
kappa-geometry goniometer. This dictates requirements on physical
dimensions of the x- y-z positioning stage, to be able to accommodate
unlimited rotations about the 4 goniometer axes. Different applications
of this goniometer-positioning stage define requirements for
positioning precision, repeatability and accuracy, as listed on the
following specification list. Current research programs in the
Materials Reliability Division include studies of combinatorial
libraries of materials and residual stresses on the micrometer scale,
both requiring high- resolution spatially resolved measurements. In
addition to the specifications below, we would prefer a z-stage with a
horizontal top surface; however, the total height of the x-y-z
assembly cannot exceed the specified height limitation. The following
specifications must be met. Stage travel for all three axes shall be
greater than 20 mm. Positioning for each axis shall be: 1. Precision:
better than 0.1 micrometer, 2. Repeatability: better than 0.7
micrometer, 3. Accuracy: better than 1 micrometer. Dimension
requirements are: 1. Both length and width of the complete x-y-z stage
(in the plane of the mounting surface) shall be less than 160 mm, 2.
Height of stacked x-y stages shall be less than 60 mm, 3. If z-stage is
mounted at 90 degrees to the mounting surface (angle bracket), the
total x-y-z stage height may exceed 100 mm, 4. If z-stage is mounted on
the top of the x-y stages with the specimen mounting surface parallel
to the base mount, the total x-y-z stage height shall be less than 100
mm (this configuration is preferred). Stage load capacity shall be
greater than 0.5 kg. Computer control: shall accept commands from
within Linux OS environment and SPEC diffractometer-control software.
Interested sources proposing to supply the specified equipment must
submit a written qualification statement clearly establishing
capability to meet the above requirements within 15 days from the date
of this synopsis. This best value procurement will be evaluated on
technical merit and price. Responses are due via mail or FAX by June
19, 2000. No RFQ package is available. This procurement is being
conducted per Simplified Acquisition Procedures FAR Part 13 (NTE
$100,000). FAR Clauses in full text are at http://www.arnet.gov/far/.
Terms and conditions and agency level protest procedures are referenced
at http://oamweb/osec.doc.gov/conops/#REFERENCE Posted 05/30/00
(D-SN459677). (0151) Loren Data Corp. http://www.ld.com (SYN# 0326 20000601\66-0017.SOL)