COMMERCE BUSINESS DAILY ISSUE OF JUNE 26,2000 PSA#2629Contracting Officer (90C), Department of Veterans Affairs, North Texas
Health Care System, 4500 South Lancaster Road, Dallas, TX 75216 65 -- RECESSED PATIENT WALL STATIONS SOL 549-68-00 DUE 071100 POC
Belinda L. Cook, 214-857-0037 Send completed Quotations referring the
solicitation number with attention to Contracting Officer to: VA North
Texas Health Care System located at 4500 S. Lancaster Rd., Dallas, TX
75216, here in after referred to as VANTHCS contemplates the purchase
of a Medical Solutions International or equal (1) Bicarb Mixing and
Distribution and Acid Distribution Systems for Hemodialysis. The
successful offeror shall be required to provide; (1) FOB destination
delivery, (2) three of each operating and service manual; (3) "on-site"
system training; (4) Installation; (5) and a full year parts and labor
warranty to include a preventive maintenance and emergency repair
system at no extra cost to VANTHCS. The product must meet the minimum
salient characteristics as defined in this solicitation. Each offeror
shall be required to provide descriptive literature with his/her offer.
GENERAL MINIMUM SPECIFICATION Bicarb Mixing andDistribution and Acid
Distribution Systems for Hemodialysis Consisting of: (1 Each)
Bicarbonate mixing and distribution system to service 27 patient
stations. (2 each) Acidified concentrate distribution system to service
27 patient stations. (2 Each) 300 gallon acid storage tanks The system
must have the capability of easy sanitation and clean up. The system
must be low pressure 2 PSI at each station to accommodate Fresenius
2008H dialysis equipment with a flow rate of 800ml/min (gravity or pump
fed with backup pump in line). All pumps must be backed up with
redundant pumps to insure any pump malfunction may be back on line
within five minutes. Recessed patient walls stations to house two each
acid connectors and one each bicarb connector A 800 ml min dialysate
flow at each 2008H Fresenius machine Bicarb flow rate 1 oz per minute
per machine Acid flow rate 1 oz per minute per machine All plumbing,
pumps, regulators, monitors, values and returns must be installed by
the vendor. Electrical and any core drilling will be performed by our
Engineering Department. All Quotations will be reviewed by VA Technical
Representative to determine if minimum specification have been met.
Award will be made to the lowest priced offer meeting all requirements
of this Quotation. Line Product Description QTY UN Unit Total Item
Cost Cost 1. Recessed Patient Wall Stations 30 Ea. $ $ 2. Bicarbonate
Valving 30 Ea. 3. Acidified Valving 60 Ea. 4. End Point Interface
Fitting 90 Ea. 5. Coiled Concentrate Tubing Lines 60 Ea. 6. Bicarbonate
Mixing and Distribution System 1 Ea. 7. Acidified Distribution System
1 Ea. 8. Acidified Bulk Storage Tanks 3 Ea. 9. Concentrate Loops (3
Each 700 Foot Loops) 3 Ea. Bidding On: Manufacturer
name_________________________________________________ Brand Name &
Product No.__________________________________________ Posted 06/22/00
(W-SN467475). (0174) Loren Data Corp. http://www.ld.com (SYN# 0278 20000626\65-0001.SOL)
65 - Medical, Dental and Veterinary Equipment and Supplies Index Page
|
|