|
COMMERCE BUSINESS DAILY ISSUE OF JULY 10,2000 PSA#2638National Institutes of Health, National Heart, Lung, and Blood
Institute, Contracts Operations Branch, 6701 Rockledge Drive, Room
6100, MSC 7902, Bethesda, MD 20892 66 -- MAGNET AND INTEGRATED GRADIENT SHIMS SYSTEM SOL
NHLBI-PS-2000-631 DUE 072100 POC Mr. Livie V. Spearman, III, Contract
Specialist, (301) 435-0352; Ms. Debra C. Hawkins, Chief, Procurement
Section, (301) 435-0366 WEB: Requests For RFP, spearmal@nhlbi.nih.gov.
E-MAIL: hawkinsd@nhlbi.nih.gov, spearmal@nhlbi.nih.gov. It is the
intent of the National Heart, Lung, and Blood Institute to purchase a
Magnet and its integrated gradient and shim system, to be part of a
1.5T MRI Scanner for the Laboratory of Cardiac Energetics, Division of
Intramural Research. The specifications for the Magnet and Gradient
are as follows: Section I -- Specifications for the 1.5 Tesla Magnet,
a) The magnet should provide a constant magnetic field of 1.5T at its
center. The diameter of its inner open bore should be 900 mm or larger.
Its overall length should be less than 2m. b) The system must include
all cryogenic equipment, such as cryocooler, compressor and cryogen
liquid level monitoring devices. It should also include emergency
quench units, magnet power supply, and any other equipment that is
needed to bring the magnet to, and maintain it at, its constant field
level, & ensure safe ramp down in the case of quenching. c)
Installation of magnet on site and consultation with NIH staff to
ensure that the proper site requirements for the magnet are met. This
includes bringing the magnetic field to a persistent level of 1.5 T,
and installation of all associated equipment, as described in section
b). On-site acceptance test procedure, including field homogeneity and
stability tests are required. d) Magnet shimming and field homogeneity
tests should be performed with the gradient coil in place. e) Field
homogeneity specifications with the gradient coil in place: the
standard deviation in a 35 cm DSV should be below 20 Hz (or 10Hz). The
peak-to-peak range in a 35 cm DSV should be less than 2 ppm. f) The
magnet's field drift should be less than 0.1 ppm/hour. g) The minimum
time interval between helium refilling should be more than 6 months,
with helium boiloff rate. h) Warranty information and maintenance
contracts related to the magnet are to be specified. Section II --
Specifications for the Gradient and Room-Temperature Shim Systems -- A
40cm inner-bore actively shielded gradient coil/RT shim coil system
should be installed in the 1.5T magnet. The gradient coil can be driven
with our existing amplifiers of 200 A peak current, 300 V peak voltage,
or by an amplifier system you supply. An existing chilled-water cooling
system (Neslab Coolflow System II) is available for cooling the
gradient coil. The following are requirements on the gradient coil and
RT shims: a) The gradient coil should have a 40 cm diameter inner
clear bore. b) The room-temperature shim coils should include z0 (for
B0 eddy current compensation), and second order shims including z2, zx,
xy, xz, x2-y2. The z0 coil should be able to provide 30 ppm field with
2 A input. The second order shim coils should each be able to provide
more than +/-10ppm field in 40 cm DSV with 2 A input. The response
time of all shim channels should be 10 microsec or less. c) For each
axis, the gradient field should be within +/- 3% of the linear function
within a 20cm DSV around the iso-center. d) The gradient strength of
each axis should exceed 5 G/cm/100A. e) The maximum gradient ramp rate
should exceed 500 mT/m/ms. This should be realized either with our
existing amplifiers of 200A peak current and 300V peak voltage, or an
amplifier system you supply. f) Sufficient cooling measures and
temperature monitoring should be built into the gradient coil so that
it can operate at near peak gradient strengths and slew rates on all 3
axes continuously. The gradient coil should have embedded
thermosensors for temperature monitoring. g) To assess potential
external eddy currents, information on the external fields generated by
the gradient coils, such as vector field plots external to the gradient
coil is required. h) To assess potential eddy currents, actual
measurements or tolerances on couplings between the gradient coils and
shim coils is required. i) Information on the isolation between the
three gradient channels, e.g., mutual inductances between the gradient
coils is required. j) Performance references, if they exist, on prior
identical or similar gradient coils built by your company is required.
k) The gradient coil should be rigidly mounted in the magnet. The
solicitation, negotiation and award of this requirement will be in
accordance with Federal Acquisition Regulations (FAR) Part 12,
Acquisition of Commercial Item procedures. The award of a firm,
fixed-price type contract is anticipated. There are no set-aside
restrictions for this requirement. The applicable Standard Industrial
Classification code (SIC) is 3845, size standard 500 employees.
Interested parties must request a copy of the solicitation, number
NHLBI-PS-2000- 631, from the following address: Attention: Mr. Livie V.
Spearman, III., U.S. Department Of Health & Human Services, Public
Health Service, National Institutes of Health, National Heart, Lung, &
Blood Institute, Division of Extramural Activities, Contracts
Operations Branch, Procurement Section, Rockledge II Centre, 6701
Rockledge Drive, Room 6148, Bethesda, MD 20892-7902. Requests for
copies of the solicitation must be in writing and may be requested by
FAX to (301) 480-3345 and will only be accepted if dated & signed by an
authorized company representative. The solicitation will be available
approximately ten (10) calendar days from the publication date of this
synopsis. The solicitation will be released to all offerors requesting
a copy. All responsible sources may submit a proposal, which if timely
received shall be considered by the agency. Posted 07/06/00
(W-SN472037). (0188) Loren Data Corp. http://www.ld.com (SYN# 0223 20000710\66-0001.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|