COMMERCE BUSINESS DAILY ISSUE OF JULY 12,2000 PSA#2640 DOC; Mountain Administrative Support Center; Acquisition Management
Division; 325 Broadway MC3; Boulder, CO 66 -- MILLIMETER WAVE MODULES SOL NB813000003215MB DUE 080400 POC
Michelle Bernal, Purchasing Agent, (303) 497-3983 or FAX (303) 497-3163
E-MAIL: NOAA; MASC Acquisition Management, Michelle.A.Bernal@noaa.gov.
The National Institute of Standards & Technology(NIST) has a
requirement for millimeter wave modules as part of a millimeter wave
measurement system (50 to 110 Ghz). The critical specifications are as
follows: (A) WR-15 Specifications (transmit): 1) System operating
frequency equals 50 to 75Ghz; (receiver) 50 to 75Ghz. 2) Source input
frequency equals 12.5 to 20 Ghz; (receiver) n/a. 3) Source input power
equals plus 5dbm to plus 13dbm; (receiver) n/a. 4) test port output
power equals plus 2dbm minimum plus 7 dbm; (receiver) n/a. 5) Test port
input power equals minus 10 dbm input signal; (receiver) minus 10dbm
input signal at 0.1 db comp. 6) Mixer LO frequency equals 10 to 15Ghz
(x5); (receiver) 10 to 15Ghz (x5). 7) Mixer LO drive equals plus 5 to
plus 13 dbm; (receiver) plus 5 to plus 13 dbm. 8) Test dynamic range
equals 80dbm minimum; (receiver) 80dbm minimum. 9) If frequency range
equals 10 to 300 Mhz; (receiver) 10 to 300 Mhz. 10) Coupler directivity
equals 35db minimum; (receiver) n/a. 11) Testinterface equals
MIL.-f-3922/67b-008(ug 385/u) for MMW test port. (B) WR-10
Specifications (transmit): 1) System operating frequency equals 75 to
110 Ghz; (receiver) 75 to 110 Ghz. 2) Source input frequency equals
12.5 to 18.0 Ghz(x6); (receiver) n/a. 3) Source input power equals plus
5dbm to plus 13 dbm; (receiver) n/a. 4) Test port output power equals
minus 5dbm minimum; (receiver) n/a. 5) Test port input power equals
minus 10 dbm at .1db; (receiver) minus .10 dbm at .1 db. 6) Mixer LO
frequency equals 9.3 to 13.8 Ghz (x8); (receiver) 9.3 to 13.8 Ghz (x8).
7) Mixer LO drive equals plus 5 to plus 13 dbm; (receiver) plus 5 to
plus 13 dbm. 8) Test dynamic range equals 80 db minimum; (receiver) 80
db minimum. 9) IF frequency range equals 10 to 300 Mhz; (receiver) 10
to 300 Mhz. 10) Coupler directivity equals 35 dbm; (receiver) n/a. 11)
Test interface equals MIL.-F-3922/67B-010(UG 387/U-M) for MMW test
port. Note: These modules must be self contained, not in individual
parts, with an RF input and LO input also a Test Port output and Test
Channel output . Technical questions must be submitted in writing to
the Contracting Officer by fax or by email. This procurement is being
conducted per FAR Part 13, Simplified Acquisition Procedures. This
requirement is set-aside for small business unless small business
cannot meet the Governments need at a reasonable cost. You may submit
your proposal by fax or by mail. Proposal must be received on or before
close of business August 4, 2000. Please include delivery date, F.O.B.
point, business size, Federal Taxpayer Identification Number, and Duns
and Bradstreet Number. Terms and Conditions and agency level protest
procedures can be fond on web site http: oamweb.osec.doc.gov/conops
#REFERENCE. Hard copy in full text available from office address and
phone number listed in notice. Anticipated award date August 8, 2000.
Posted 07/10/00 (D-SN472882). (0192) Loren Data Corp. http://www.ld.com (SYN# 0289 20000712\66-0009.SOL)