COMMERCE BUSINESS DAILY ISSUE OF AUGUST 4,2000 PSA#2657 USDA, ARS, BHNRC, Building 308, Room 226A, BARC-EAST, Beltsville, MD
20705 66 -- FLUORESCENCE SPECTROPHOTOMETER WITH ATTACHED MICROPLATE READER
SOL 001-1235-125-40043 DUE 082100 POC Kim T. Horton 301/504-8557 FAX
301/504-5235 E-MAIL: Kim Horton, Contracting Officer, Beltsville Human,
hortonk@ba.ars.usda.gov. This is a combined synopsis/solicitation for
a commercial item prepared in accordance with the format in Subpart
12.6, as supplemented with additional information included in this
notice. This announcement constitutes the only solicitation; quotes are
being requested and a written solicitation will not be issued.
Solicitation number 001-1235-125-40043 is issued as a request for
quotation (RFQ). This solicitation document is reserved for 100% small
business. The SIC Code is 3826, and the small business standard in
500. The Agency need is for a Fluorescence Spectrophotometer with
Attached Microplate Reader. The following specifications are necessary
for the Fluorescence Spectrophotometer: (1) Wavelength ranges --
Excitation: 200-900 nm. Zero order selectable. Emission: 200-900 nm.
Zero order selectable. (2) Limit of detection: <1.0 pM fluorescein
using standard cell; <10 pM fluorescein using 40 ul microcell;
<25 pM QBS in standard cell. (3) Minimun sample volume: 0.5 ml
(using standard 10mm fluorescence cell) (4) Signal averaging:
Fluorescence 0.0125 to 999 sec; Phosphorescence 1 usec to 10 sec (gate
time); Bio/Chemi-Luminescence 40 usec to 10 sec (gate time) (5)
Temperature monitors (temperature probe inside curvette) Specifications
for the Microplate Reader include the following: (1) Reading Modes:
Fluorescence; Phosphorescence; Bio/Chemi-Luminescence (2) Measuring
Types: Wavelength scanning; Kinetics time scanning; single and
multiwavelength reads; time-resolved fluorescence and phosphorescence.
(3) Sensitivity: 3 fmol/well fluorescein in a black well. Well-to-well
repeatability <2.0% for every well of a 96 well microplate. (4) Must
be able to read up to 384 well format. FOB point is destination. Please
include installation and any accessories in the price. The following
provisions and clauses are included in the solicitation: 52.202-1,
52.203-5, 52.203-6, 52.203-7, 52.204-6, 52.212-1, 52.212-2, 52.232-1.
The Government will award a contract to the responsible offeror whose
offer, confirming the solicitation, will be most advantageous to the
Government, price and other factors considered. Other provisions and
clauses included are: 52.212-3; which must be completed and returned
with your quote, and 52.212-4, 52.212-5, 52.216-24, 52.216-25,
52.225-13. Include numbered Notes 1 and 23. Solicitations can be mailed
to Kim Horton, USDA, ARS, BHNRC, Bldg 308, Rm 226, BARC-East,
Beltsville, Maryland 20705-2350 or faxed to 301/504-5235. Posted
08/02/00 (W-SN481486). (0215) Loren Data Corp. http://www.ld.com (SYN# 0528 20000804\66-0007.SOL)