COMMERCE BUSINESS DAILY ISSUE OF AUGUST 21,2000 PSA#2668 National Institutes of Health, Office of Procurement Management, 6011
Executive Blvd., Rockville, Maryland 20892-7260 66 -- DUAL CHAMBER HOT CELLS POC Catherine Hiltner, Contracting
Officer, (301) 402-3063 The National Institutes of Health, National
Institute of Mental Health intends to negotiate with Von Gahlen
International, Inc., 505 South Second Street, Chatsworth, GA 30705, to
furnish and install a quantity of three (3) dual-chamber hot-cell
systems meeting the following specifications: (1) Shall include two
laminar flow inserts; (2) overall dimensions of each dual unit hot-cell
shall be 45.5 inches deep by 112 inches wide by 97 inches high
including work stand. All external surfaces of each hot-cell and stand
shall be epoxy coated. (3) Inside dimensions of each separate hot-cell
shall be 36 inches deep by 49 inches wide by 46.5 inches high; (4)
Shall be shielded with three inches of lead in each direction including
sidewalls, floor, and roof; (5) Each separate hot-cell shall be
equipped with motor driven doors in the front and rear walls. Each rear
door shall be solid. Each front door shall contain a glass window. Each
window shall be 14.75 inch wide by 11.75 inch high by 6 inch thick,
clear 5.2 density lead glass. All fron doors shall include two each
8.25 inch glove ports located at least 14 inches apart and located
below the window. (6) All doors shall include a safety interlock; (7)
All door openings shall be 37.5 inches wide by 28 inches high; (8) In
addition to the windows in the front doors, tow of the dual units shall
include windows on both ends. The third dual unit shall include only
one window on the right end. All windows shall be 13.75 inches wide by
11.75 inches high by 6 inches thick made of 5.2 density lead glass.
Each window shall be mounted lengthwise and centered on each end panel.
Each end having a window shall be equipped with a 3 inch diameter
stepped plug for access lines located in the center below windows. (9)
Each separate hot-cell shall be lined with 16 gage stainless steel.
The dual unit having one end window shall have both liners air tight
and be GMP certified. (10) Each separate hot-cell shall be equipped
with internal lighting fixtures, mounted on ceiling center. (11) Each
separate hot-cell shall be equipped with a 6 inch diameter exhaust pipe
located in the center back of the ceiling. (12) Each separate hot-cell
shall include 16 individually switched electrical outlets located on
sides and back of work area, a minimum of two inches above floor. (13)
Each separate hot-cell shall include a low voltage control panel to
operate lights. The provided equipment and services are proprietary and
commercial as defined in FAR Part 2. Von Gahlen International, Inc. is
the only known source capable of providing the required equipment.
This action is under authority of 41.U.S.C. 253(c)(1), as set forth in
FAR 6.302-1 and HHSAR 306.302-1. Only one responsible source and no
other suplplies or services will satisfy agency requirements.
Interested parties may identify their interest and capability to
respond to the requirement or submit proposals. This notice of intent
is not a request for competitive proposals. However, all proposals
received in the contracting office identified herein within 15 days
after publication of this synopsis will be considered by the
Government. A determination by the Government not to compete this
proposed contract based upon responses to this notice is solely within
the discretion of the Government. Information received will normally
be considered solely for the purpose of determining whether to conduct
a competitive procurement. Posted 08/17/00 (W-SN487091). (0230) Loren Data Corp. http://www.ld.com (SYN# 0270 20000821\66-0008.SOL)